DP-40K HIGH VOLTAGE DIFFERENTIAL PROBE
1. FEATURES
The DP-40K differential voltage probe provides a safety means of
measuring floating potentials for all models of oscilloscopes
incomplete safety (MAX. Bandwidth: 75 MHZ).
It converts the high differential voltage (≦ 40 KV peak) into a low
voltage (≦ 4.0 V) with reference to the earth for display on the
oscilloscopes. It is designed specially for high voltage using.
The BNC output is designed to operate on an input with an
impedance of 1 MΩ. It is 2 times of the 50 Ω.
DP-40K can observe more exact measured testing voltage from
DMM. (Oscilloscope accuracy is 3%, and DMM is 10 times).
2. SPECIFICATIONS
(1) Bandwidth:
DC - 75 MHz (-3 dB) for x 500
DC - 20 MHz (-3 dB) for x 5000
(2) Attenuation: x 500, x 5000
(3) Accuracy: ± 2%
(4) Voltage Input Ranges (DC + AC peak to peak)
≦ 40 KV for x 5000, (i.e about 14 KV RMS or ± 20 KV DC)
≦ 4 KV for x 500, (i.e about 1.4 KV RMS or ± 2 KV DC)
(5) Permitted Max Input Voltage
Max differential voltage: 40 KV (DC + AC peak to peak)
Max voltage between each input terminal and ground: 10 KV RMS
(6) Input Impedance:
Differential: 108.8 MΩ // 1 pF
Between terminals and ground: 54.4 MΩ // 2 pF
(7) Output: ≦ ± 4.0 V
(8) Output Impedance: 50 Ω
(9) Rise Time: 4.7 ns for x 500 , 17.5 ns for x 5000
2