Nice NICE1 Manuel utilisateur

NICE1 Cold + Compression
Therapy System


Thank you for choosing the NICE1
Cold + Compression Therapy
System. The NICE1 uses advanced
technology to greatly improve the
convenience and ecacy of cold +
compression therapy.
This manual has all the information
you need to properly and eectively
operate the NICE1.
Consult a healthcare professional before using NICE1.

QUICK START GUIDE
• Connect the power cord to the
external power supply.
• Plug the circular connector into
the back of the Cooling Unit. You
will hear a click when properly
connected.
• Plug the power cord into the wall
outlet.
Consult a healthcare professional before using NICE1.
It is recommended to read entire user manual before beginning. Refer to the
individual wrap guides for specific information regarding wrap use.
• Connect the hose to the Cooling Unit.
You will hear a click when properly
connected.
• Place the wrap on the appropriate
body part. Secure with hook
and loop.
• Connect the hose to the wrap.
You will hear a click when properly
connected.

Settings for temperature, compression, and duration should be used as directed
by your physician or medical professional.
STEP 1
Turn on the power switch on the back of the
device. The touch screen will illuminate and
display the “Nice” logo. Push “NEXT” . Review
the checklist on the touch screen. Confirm and
press “NEXT”.
STEP 2
Select desired COLD LEVEL and press NEXT.
Cold Settings are as follows:
Level 1 = 58F (14.4C)
Level 2 = 54F (12.4C)
Level 3 = 50F (10C)
Level 4 = 46F (7.7C)
Level 5 = 42F (5.5C)
STEP 3
Select desired COMPRESSION LEVEL and
COMPRESSION TYPE and press NEXT.
STEP 4
Choose a MANUAL time setting of 5 – 40 min-
utes for a single therapy session OR choose a
PRESET PROGRAM for multiple sessions and
press NEXT.
STEP 5
Review and confirm your settings by pressing
START or press BACK to change the settings.
STEP 6
During your therapy session you can change
the cold or compression settings simply by
pressing 1 – 5 or HIGH, MED
or LOW. To Dim, Press Dim switch, to brighten
- touch screen.

Programmable intermittent and
constant compression greatly improves
the eectiveness of cold therapy and
promotes healing.
COMPRESSION
By combining the therapeutic benefits of
cooling and compression in a single easy-
to-use device, recovery times are greatly
improved.
RECOVERY
State-of-the-art technology delivers
therapeutic cooling without ice.
The most convenient way to apply
cold therapy.
COOLING

1. Introduction
2. NICE1 description
3. Indications for use
4. Contraindications, cautions, and warnings
5. Unpacking NICE1
6. Environmental conditions
7. How to set up your NICE1 for therapy
8. Operating instructions
9. Maintenance and cleaning
10. Draining the water from the unit
11. Storage and re-packing the unit
12. Troubleshooting guide
13. Service and customer support
14. Warranty and disclaimer information
15. Electromagnetic compatibility
8
8
10
10
12
13
13
15
18
18
20
21
23
24
25
TABLE OF CONTENTS

8NICE RECOVERY SYSTEMS
1 INTRODUCTION
NICE1 is designed to provide cold and compression therapy as specified in
this manual. If the system is used in a manner other than as specified, its
operation or the safety protection may be impaired.
Please read the entire manual carefully before operating the NICE1. The
patient is the intended operator. The patient can safely use and maintain
the NICE1 following these instructions.
CAUTION: Federal law restricts this device to sale by or on the order of a physician.
2NICE1 DESCRIPTION
NICE1 is a cold and compression therapy device used to aid recovery and
reduce pain associated with soft tissue injuries. The device works by circulating
cooled water and air through a therapy wrap that is placed on the body part.
The cooled water circulates through the therapy wrap and provides cold
therapy; the air inflates the therapy wrap causing it to compress around the
injured body part.
2.1 FEATURES
• Fluid therapy temperature range between 42°F - 58°F (14.4°C - 5.5°C)
• Treatment for edema with alternating compressions of
Low (20 mmHg), Medium (35 mmHg) and High (50 mmHg)
• Programmable therapies
• Lightweight and portable package
• User-friendly interface
• Easy to use and read touch screen display
2.2 GENERAL SPECIFICATIONS
• Weight: 10lbs when full of water
• Hose Length: 6 ft.
• Specified Medical Grade Power Supply (supplied with the system)
• Dimensions: 8.0”W x 7.64”H x 8.0”D
• Operating Fluid: Distilled water, tap water or
glycol/alcohol solution
• Flow Rate: .5 liters per minute
• Water Pressure: <30 PSI
• Water Reservoir Capacity: 300 mL
• Safety: IEC 60601-1:2012 and IEC 60601-1-2:2007
• Air Pump Pressure: 0-50 mm Hg
• Intermittent Compression Setting:
o High: inflate to 50 mmHg for 2 minutes and then deflate
to 5 mmHg for 30 seconds.
o Medium: inflate to 35 mmHg for 2 minutes and then deflate
to 5 mmHg for 30 seconds.
o Low: the NICE1 inflate to 20 mmHg for 2 minutes and then
deflate to 5 mmHg for 30 seconds.
• Cooling Temperature Range:
Level 1 = 58F (14.4C)

9
NICERECOVERY.COM
Level 2 = 54F (12.4C)
Level 3 = 50F (10C)
Level 4 = 46F (7.7C)
Level 5 = 42F (5.5C)
• Use Operating Temperature: 60°F - 80°F (15.5°C - 26.7°C)
• Relative Humidity : 10% - 95% non-condensing
• Storage/Transportation Temperature Range: 40°F - 105°F (4°C - 41°C)
• Maximum Sound Pressure Level: 75dBA (with air pump running)
• Maximum Altitude: 4000m (13,123 ft)
• Display accuracy: +/- 2%
2.3 MEDICAL SYSTEM SPECIFICATIONS
• Specified Medical Grade Power Supply. The following
Power Supply is specified for use with the NICE1 unit:
o Class II Power Supply: DC output rated 15Vdc 12A
NOTE: A detachable type power supply cord (rated 125V 10A) is
provided with each power supply, suitable for use in the USA and Canada.
• NICE1 Cooling Unit: Electrical Input Rating of 15Vdc 12A
• Patient Therapy Wrap (patient applied part): Classification
of Patient Applied Part, Type BF
• Insulated Air/Water Hose Assembly for connection between
Control Unit and Patient Therapy Wrap
• Leakage Currents (per IEC 60601-1):
o Touch (Enclosure) Current:
<28uA in Normal Condition
<50uA in Single Fault Condition
o Patient (Applied Part) Leakage Current:
<28uA in Normal Condition
<50uA in Single Fault Condition
2.4 OPTIONS
• Non-Sterile Patient Therapy Wraps (knee wrap, etc.)
• Carrying Case
Caution: Consult accompanying documents Lot or Batch Number
Refer to User Manual and Labels.
Follow instructions for use.
Serial Number
Caution: Federal law restricts this device to sale
by or on the order of a physician.
OFF position for the DC Power Switch
at the rear of the Control Unit
Type BF Patient Applied Part (therapy wrap) ON position for the DC Power Switch
at the rear of the Control Unit
SN
LOT
RxOnly
!

10 NICE RECOVERY SYSTEMS
3INDICATIONS FOR USE
NICE1 combines cold and compression therapies. It is intended to treat
post-surgical and acute injuries to reduce edema, swelling, and pain
where cold and compression are indicated. It is intended to be used by or
on the order of a healthcare professional in hospitals, outpatient clinics,
athletic training settings, or home settings.
4CONTRAINDICATIONS, CAUTIONS, AND WARNINGS
4.2 CONTRAINDICATIONS FOR COLD THERAPY
NICE1 or any cold therapy device should not be used
by patients that:
• have significant vascular impairment in the aected region
(e.g., from prior frostbite, diabetes, arteriosclerosis or ischemia).
• have acute paroxysmal cold hemoglobinuria or cryoglobulinemia.
• have Raynaud’s disease or cold hypersensitivity (cold urticarial).
4.3 PRECAUTIONS
Basic safety precautions should always be followed when using
NICE1 to reduce the risk of fire, electric shock, and personal injury.
Please read the entire User Manual carefully before operating
the unit.
4.4 CAUTION
• Never push objects of any kind into the therapy unit through
the frame.
• Never spill liquid of any kind on the therapy unit.
• Do not overfill the water reservoir.
• If the unit gets wet, unplug the unit from the wall and allow
the unit to dry before use.
• Only use the supplied power supply and power cord.
• Do not operate the unit if it has any noticeable or physical
damage or is leaking fluid.
• Do not operate the unit with a damaged or frayed power cord.
• The therapy unit is not intended to be used in a wet environment
or when relative humidity is greater than 60%.
• Do not spray the unit with any water solvents or cleaners.
• Do not drop the therapy unit or cause impact to the unit.
• Do not pull or otherwise put undue stress on the hoses.
• Do not use near equipment that generates electromagnetic
or other interference, as this may be harmful to the therapy unit.
• Do not smoke while using therapy wraps or use wraps by
an open flame.
• Do not stick a finger or any other foreign objects into
the reservoir.
• Do not stick a finger or any other foreign objects into the fan.
• Do not drink or ingest the coolant or water that has been
circulating in the system.
• Do not service or perform maintenance while the equipment
is in use.