Hardness value of Leeb
standard test block
Error of
displayed value
Repeatability of
displayed value
760±30HLD ±6 HLD 10 HLD
530±40HLD ±10 HLD 10 HLD
3.1.4 Measuring range:170~960HLD;
3.1.5 Measuring direction:360°;
3.1.6 Scales:HL、HB、HRB、HRC、HV、HS;
3.1.7 Display:128×32 dot matrix LCD;
3.1.8 Data memory:270 average data in 9 files;
3.1.9 Upper and lower limit setting:same to measuring range;
3.1.10 Power:AAA(7#)1.5V 2pc;
3.1.11 Continual working time :about 150 hours(without backlight);
3.1.12 Communication interface:USB2.0.
3.2 Dimension and weight
3.2.1 Dimension:155mm×55mm×25mm
3.2.2 Weight: about 166g.
4 Using
4.1 Preparation before using
4.1.1Preparation for sample surface
Preparation for sample surface should conform to the relative requirement in figure 3.
●In the preparation process for sample surface, the hardness effect of being heated or cold
processing on the surface of sample should be avoided;
●Too big roughness of the measured surface could cause measure error. So, the measured surface
must be metallic luster, smoothing and polish, without oil stain;
●Curved surface: The best testing surface of sample is flat. When the curvature radius R of the
surface to be tested is smaller than 30mm,the small support ring or the shaped support rings
should be chosen, see figure 2.
Figure 2
●Support of test sample
Support is not necessary for heavy sample.
Medium-weight parts must be set on the smoothing and stable plane. The sample must set