10
NOTE! If there is a fault at the probe connection (1) or the pneumatic unit, a red arrow on
the probe connection (1) lights up.
WARNING! When operating on mains power, check that the mains control lamp (7) is on.
Otherwise, the ICP monitor runs in battery mode and discharges.
Reading the ICP values
The median ICP in mmHg (3) is shown in large digits on the display of the ICP monitor. The systolic
(5) and diastolic (4) ICP can be read in mmHg to the right.
NOTE! The display values of the systolic (5) and diastolic ICP (4) must be used to confirm
the median ICP (3).
NOTE! If the amplitude between the systolic (5) and diastolic ICP (4) is very low or non-
existent, this indicates an error. In this case, check the probe for kinks, constriction,
pinching or a faulty connection to the ICP monitor and check whether the ICP monitor is
switched on.
NOTE! An unstable signal path indicates a leakage in the pneumatic system. In this case,
shut down the ICP monitor and contact the manufacturer.
Connecting to a patient monitor
NOTE! Only cables authorized by Spiegelberg may be used for connecting the ICP
monitor.
1. Connect the monitor cable to the rear connector "Monitor" (11) and the ABP input of the patient
monitor.
2. Switch on the patient monitor and the ICP monitor (if not already done).
3. Press the "Zero" button (2) on the ICP monitor. The message "0.0" flashes on the display (3).
4. Zeroing of the patient monitor within 10 s according to the manufacturer's specifications.
5. Compare the median ICP on the ICP monitor (3) with the display on the patient monitor.
NOTE! The ICP monitor must always be switched on when in use with a patient monitor.
If the patient monitor is connected to an ICP monitor which is switched off, the patient
monitor may display a pressure signal without amplitude. This does not reflect the actual
measured value.
NOTE! During initialisation of the ICP monitor (zeroing the probe), a display of up to 100
mmHg may appear on the patient monitor. This does not reflect the actual measured
value.
NOTE! The screw connections of the monitor cables must be tightened firmly to prevent
disconnection when moving the ICP monitor.
Connecting to a computer
NOTE! Only cables authorized by Spiegelberg may be used for connecting the ICP
monitor.
Connect the computer cable to the "RS 232" (12) connector on the back of the ICP monitor and the RS
232 connector of the computer.
NOTE! The screw connections of the computer cables must be tightened firmly to prevent
disconnection when moving the ICP monitor.