8161-N Jan 2020
Case Management and Modifications
The Module disposable and Smart Syringe contain batteries. Battery icons
display status as green (OK), yellow (low- approximately 20 minutes remaining)
or red (depleted). A warning will appear on the screen with a countdown timer
when the disposable has 5 minutes of battery life remaining. If depleted, Display
will automatically require replacement of the disposable or end case.
During a case, if required the Smart Syringe or Module disposable may be
replaced.
To Replace Disposable
1. Select New (Syringe) or New (Module).
2. Confirm Yes to replace or No to cancel. The system will begin scanning for
the new disposable.
3. Physically replace the disposable.
4. Prime the disposable DO NOT PULL BATTERY TAB OF DISPOSABLE
UNTIL PRIMING IS COMPLETED. System will continue scanning until a
connection is made.
5. Pull battery tab of new disposable and confirm wireless connection.
Possible Alarm Messages
Alarm Message User Actions
Threshold % Alarm and Threshold
Reached Alarm No action required. This alarm is for
notification purposes only if warning tones
are enabled.
Low Battery Alarm (Disposables) If warning tones are enabled, alarm will
sound for approximately 2 seconds. User
must replace the disposable, end the case
or allow the battery to continue depleting
while continuing the case. Monitoring
capability will continue until the No Battery
Alarm sounds.
No Battery Alarm (Disposables) User must replace the disposable or
continue case without monitoring capability.
If warning tones are enabled an audible
alert will sound in addition to the warning on
the screen.
Possible Error Messages
Error Message User Actions
“The connected Syringe is not
calibrated properly. Please connect
another device to continue.”
Replace the Smart Syringe.
“The connected DyeVert is not
calibrated properly. Please connect
another device to continue.”
Replace the DyeVert Plus or DyeVert Plus
EZ Module disposable
“No syringe or DyeVert data
received.” End case and replace the Smart Syringe
and DyeVert Plus or DyeVert Plus EZ
Module disposable
“No syringe or pressure module
data received.” End case and replace the Smart Syringe
and Pressure Module disposable
“Lost connection with the Syringe.
Attempting to reconnect.” No action required
“Lost connection with the DyeVert.
Attempting to reconnect.” No action required
“Lost connection with the DyeVert
EZ. Attempting to reconnect.” No action required
“Lost connection with the Pressure
Module. Attempting to reconnect.” No action required
Error Message User Actions
“Bluetooth Fault Detected. Turn
system OFF and ON: 1. OFF:
Press & hold blue power button >
5 secs. 2. ON: Press & release blue
power button. “
Perform a hard shutdown of the Display by
holding the power button down for >5
seconds, then restart the Display by
pressing the power button once. If the
problem persists, contact Osprey Medical
“Software Fault Detected. Turn
system OFF and ON: 1. OFF:
Press & hold blue power button >
5 secs. 2. ON: Press & release blue
power button. “
Perform a hard shutdown of the Display by
holding the power button down for >5
seconds, then restart the Display by
pressing the power button once. If the
problem persists, contact Osprey Medical
“Cancelling Connection Attempt” No action required
“The syringe cannot be connected.
Replace the disposable and
continue the case, or end the
case?”
Replace the Smart Syringe or end case
“The syringe cannot be connected.
Replace the disposable to start the
case, or return to main menu?”
Replace the Smart Syringe and continue
the case or select main menu and cancel
device use
“The DyeVert cannot be connected.
Replace the disposable and
continue the case, or end the
case?”
Replace the DyeVert Plus Module
disposable or end case
“The DyeVert cannot be connected.
Replace the disposable to start the
case, or return to main menu?”
Replace the DyeVert Plus Module and
continue the case or select main menu and
cancel device use
“The DyeVert EZ cannot be
connected. Replace the disposable
and continue the case, or end the
case?”
Replace the DyeVert Plus EZ Module
disposable or end case
“The DyeVert EZ cannot be
connected. Replace the disposable
to start the case, or return to main
menu?”
Replace the DyeVert Plus EZ Module and
continue the case or select main menu and
cancel device use
“The Pressure Module cannot be
connected. Replace the disposable
and continue the case, or end the
case?”
Replace the Pressure Module disposable or
end case
“The Pressure Module cannot be
connected. Replace the disposable
to start the case, or return to main
menu?”
Replace the DyeTect Pressure Module and
continue the case or select main menu and
cancel device use
“There is no remaining battery
power for the syringe.” Replace the Smart Syringe
“Syringe Battery Level Critical
Replace the Syringe within <timer>“ Replace the Smart Syringe within time
remaining on displayed countdown timer.
“There is no remaining battery
power for the DyeVert.” Replace the DyeVert Plus Module
disposable.
“There is no remaining battery
power for the DyeVert EZ.” Replace the DyeVert Plus EZ Module
disposable.
“DyeVert Battery Level Critical
Replace the DyeVert within
<timer>“
Replace the DyeVert Plus or DyeVert Plus
EZ Module disposable within time
remaining on displayed countdown timer.
There is no remaining battery
power for the Pressure Module.” Replace the DyeTect Pressure Module
disposable.
“Pressure Module Battery Level
Critical. Replace the Pressure
Module within <timer>“
Replace the DyeTect Pressure Module
disposable within time remaining on
displayed countdown timer.
“Would you like to continue the
current case started on <date> at
<time>?”
Select “Yes” to restart the case, “No” to end
the case
“Unable to connect to the Syringe.
Please replace the unit.” Replace the Smart Syringe
“Unable to connect to the DyeVert.
Please replace the unit.” Replace the DyeVert Plus or DyeVert Plus
EZ Module disposable
“Unable to connect to the Pressure
Module. Please replace the unit.” Replace the DyeTect Pressure Module
disposable.
“Please wait as we reconnect to
previously used devices.” No action required
“Please wait as we dispose of
previously used devices.” No action required
“Database Error: failed to initialize
database.” Perform a hard shutdown of the Display by
holding the power button down for >5
seconds, then restart the Display by
pressing the power button once. If the
problem persists, contact Osprey Medical
“Database Error: failed to add
case.” Contact Osprey Medical
“Database Error: failed to update
case.” Contact Osprey Medical
“Database Error: failed to mark all
unfinished cases as abandoned.” Contact Osprey Medical
DyeVert Plus or DyeVert Plus EZ Contrast Reduction System and Contrast
Monitoring System FCC Information
Display Contains FCC ID: Z64-WL18SBMOD
DyeVert Plus and DyeVert Plus EZ Module FCC ID: 2AHUPDV
Device Status & Replacement