DB-1 –USER MANUAL
1 Safety
The DB-1 demonstration board has been designed for the purpose of simulation of measurements,
whose results determine the safety conditions of a virtual installation. Therefore, in order to guarantee ap-
propriate operation and correctness of the obtained results, the following recommendations must be fol-
lowed:
Before you proceed to operate the device read thoroughly the present manual and observe the safe-
ty regulations and instructions specified by the manufacturer.
The device has been designed for operation at the nominal voltage of 230 V 50 Hz.
The device cannot be used in the grid and with other devices in special interiors, for example those
of a dangerous atmosphere regarding the risk of explosion or fire.
It is unacceptable to operate the following:
a damaged device which is completely or partially out of order
a device with damaged power cable insulation,
a device stored for an excessive period of time in disadvantageous conditions (e.g. excessive
humidity). If the meter has been transferred from a cool to a warm environment of a high level of
relative humidity, do not perform the measurements until the meter has been warmed up to the
ambient temperature (approximately 30 minutes).
a device with damaged housing.
Before measurement of connection resistance or insulation resistance, make sure the measured ob-
ject has been disconnected from the power supply. Disconnect the power cable of the device from
the power supply contact and turn the RCD off.
Repairs may be performed solely by an authorised service point. Before the device is connected to the
power supply network, make sure the power supply socket is equipped with a properly connected earth
pin. The board will not function if it is powered from a socket without an additional PE connection.
The board is equipped with a protection constituted by two F 4A 250 V fuses (T 3,14A 250 V fuse are
acceptable). The fuses are located in the device power supply socket. Should the fuse elements be
damaged, they must be replaced.
It is unacceptable to use other fuses.
WARNING
The DB-1 demonstration board may be used solely by qualified personnel who
have appropriate certificates that entitle them to work on electric installations, or
by persons working under their supervision. Using the board by unauthorised per-
sonnel or without supervision of such personnel may cause damage of the device
and be a source of a serious danger for the user.
Before the fuses are replaced, the device must be disconnected from the 230 V
power supply network.
NOTE!
Before you proceed to operate the device, read thoroughly the present manual and ob-
serve the safety regulations and instructions specified by the manufacturer.
The demonstration board has been designed for the purpose of simulation measure-
ments of alternating voltage, impedance, short circuit, earth resistance, parameters of
the RCD switch, ground resistivity, insulation resistance, resistance of equipotential
bondings. Any application that differs from those specified in the present manual may
cause damage of the device and be a source of a serious danger for the user.
Before and during measurements of ground resistivity, insulation resistance, resistance
of equipotential bondings, earth resistance (by means of the technical methods) discon-
nect the power cable of the device from the power supply socket and turn the RCD
switch off. The power supply of the board is signalled by the „POWER ON” diode.