Swann AMI Oxytrace Manuel utilisateur

AMI Oxytrace
AMI Oxytrace QED
Version 6.20 and higher
A-96.250.531 / 300617
Operator’s Manual

© 2016, SWAN ANALYTISCHE INSTRUMENTE AG, Switzerland, all rights reserved
subject to change without notice.
Customer Support
SWAN and its representatives maintain a fully trained staff of technical specialists
around the world. For any technical question, contact your nearest
SWAN representative, or the manufacturer:
SWAN ANALYTISCHE INSTRUMENTE AG
Studbachstrasse 13
8340 Hinwil
Switzerland
Internet: www.swan.ch
E-mail: [email protected]
Document Status
Title: Monitor AMI Oxytrace QED Operator’s Manual
ID: A-96.250.531
Revision Issue
00 Jan. 2008 First Edition
01 Nov. 2008 Including AMI Oxytrace QED
02 July 2011 Chapter Quality Assurance added
03 Oct. 2013 New mainboard V2.4, update to FW Release 5.4
04 Sept. 2016 New mainboard V2.5, update to FW Release 6.20

AMI Oxytrace QED
A-96.250.531 / 300617 1
Table of Contents
1. Safety Instructions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4
1.1. Warning Notices . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5
1.2. General Safety Regulations . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7
2. Product Description . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8
2.1. Description of the System. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8
2.2. Technical Data . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 13
2.3. Instrument Overview. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 15
2.4. Single Components . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 17
2.4.1 AMI Oxytrace Transmitter. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 17
2.4.2 Sensor OXYTRACE G . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 18
2.4.3 QV-Flow PMMA OTG . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 19
2.4.4 B-Flow SS316L OTG . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 20
3. Installation. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 21
3.1. Installation Check List. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 21
3.2. Mounting of Instrument Panel. . . . . . . . . . . . . . . . . . . . . . . . . . . . 22
3.3. Connecting Sample Inlet and Outlet. . . . . . . . . . . . . . . . . . . . . . . 23
3.3.1 Swagelok Fitting Stainless Steel at Sample Inlet . . . . . . . . . . . . . 23
3.3.2 FEP Tube at Sample Outlet . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 24
3.4 Install the Swansensor Oxytrace G . . . . . . . . . . . . . . . . . . . . . . . 24
3.5. Electrical Connections . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 25
3.5.1 Electrical Connection Diagram AMI Oxytrace . . . . . . . . . . . . . . . 27
3.5.2 Electrical Connection Diagram AMI Oxytrace QED . . . . . . . . . . . 28
3.5.3 Power Supply . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 29
3.6. Input . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 30
3.7. Relay Contacts . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 30
3.7.1 Alarm Relay . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 30
3.7.2 Relay 1 and 2 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 31
3.8. Signal Outputs . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 33
3.8.1 Signal output 1 and 2 (current outputs) . . . . . . . . . . . . . . . . . . . . 33
3.9. Interface Options . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 33
3.9.1 Signal Output 3. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 34
3.9.2 Profibus, Modbus Interface. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 34
3.9.3 HART Interface. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 35
3.9.4 USB Interface . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 35
4. Instrument Setup . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 36
4.1. Establish Sample Flow . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 36
4.2. Programming . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 36

2A-96.250.531 / 300617
AMI Oxytrace QED
5. Operation. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 37
5.1. Keys . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 37
5.2. Display . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 38
5.3. Software Structure . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 39
5.4. Changing Parameters and values. . . . . . . . . . . . . . . . . . . . . . . . . 40
6. Maintenance . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 41
6.1. Maintenance Table . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 41
6.2. Stop of Operation for Maintenance. . . . . . . . . . . . . . . . . . . . . . . . 41
6.3. Maintenance of the Oxygen Sensor . . . . . . . . . . . . . . . . . . . . . . . 42
6.3.1 Electrolyte exchange. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 42
6.3.2 Clean Swansensor Oxytrace G and Flow Cell . . . . . . . . . . . . . . . 44
6.4. Maintenance of the Faraday electrode . . . . . . . . . . . . . . . . . . . . . 45
6.5. Calibration. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 47
6.6. Zero-Verification . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 49
6.7. Faraday Verification . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 50
6.8. Quality Assurance of the Instrument. . . . . . . . . . . . . . . . . . . . . . . 51
6.8.1 Activate SWAN Quality assurance procedure . . . . . . . . . . . . . . . 52
6.8.2 Pre-test . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 53
6.8.3 Connect the sample lines . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 53
6.8.4 Carry out comparison measurement. . . . . . . . . . . . . . . . . . . . . . . 55
6.8.5 Completion of the measurement. . . . . . . . . . . . . . . . . . . . . . . . . . 56
6.9. Replacing Fuses . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 57
6.10. Longer Stop of Operation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 58
7. Error List . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 59
8. Program Overview . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 63
8.1. Messages (Main Menu 1) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 63
8.2. Diagnostics (Main Menu 2) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 64
8.3. Maintenance (Main Menu 3) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 65
8.4. Operation (Main Menu 4) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 66
8.5. Installation (Main Menu 5). . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 67
9. Program List and Explanations. . . . . . . . . . . . . . . . . . . . . . . . . 69
1 Messages. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 69
2 Diagnostics . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 69
3 Maintenance . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 71
4 Operation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 73
5 Installation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 74
10. Material Safety Data sheets . . . . . . . . . . . . . . . . . . . . . . . . . . . . 89
10.1. Reagents. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 89

AMI Oxytrace QED
A-96.250.531 / 300617 3
11. Default Values. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 90
12. Index . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 93
13. Notes . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 94

4A-96.250.531 / 300617
AMI Oxytrace QED
Safety Instructions
AMI Oxytrace QED -
Operator’s Manual
This document describes the main steps for instrument setup, oper-
ation and maintenance.
1. Safety Instructions
General The instructions included in this section explain the potential risks
associated with instrument operation and provide important safety
practices designed to minimize these risks.
If you carefully follow the information contained in this section, you
can protect yourself from hazards and create a safer work environ-
ment.
More safety instructions are given throughout this manual, at the
respective locations where observation is most important.
Strictly follow all safety instructions in this publication.
Target
audience
Operator: Qualified person who uses the equipment
for its intended purpose.
Instrument operation requires thorough knowledge of applications,
instrument functions and software program as well as all applicable
safety rules and regulations.
OM Location The AMI Operator’s Manual shall be kept in proximity of the instru-
ment.
Qualification,
Training
To be qualified for instrument installation and operation, you must:
read and understand the instructions in this manual as well as
the Material Safety Data Sheets.
know the relevant safety rules and regulations.

AMI Oxytrace QED
Safety Instructions
A-96.250.531 / 300617 5
1.1. Warning Notices
The symbols used for safety-related notices have the following sig-
nificance:
DANGER
Your life or physical wellbeing are in serious danger if such
warnings are ignored.
Follow the prevention instructions carefully.
WARNING
Severe injuries or damage to the equipment can occur if such
warnings are ignored.
Follow the prevention instructions carefully.
CAUTION
Damage to the equipment, minor injury, malfunctions or incor-
rect process can be the consequence if such warnings are ig-
nored.
Follow the prevention instructions carefully.
Mandatory
Signs
The importance of the mandatory signs in this manual.
Safety goggles
Safety gloves

6A-96.250.531 / 300617
AMI Oxytrace QED
Safety Instructions
Warning Signs The importance of the warning signs in this manual.
Electrical shock hazard
Corrosive
Harmful to health
Flammable
Warning general
Attention general

AMI Oxytrace QED
Safety Instructions
A-96.250.531 / 300617 7
1.2. General Safety Regulations
Legal
Requirements
The user is responsible for proper system operation.
All precautions must be followed to ensure safe operation
of the instrument.
Spare Parts
and
Disposables
Use only official SWAN spare parts and disposables. If other parts
are used during the normal warranty period, the manufacturer’s
warranty is voided.
Modifications Modifications and instrument upgrades shall only be carried out by
an authorized Service Technician. SWAN will not accept responsi-
bility for any claim resulting from unauthorized modification or alter-
ation.
WARNING
Risk of Electrical Shock
If proper operation is no longer possible, the instrument must be
disconnected from all power lines, and measures must be taken
to prevent inadvertent operation.
To prevent from electrical shock, always make sure that the
ground wire is connected.
Service shall be performed by authorized personnel only.
Whenever electronic service is required, disconnect instru-
ment power and power of devices connected to.
– relay 1,
– relay 2,
– alarm relay
WARNING
For safe instrument installation and operation you must read
and understand the instructions in this manual.
WARNING
Only SWAN trained and authorized personnel shall perform the
tasks described in this document.

8A-96.250.531 / 300617
AMI Oxytrace QED
Product Description
2. Product Description
This manual describes the functions of the following monitors
AMI Oxytrace
AMI Oxytrace QED
Both monitors are mainly identical except that AMI Oxytrace QED
includes a faraday verification.
2.1. Description of the System
Application AMI Oxytrace is used to measure low levels of oxygen in high-puri-
ty water. Especially in power plant water cycles (e.g. feedwater), a
very low level of oxygen is needed to prevent corrosion.
Measuring
principle
Clark principle:
The sensor consists of one noble metal electrode (e.g. platinum or
gold), a reference electrode (mostly Ag/AgCl) and optionally a met-
al guard electrode.
The Clark-type electrode is the most widely used oxygen sensor for
measuring oxygen dissolved in a liquid. The basic principle is that
there is a cathode and an anode submersed in an electrolyte and a
voltage is applied between the two parts. Oxygen enters the sensor
through a permeable membrane by diffusion, and is reduced at the
cathode according to
O2 + 4e- + 2 H2O --> 4 OH-
This reaction creates a measurable current. There is a linear cor-
relation between the oxygen concentration and the electrical cur-
rent.
The guard electrode is on the same voltage level as the cathode
but there is no current measurement. Oxygen which diffuses from
the electrolyte to the cathode is consumed by the guard electrode.
As a consequence, residual oxygen in the electrolyte will no more
disturb the measurement signal and the response time to low oxy-
gen levels will be shorter.
Temperature
compensa-
tion
The measuring signal depends on temperature, but is automatically
compensated to 25 °C. The sample temperature is determined con-
tinuously by a temperature sensor inside the oxygen electrode.
Autres manuels pour AMI Oxytrace
1
Ce manuel convient aux modèles suivants
1
Table des matières
Autres manuels Swann Moniteur

Swann
Swann SW248-LM8 Manuel utilisateur
Swann
Swann AMI Solicon4 Manuel utilisateur

Swann
Swann DP885C Mode d’emploi

Swann
Swann AMI Turbitrack Manuel utilisateur

Swann
Swann SW248-8LM Manuel utilisateur

Swann
Swann SW244-LMP Manuel utilisateur

Swann
Swann SW346-7MC Manuel utilisateur

Swann
Swann SW248-8LM Manuel utilisateur

Swann
Swann SW248-LM8 Manuel utilisateur

Swann
Swann SW248-LM8 Manuel utilisateur