Preface
EC300XL i

Important Read this instruction manual. Failure to read,
understand and follow the instructions in this manual may
result in damage to the unit, injury to operating personnel,
and poor equipment performance. s
Caution All internal adjustments and maintenance must be performed by qualified service personnel. s
This equipment has been designed and tested to conform to CSA1010 safety standards, as applicable to
laboratory instrumentation. This applies only to the EC300 when used as specified in the documentation, in its
intended applications, with Thermo Fisher Scientific approved electrophoresis apparatus only. Usage in any other
manor may not provide similar performance or safety protection.
This equipment is provided with a 3-conductor, grounded AC line cord. The protective earth ground is necessary
for safe operation. Do not use any other AC line cord with this instrument.
The EC300 is a high voltage power supply capable of generating dangerous levels of voltage and current during
operation. Exercise caution when working around and with the electrical connections of this equipment. Always
check electrical connectors, wires, and associated apparatus for any signs of wear or damage before using with
this equipment. Be sure to use only electrophoresis equipment that is suitably rated for the voltage and current
capabilities of the EC300 power supply.
The output of the EC300 power supply is intended for connection to electrically isolated electrophoresis
apparatus only. Use only with electrically isolated electrophoresis apparatus with minimum isolation of 600V. Do
not connect any terminal of the EC300 output to earth ground. This may impair the safety protection provided
by the equipment, or cause equipment damage.
The high voltage output of the EC300 power supply takes some amount of time to decay when unloaded or
lightly loaded. Wait a minimum of 60 seconds after stopping a run before touching the power supply leads.
This equipment has a protective ground leakage current of less than 0.5mA using test methods defined in
CSA1010 and CSA151.
This equipment is for indoor use only.
Material in this manual is for information purposes only. The contents and the product it describes are subject to
change without notice. Thermo Fisher Scientific makes no representations or warranties with respect to this
manual. In no event shall Thermo be held liable for any damages, direct or incidental, arising out of or related to
the use of this manual.
©2012 Thermo Fisher Scientific. All rights reserved.
Thermo Scientificii EC300XL
Preface

Thermo Scientific EC300XL iii
Preface
Important operating and/or maintenance instructions. Read the accompanying text carefully.
Potential electrical hazards. Only qualified persons should perform procedures associated with this
symbol.
Equipment being maintained or serviced must be turned off and locked off to prevent possible injury.
Hot surface(s) present which may cause burns to unprotected skin, or to materials which may be
damaged by elevated temperatures.
Marking of electrical and electronic equipment, which applies to electrical and electronic equipment
falling under the Directive 2002/96/EC (WEEE) and the equipment that has been put on the market
after 13 August 2005.
This product is required to comply with the European Union’s Waste Electrical & Electronic
Equipment (WEEE) Directive 2002/96/EC. It is marked with the WEEE symbol. Thermo Fisher
Scientific has contracted with one or more recycling/disposal companies in each EU Member State
European Country, and this product should be disposed of or recycled through them. Further
information on Thermo’s compliance with this directive, the recyclers in your country and
information on Thermo products will be available at www.thermofisher.com.
4Always use the proper protective equipment (clothing, gloves, goggles, etc.)
4Always dissipate extreme cold or heat and wear protective clothing.
4Always follow good hygiene practices.
4Each individual is responsible for his or her own safety.

Thermo Scientificiv EC300XL
Preface
Do You Need Information or Assistance on
Thermo Scientific Products?
If you do, please contact us 8:00 a.m. to 6:00 p.m. (Eastern Time) at:
1-740-373-4763 Direct
1-800-438-4851 Toll Free, U.S. and Canada
1-877-213-8051 FAX
http://www.thermoscientific.com Internet Worldwide Web Home Page
service.led.marietta@thermofisher.com Tech Support Email Address
Certified Service Web Page
Thermo Fisher Scientific
401 Millcreek Road, Box 649
Marietta, OH 45750
Our staff can provide information on pricing and give you quotations. We can
take your order and provide delivery information on major equipment items or make
arrangements to have your local sales representative contact you. Our products are listed on the
Internet and we can be contacted through our Internet home page.
Our staff can supply technical information about proper setup, operation or
troubleshooting of your equipment. We can fill your needs for spare or replacement parts or
provide you with on-site service. We can also provide you with a quotation on our Extended
Warranty for your Thermo Scientific products.
Whatever Thermo Scientific products you need or use, we will be happy to discuss your
applications. If you are experiencing technical problems, working together, we will help you
locate the problem and, chances are, correct it yourself...over the telephone without a service
call.
When more extensive service is necessary, we will assist you with direct factory trained
technicians or a qualified service organization for on-the-spot repair. If your service need is
covered by the warranty, we will arrange for the unit to be repaired at our expense and to your
satisfaction.
Regardless of your needs, our professional telephone technicians are available to assist you
Monday through Friday from 8:00 a.m. to 6:00 p.m. Eastern Time. Please contact us by
telephone or fax. If you wish to write, our mailing address is:
International customers, please contact your local Thermo Scientific distributor.
Sales Support
Service Support
www.unitylabservices.com

EC300XL vThermo Scientific
Table of Contents
Introduction . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .1-1
Unpacking the Power Supply . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .1-2
Specifications . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .1-2
Set-Up and Operation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .2-1
Using the Power Supply . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .2-2
Constant Current Operation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .2-3
Constant Voltage Operations . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .2-3
Timed Operations . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .2-4
Automatic Power-fail Restart . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .2-4
Troubleshooting . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .3-1
Maintenance . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .4-1
Cleaning . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .4-1
Replacing a Fuse . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .4-1
Replacement Parts . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .4-1
Accessories . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .4-1
Section 1
Section 2
Section 3
Section 4


EC300XL 1-1Thermo Scientific
Section 1 Introduction
Thank you for selecting a Thermo Scientific EC300 Electrophoresis power
supply. This manual describes the operation of the EC300. The power
supply that you have purchased is the most productive and easy-to-use unit
available anywhere. This manual should answer any questions that might
arise in operating your power supply; however, don't hesitate to call our
Technical Services department if you need any assistance.
The EC300 power supply is designed to provide constant voltage or
constant current output to apparatus used in electrophoresis applications.
One to four sets of electrophoresis cells can be connected in parallel and
run simultaneously. The EC300 can deliver up to 75W of total output
power. When operating in constant voltage or constant current mode, the
power supply automatically limits the other parameter to either the power
supply maximum, or a lower limit if set by the user. If this non-constant
limit is reached, the power supply will automatically switch control modes,
from constant voltage to constant current, or vice versa. In this way, the
EC300 protects your electrophoresis cells from damaging over power
conditions. The EC300 power supply also provides for timed operation in
either voltage or current modes, and allows an automatic completion in
the event of a power loss if enabled by the user.
• 10-300V, adjustable in 1 volt steps
• 4-400mA, adjustable in 1mA steps
• 75W maximum output
• Automatic control mode crossover
• 0-999 minute timed run
• Automatic restart if loss of AC power (if enabled)
• Bright 3-digit display

1-2 EC300XL Thermo Scientific
Section 1
Introduction
Specifications
When unpacking your EC300 power supply, be sure you have received
the EC300 unit plus the AC line cord.
Inspect your equipment and packaging material for signs of damage.
Damage to the shipping container may indicate rough handling which
could cause internal damage to the power supply. If you suspect shipping
damage to the power supply, contact your carrier for instructions on filing
a claim. If you are missing any of the above items, contact your supplier
for instructions.
AC Input Power
100-120VAC, 50-60Hz, 100VA Max
Environmental
Operating temperature: 0-40°C, 0-95% R.H. non-condensing
Altitude: 2000m
Overvoltage category II, IEC664
Pollution degree 2, IEC664
DC Output Power
10-300VDC, 75W Max
4-400mA, 75W Max
Ripple: ±1%
Drift: ±1%, after 30-minute warm-up
Compliance
CSA 1010
This equipment has been designed and tested to conform to CSA1010
safety standards, as applicable to laboratory instrumentation. This
applies only to the EC300 when used as specified in the documenta-
tion, in its intended applications, with Thermo Scientific approved elec-
trophoresis apparatus only. Usage in any other manner may not provide
suitable protection.
Unpacking the Power
Supply

EC300XL 2-1Thermo Scientific
Section 2 Set-Up and Operation
Select a location that allows for 3" clearance behind the power supply, and
comfortable reach of the front panel controls and cell connections. Do not
block the vented area of the case - on the front bottom of the unit, or the
fan area at the rear. Connect the electrophoresis apparatus to the power
supply, making sure to match the red positive lead to the red positive jack,
and the black negative lead to the black negative jack. Power the unit on
by connecting the AC power cord to the power entry on the rear panel
then plugging the 3 prong AC plug into a power source. Connect the
power supply to a 3-prong grounded AC outlet, using the AC cord
provided with the unit only.
Red positive
jacks
Up/Down arrows
Constant key (K)
Black negative
jacks
Select key (Set)
Run key
Blue On/Off key
3 digit LED display
Figure 2-1. Control Panel
Table des matières
Autres manuels Thermo Scientific Alimentation

Thermo Scientific
Thermo Scientific invitrogen PowerEase Touch Manuel utilisateur

Thermo Scientific
Thermo Scientific PowerEase PS0300 Manuel utilisateur

Thermo Scientific
Thermo Scientific EC200XL Manuel utilisateur

Thermo Scientific
Thermo Scientific Invitrogen PowerEase Touch PS0120 Manuel utilisateur

Thermo Scientific
Thermo Scientific Invitrogen PowerEase Touch PS0350 Manuel utilisateur