WARNINGS AND CAUTIONS
• This vaporiser is intended for Veterinary Use Only
• Do not ll the vaporiser with any agent other than the one specied on the front
label. Vaporisers are specically designed dependent on agent type. Therefore, any
other agent used can prove to be dangerous to a patient.
• Do not use this vaporiser until it is mounted upright, vertical and out of operation
for a minimum of (1) hour aer initial lling to allow for proper initial wick
absorption. Failure to allow wicking absorption time may result with inaccurate
output at the selected dial setting.
• Do not carry vaporiser by control dial. Handle and transport vaporiser with care
by grasping the vaporiser rmly with two hands.
• Do not modify, tamper, or disassemble the vaporiser. There is a probable
danger of damaging the vaporiser and altering the calibration accuracy.
• Do not put vaporiser into any liquid, including water.
• Do not attempt to sterilise vaporiser.
• Do not drain anaesthetic agent into any container other than a properly
marked container.
• Do not tilt or tip vaporiser beyond a 45-degree angle while lled with liquid
agent. If unit is accidentally tipped on its side, please call the manufacturer
for specic instructions.
• Do not have vaporiser serviced by anyone other than a SOMNI authorised
service centre.
• Do not operate vaporiser prior to leak testing the anaesthetic equipment,
ensuring secure connections to prevent unnecessary anaesthetic exposure
• Turn the vaporiser OFF when not in use.
• It is recommended that the vaporiser is kept upright at all times aer installation.