Nsense probe dimmensions
2.Configuration Already completed by Novatron prior to delivery
With the next 4 steps you assign the input sensors to output channels
Step 1: Connect nSens to Quantadat and start detection
Step 2: Assign nSens to measuring points A to D
Step 3: Assign measuring points and parameters to channels 1 to 4
Channel 1 equals analog output 1 and display line 1
Step 4: Configure analog signals according your specification
analog output CH1 linear scale 4-20mA, -20 to +80 °C
analog output CH 2 linear scale 4-20mA, 0 to 100% rh ( supplied already completed)
Step 5: If the parameters need to be changed from standard configuration in Step 4 above
then please follow these instructions from the three front menu keys UP, DOWN, ENTER:-
Set up of g/ Kg ( mixing ratio) Scaling range of output in g/Kg (mixing ratio)
ENTER ENTER
‘Measurement’ DOWN
ENTER ‘Analogue Outs’
DOWN ENTER
‘Channel 1’ DOWN
ENTER ‘Analogue Out 1’
‘Measurement Point A’ ENTER
DOWN DOWN
‘Mixing Ratio’ ‘Range L 0.00 g/ Kg’
DOWN DOWN
EXIT ‘Range H 100.00 g/Kg ( change as required)
EXIT
Note: If you get stuck in the menu just down-power momentarily to return to
standby mode. On first power up always press ENTER.
The following steps 1 –4 are only needed if you need to configure the
Quantadat differently from the original specified settings shown in the previous
section 2 ( direct replacement for old TR200 Triplex system) or if the
configuration, carried out by Novatron prior to delivery, has been altered in
error . Otherwise, please ignore the following steps, you are now ready to skip
to section 3 Calibration.