Terumo FineTouch Manuel

1
Product list
Blood glucose monitoring system for self-testing
mmol/L
Blood Glucose Meter
Guide Book


3
• Before using the FINETOUCH Blood Glucose Meter carefully read through this “Guide Book” and carry
out the measurements in accordance with the instructions from your physician.
• Discuss any questions you might have with your physician prior to your first measurement.
• For your convenience an abridged guide book is supplied with shortened instructions for you to carry
with you.
• If you have any doubts about the measurement results, make sure to consult your physician.
• Let your physician guide you in the management of your diabetes treatment regimen. In particular, do
not alter the dosage and administration frequency of your oral medication and/or insulin at your own
discretion or at the judgment of any member of your family unless he/she is a licensed physician.
• Carefully read the instruction manual of the medical devices to be used prior to testing.
• In case of suspected hypoglycaemia, collect the blood from the finger or earlobe. Blood glucose test
results using samples taken from the arm may differ from fingertip or earlobe samples.
Caution: Carefully follow these instructions:

4
Attach the
Test TIP
Prick
Draw up blood
Test TIP Meter
Lancet Lancing Device
Meter
Beep
Completes the
measurement
in 10
seconds
After blood is applied to the Test TIP, glucose in the blood
reacts with the reagent on the Test TIP, and colours the
reagent pad. The Meter measures this colour concentration,
and shows the blood glucose level on the display.
Outline of how to use

5
Contents
Before use ................................................................................................................................................... 6
How to use the FINETOUCH
Blood Glucose Monitoring System
................................................................................13
Problem solving .................................................................................................................................32
(Read this section when you receive the wrong indication
or when you are not able to measure in the usual way)
Maintenance of the Meter
........................................................................................................44
Additional functions
........................................................................................................................46
Cautions (for safety / on use / for storage) ......................................................57
Waste and Recycle of Meter and batteries ...................................................62
Meter Specifications ...................................................................................................................63
Product list ..............................................................................................................................................64
Warranty information ..................................................................................................................67
Page

6
Before use
After blood is applied to the FINETOUCH Test TIP, glucose in the blood reacts with the reagent
(enzyme glucose oxidase) on the Test TIP, and colours the reagent pad. The intensity of the colour
is proportional to the amount of glucose in the blood. The FINETOUCH Meter measures this colour
concentration and shows the blood glucose level on the display. Test results show “plasma” blood
glucose concentration as per recommendation of the International Federation of Clinical Chemistry
and Laboratory Medicine (IFCC) 1). This makes it easier for you and
your diabetes-care team to compare your meter results with laboratory
tests. If you have been using another type of meter – one that provides
whole-blood results – you may notice that your test results with the
FINETOUCH Blood Glucose Monitoring System are about 11% higher.
1) Clin Chem Lab Med 2006; 44 (12): 1486-1490
Before use
Diabetes mellitus is a disease resulting in a high glucose level in the blood. Untreated diabetes
mellitus has been known to lead to the impairment of the liver, kidney, eyes and nervous system.
There are many treatment procedures for diabetes mellitus today, and the first important step is not
to panic but consult a physician. It is essential that diabetics measure their own blood glucose level
and keep the doctor informed about any changes in their blood glucose levels. This is the best way
to ensure a more effective course of treatment.
Important! There are two different units to indicate blood glucose
levels (mg/dL and mmol/L). Carefully check if your Meter is set on
mmol/L. Otherwise, you can misinterpret results. In case your meter is
set at the wrong unit, please contact Terumo immediately.

7
Before use
Make sure to follow the instructions in
order to ensure safe/correct use
Caution:
• Exposure to blood may spread infectious diseases.
Make sure that blood or devices contaminated with blood, including
tissue paper and other items you have used, do not come into
contact with any other person.
• Always keep FINETOUCH products away from children.
There is a risk of children accidentally swallowing the lancet,
batteries, Blood Glucose Test TIP, Blood Glucose Test TIP case, etc.
In case that any of these parts have been swallowed, immediately
consult a physician.
• Do not reuse used lancets.
This may result in blood infection.
• Dispose of the used Test TIP and used Lancet as instructed by your
physician or in accordance with local regulations for disposal of
contaminated materials.

8
Before use
• Meter • Lancing Device
• Guide Book
• Test TIP
• Lancet
• Meter-Check TIP• Lithium battery
(2 x CR2032)
TIP attachment
“POWER” button “ ”
Display
“RECALL” button “ ”
“TIME SET”
button “ ”
Interface
Eject button
Protective cap
Dial
Set Lever
Push Button
Battery
box
Back
Make sure to use FINETOUCH Test TIP with BLUE lm-seal.
Check if the contents of the kit are complete. If something is missing, please contact Terumo (you will nd contact informa-
tion at the end of the guide book). Do not use if damaged.
• Soft bag
FINETOUCH Blood Glucose Monitoring System
Composition and Content

9
Before use
12
Placing the
batteries
Close the
battery box
Slide the lid whilst
pressing part.
Place the 2 batteries
with the positive end
facing upwards.
Slide the lid along
the grooves to
close.
• A new battery lasts around 1000 measurements or approximately a year.
• Please obtain the replacement Lithium batteries (DL or CR2032) at a local electrical appliance shop.
• It is recommended to have back-up batteries available.
• Dispose of batteries according to your national regulations.
Keep pressing
part.
• Lancing Device
Installing or replacing the batteries

10
Before use
Setting time and date
1Press the “ ” button.
The indication in the
display (see above)
appears for 1 second.
The date and time
are displayed.
After placing the batteries, make sure you set the date and time.
Beep
Autres manuels pour FineTouch
1
Table des matières
Autres manuels Terumo lecteur de glycémie
Manuels lecteur de glycémie populaires d'autres marques

ReliOn
ReliOn Prime Manuel de l'opérateur

AgaMatrix
AgaMatrix Presto Manuel utilisateur

Polymer Technology Systems
Polymer Technology Systems A1C Now Self Check Manuel utilisateur

Medisana
Medisana GlucoDock Manuel utilisateur

Arkray
Arkray ASSURE PLATINUM Manuel utilisateur

Dexcom
Dexcom G4 Professional Manuel utilisateur

LabStyle
LabStyle Dario Manuel utilisateur

ACONLABS
ACONLABS ON CALL EXPRESS Manuel utilisateur

Higi Technology
Higi Technology iGlucose Manuel utilisateur

Bayer HealthCare
Bayer HealthCare CONTOUR NEXT Manuel utilisateur

Accu-Chek
Accu-Chek ADVANTAGE Manuel utilisateur

Sanofi-aventis
Sanofi-aventis BG STAR Caractéristiques techniques