SciCan STATIM 7000 Manuel utilisateur

•
Service Manual
STATIM 7000
CASSETTE AUTOCLAVE
P/N xx-xxxxxx Rev. 1.0
STAT
IM
7000
Operator's ma
nual
Copyright ©2008 SciCan Ltd. All rights reserved.

Table of Contents
1. Cassette........................................................................................................................1
2. Cover.............................................................................................................................4
3. Rear Chassis Cover Plate ...........................................................................................7
4. Membrane Keypad and LCD .......................................................................................8
5. Armature.....................................................................................................................10
6. Isoplate .......................................................................................................................12
7. Steam Generator........................................................................................................13
8. Steam Generator Check Valve.................................................................................. 16
9. Steam Generator Thermal Fuse................................................................................18
10. Pressure Relief Valve ..............................................................................................21
11. Chamber Solenoid Valve.........................................................................................22
12. Validation Solenoid Valve .......................................................................................25
13. Thermocouples........................................................................................................27
14. Pump.........................................................................................................................35
15. Auxiliary Heater .......................................................................................................36
16. Adaptor Board.......................................................................................................... 38
17. Controller Board ...................................................................................................... 39
18. Microprocessor........................................................................................................40
19. Auxiliary Heater Board ............................................................................................42
20. Solid State Relay......................................................................................................43
21. Mains Components..................................................................................................44
22. Line Filter..................................................................................................................46
23. Reservoir Inlet Assembly........................................................................................ 48
24. RFID Adaptor Module ..............................................................................................50
25. Compressor..............................................................................................................51
26. Conductivity Calibration .........................................................................................53
27. S Class Calibration ..................................................................................................55
28. Calibration................................................................................................................60
SciCan Ltd.
1440 Don Mills Road
Toronto, Ontario
M3B 3P9
Canada
Local: 416.445.1600
Toll Free: 1.800.667.7733
Fax: 416.445.2727
For all Service and Repair inquiries:
Canada: 1-800-870-7777
USA: 1-800-572-1211
EU: +49-(0)821-56-7456
International: 1-416-445-1600
Email: techservice.ca@scican.com

STATIM 7000 Service Manual - Cassette
1
1. Cassette
1.1 Replacing the Cassette Seal and O-Rings
To ensure optimum performance of the Statim 7000, the seal is equipped with an
electronic tag to ensure only authentic seals are used and also to provide the
user a LCD message to replace the seal after 1000 cycles. The seal and O-rings
should be changed every 1000 cycles or every 12 months.

STATIM 7000 Service Manual - Cassette
2
Replacement seals are available from SciCan (p/n 01-110295S Lid Seal).
When the cassette seal has undergone 1000 cycles, an early message on the
LCD will display SEAL LIFE WARNING - 100 CYCLES REMAINING, with a
warning beep. The seals should be changed immediately. After 1100 cycles, your
unit will stop operating and a message will be displayed SEAL LIFE EXPIRED,
REPLACE. Replacement seals are available from your SciCan dealer.
To change the cassette lid seal, follow these steps:
1. Place the cassette lid and the new seal on a clean work surface.
2. Examine the position of the old seal in the cassette lid and arrange the
new seal in the same orientation, next to the lid.
3. Remove the old seal and discard.
4. Clean any residue out of the seal channel and flush out the channel with
distilled water.
5. Lubricate the new seal with the liquid seal lubricant provided.
6. Align the RFID tag in the new seal with the holes in the lid.
7. Insert the rounded edge of the seal under the round lip of the lid.
NOTE: When inserting the seal, seven round nibs should be visible. The nibs
should fit flush with the lid’s outer surface. Ensure the seal is completely inserted.
Feel around the periphery to ensure the seal is securely seated.
To change the cassette tray seals (O-Rings), follow these steps:
1. Remove the Perforated Rack (01-110294S)
2. Remove the Tray exhaust duct (01-110297S)
3. Remove the Steam Deflector (01-110293S)
4. Unscrew the first coupling (Exhaust Coupling, 01-110292S) and carefully
remove the two O-Ring seals.
5. Find the matching seals from the kit and install them on to the port valve.
6. Secure the first coupling to the cassette ensuring the metal spring is on
the outer wall of the cassette.

STATIM 7000 Service Manual - Cassette
3
7. Unscrew the second coupling (Inlet Coupling, 01-110291S) and carefully
remove the two O-Ring seals.
8. Find the matching seals from the kit and install them on to the port valve.
9. Secure the second coupling to the cassette ensuring the metal spring is
on the outer wall of the cassette.
10.Replace the Tray Exhaust duct, steam deflector, and perforated rack.
NOTE: During a cycle, steam may appear between the lid and the tray. If this
persists, remove the cassette and check that the seal is correctly installed.
Be careful. The metal parts may be hot, and the cassette may contain hot steam.
For the unit to function effectively, the steam deflector and perforated rack MUST
be placed back in the cassette in the proper position.
1.2 Replacing the Cassette Handle
Cracked or otherwise broken Statim 7000 cassette handles can be replaced by
following these steps:
1. Remove the two screws on the underside of each piece of the handle.
2. Remove the handle.
3. Separating the handle from the cassette may require pushing down
forcefully on the handle.
4. Replace and re-attach.

STATIM 7000 Service Manual - Cover
4
2. Cover
1. With the unit off, unplug the power cord from the wall outlet and remove
the cassette and reservoir from the unit.
2. Remove the water filter or water
bypass cartridge (if the unit is using
distilled water) from the reservoir
area.
3. Remove the eight screws across the
bottom front of the unit.
4. Push the cover forward from the
back a little to loosen it.

STATIM 7000 Service Manual - Cover
5
5. Detach the LCD/keypad by reaching
up inside the cover through the
armature opening to feel for a plastic
tab located directly behind the
Rubber and Plastics key.
6. Push this tab to the left to unlock the
LCD/keypad from the cover.
7. Place the LCD/keypad on top of the
armature so that it is out of the way
as you remove the cover.

STATIM 7000 Service Manual - Cover
6
8. Push the cover all the way forward
to slide it off from the front.
9. Reverse removal instructions to replace. NOTE: replacing cover may
change to include info on cover engaging a rail.

STATIM 7000 Service Manual – Cover Plate
7
3. Rear Chassis Cover Plate
Accessing the components behind the armature will require the removal of the
rear chassis cover plate. To remove this plate, follow these steps.
1. Power OFF the unit, unplug it and remove the cover. See Cover. NOTE:
The steam generator may be hot if the unit has been operating.
2. Disconnect the fan connectors.
3. Disconnect the inlet and outlet tubes on the biological air filter.
4. Remove the three screws across
the top and the three screws
across the bottom of the plate.
5. Lift the plate to remove.
6. Reverse instructions to replace.

STATIM 7000 Service Manual – Keypad & LCD
8
4. Membrane Keypad and LCD
The membrane keypad and LCD contain circuitry that is static sensitive.
Always wear a static strap when working with or near this circuitry. Transport
electronic components in a static protected bag.
To replace the membrane keypad or LCD, perform the following steps:
1. With the unit off, unplug the power cord from the wall outlet and remove
the cassette and reservoir from the unit.
2. Remove the water filter or water bypass cartridge (if the unit is using
distilled water) from the reservoir area.
3. Remove the eight screws across the bottom front of the unit.
4. Push the cover forward from the back to loosen it.
5. Detach the LCD by reaching up
inside the cover through the
armature opening to feel for a
plastic tab located directly
behind the Rubber and Plastics
key.
Push this tab to the left to unlock the
LCD from the cover.