ScanSpeed Mini Manuel utilisateur

Date : 19.01.2017
Cat. No. : 7.601.314.101, 7.601.314.102, 7.601.314.103
SERVICE MANUAL
PERSONAL MICRO CENTRIFUGE
ScanSpeed Model Mini

Micro Centrifuge, mini – Service manual
2
CONFIDENTIAL
Table of contents
1 Operating Instruction ...............................................................................................................4
1.1 About this manual..............................................................................................................4
1.2 Safety label and precaution .............................................................................................4
1.2.1 Safety Label .............................................................................................................4
1.2.2 Safety Precautions..................................................................................................5
2 Installation ..................................................................................................................................6
2.1 Unpacking ........................................................................................................................6
2.2 Location ............................................................................................................................6
2.3 Supply the power ...........................................................................................................6
2.4 Power On and opening front door .............................................................................6
3 Device Information....................................................................................................................8
3.1 Special qualities ..............................................................................................................8
3.2 Technical Specifications.................................................................................................8
3.3 Outer Description ...........................................................................................................9
3.4 Operating Function of Control Panel..........................................................................9
3.5 Operating System ........................................................................................................ 10
3.6 Controller Board .......................................................................................................... 11
4 Disassembling ......................................................................................................................... 12
4.1 Rotor and Door assembly ........................................................................................ 12
4.2 Case Top ........................................................................................................................ 12
4.3 Main Controller Board ................................................................................................ 13
4.4 Display Controller Board ............................................................................................ 13
4.5 Motor assembly and Anti-vibration rubber............................................................ 14
4.6 RPM Sensor assembly................................................................................................. 14
4.7 Door lock assembly and Door Sensor ..................................................................... 15
5 Service Mode........................................................................................................................... 16
5.1 Transition into service mode..................................................................................... 16
5.2 Handling values............................................................................................................ 16
6 Error code and Troubleshooting ....................................................................................... 16

Micro Centrifuge, mini – Service manual
3
6.1 Error code...................................................................................................................... 17
6.2 Troubleshooting ........................................................................................................... 18
7 Maintenace .............................................................................................................................. 19
7.1 Cleaning and disinfection........................................................................................... 19
7.2 Device test for centrifuge .......................................................................................... 20
7.2.1 Validation of actual RPM .................................................................................... 20
7.2.2 Validation of Motor performance..................................................................... 20
8 Parts Information ................................................................................................................... 21
8.1 Assembly Drawing....................................................................................................... 21
8.2 Part List.......................................................................................................................... 25

Micro Centrifuge, mini – Service manual
4
1. Operating Instruction
1.1 About this manual
This service manual should be used by specialized engineers authorized by Gyrozen
Co., Ltd.
Any repairing work operated by non-authorized personnel cannot be protected and
guaranteed.
This service manual aims to find possible errors quickly and fix them properly.
Refer to the user’s manual for detailed operation of Centrifuge.
Do not copy or reprint without approval
1.2 Safety label and precaution
1.2.1 Safety Label
The labels attached to the device give information for safety.
Label Information Label Information
Attention label to show
risk and warning
Attention label to warn
electric shock

Micro Centrifuge, mini – Service manual
5
1.2.2 Safety Precautions
•Supply proper voltage power according to device’s power requirement.
•Let all repairing works done by authorized or qualified personnel.
•Use rotors or accessories which are approved by Gyrozen.
•Not try to open the lid and or move the device while the rotor is running.
•Operate the centrifuge with a rotor properly attached and secured to the shaft.
•Not use flammable, hazardous, explosive, or corrosive materials.
NOTE: When it is required to use toxic, radioactive materials or pathogenic micro-
organisms, which belong to the Risk Group II of WHO: “Laboratory Bio-safety Manual,”
should follow the regulation guidelines from WHO.
•Keep away hazardous materials farther than 30 cm (12 in) from the device during
centrifugation, as recommended in IEC standards 61010-2-020.
•Keep the rpm or rcf under its maximum speed in the case that the density of sample
materials is greater than 1.2 g/ml to avoid rotor failure.
•Load samples symmetrically in the rotor diagonally to make balance between the tubes.
•Balance the load on the rotor totally to prevent the damage to the device even by using
several water-filled tubes.
•Place device on a flat, level, rigid and stable surface.
•Disconnect power supply prior to maintenance and service work to avoid electrical shock.
•Use proper disinfection procedures when centrifuging bio hazardous compounds.
When a blackout takes place while the device is running, the door does not open. And the
rotor speed begins to decrease at natural level. Even if the power turns on before the rotor
stops completely, the rotor does not return to the original speed, but decreases more rapidly
with buzzer sound.
The door is closed/opened automatically by a door lock unit operated by a solenoid, and it
will not be opened while the rotor is running at all. Even if the door is opened accidentally,
a door limit switch senses it instantly to make the rotor speed decrease.
In Blackout
M
a
ke sure to
Door opening

Micro Centrifuge, mini – Service manual
6
2. Installation
2.1 Unpacking
1) Check if the box contains such parts as;
① Power cable, 1ea
② User’s manual, 1ea
③ Rotor, 1ea
④ Lid, 1ea
⑤ Rotor coupling nut, 1ea
⑥ Adaptor 0.2ml, 12ea
⑦ Adaptor 0.5ml, 12ea
2) Open the box and lift out the device carefully together with the safety padding.
3) Remove the safety padding and vinyl wrap.
4) Place the device on the flat surface.
2.2 Location
1) Install the device at the solid and flat floor or table. If you place the centrifuge at
the slope, the axis of rotation is possibly changed because of the rotor weight.
2) Install the device keeping a distance of 30cm at least from the wall.
The distance is needed for the air circulation around the device.
3) Install the device at the place with appropriate temperature and humidity.
These conditions have to be maintained constantly as soon as possible.
4) Install the device at the place without any kinds of corrosive gases.
2.3 Supply the power
1) mini model uses 110V or 220V.
Check proper voltage of the device and connect to adequate power outlet.
2) If the power input is more than +/- 10% of the recommended voltage or
fluctuating frequently, it may affect some functions of the device.
In that case it is recommended to use AVR (Automatic Voltage Regulator).
3) If you want to use the device under the other voltage range, please contact us for
safe usage.
2.4 Power On and opening front door
1) Turn on the device by the switch on the back side of the device.
2) The door opens automatically at the time of power on.
3) The door opens automatically also when the test ends.
4) To open the door in other cases, press the ‘Door’ button.

Micro Centrifuge, mini – Service manual
7
5) If it does not work (door not open), use the emergency door open tool at the
bottom of the device(as the figure)
Insert a tool like a driver into the hole and push the latch to the red arrow
direction

Micro Centrifuge, mini – Service manual
8
3. Device Information
3.1 Special qualities
High safety and low noise
Fixed angle rotor
Simultaneous display of rpm and rcf speed
Automatic alarm function for Door open, Speed trouble
High tech AC Induction Motor adopted
3.2 Technical Specifications
Max. RPM 13,500 rpm
Max. RCF(Fixed Angle) 12,300xg
Max. capacity(Fixed Angle) 12 x 1.5/2.0 ㎖, 4 x 8-tube PCR strips
Run time ≤ 30 min
Acceleration time ≤ 12ec
Deceleration time ≤ 16ec
Power supply 220V, 50/60Hz(110V optional)
Power requirement 300 VA
Dimension (WxDxH) mm 208 x 245 x 145
Weight 4.4 Kg( main body only)
CE Certification Yes

Micro Centrifuge, mini – Service manual
9
3.3 Outer Description
3.4 Operating Function of Control Panel
1) Display LCD: to show the data of each function.
2) RPM/RCF button: to set speed value and to switch the RPM/RCF display.
3) Door button: to open the door.
4) Time button: to set test time from 1 min to 30 min.
5) Pulse button: to accelerate to set RPM and decelerate rapidly.
6) Start/Stop button: to start and to stop operation.
7) Up & Down arrow button: to change input data.
1
2
3
4
5
6
7

Micro Centrifuge, mini – Service manual
10
3.5 Operating System
1) DSPIC 30F4011, MICOM; controls all devices.
2) SMPS; distributes inlet power to each part as appropriate form.
3) Inverter; transforms the single phase power to the 3 phase for running the AC
induction motor.
4) IPM FSBS5CH60; controls the AC induction motor.
5) Solenoid; opens and closes the door lid automatically.
6) Temperature sensor; measures the temperature of motor at its surface and issues E3
error if it goes too high(above 110℃).
7) Firmware program; used to update the firmware with notebook and interfacing
connector.
Autres manuels pour Mini
1
Table des matières
Autres manuels ScanSpeed Équipement de laboratoire
Manuels Équipement de laboratoire populaires d'autres marques

Agilent Technologies
Agilent Technologies 5800 ICP-OES Manuel utilisateur

Endress+Hauser
Endress+Hauser Cleanfit CPA875 Manuel utilisateur

NI
NI PXI-5422 Manuel

Collomix
Collomix Aqix Manuel utilisateur

SPEX SamplePrep
SPEX SamplePrep 6875 Freezer/Mill Series Manuel utilisateur

Ocean Insight
Ocean Insight FLAME-NIR+ Manuel utilisateur