Instruction Manual
more to myerson than teeth
www.myersontooth.com
to proof
6.17.2010

Installation of the Heat Guard 2
Section 1 Components 3
Section 2 Storage & Handling 4
Section 3 Basic Operation 4
Section 4 Editing User Profiles
Section 5 Processing using the “Run Stored Profile” 5
option (Recommended Option)
Section 6 Processing using the “Rerun Last Profile”
Section 7 Myerson DuraFlex Partial Injection Procedure 7
Section 8 Myerson DurAcetal Framework Injection Procedure 14
Section 9 Warranty & Returns 19
Section 10 Technical Data 19
Section 11 Preventative Maintenance & Cleaning 20
Section 12 Service & Repairs 20
Section 13 Declaration of Conformity 19
Section 14 Important Safety Warnings 21
Section 15 Notes 22
Installation of the HEAT GUARD
To prevent damage, the myerson FlexPress™ has been shipped
with the heat guard not installed. To install the heat guard, place
the heat guard over the corner posts with the warning label UP.
Install the (4) black acorn nuts nger tight. The acorn nuts are
simply holding the heat guard down, so they do not need to be
wrenched tight.
The heat guard can be removed periodically for cleaning.
Just follow the above steps in reverse order to remove the
heat guard.
Thank you for your purchase. Feel free to call myerson with
any questions.
2

myerson Flexpress™ x x
(1) Injection Flask BS-0104
x x
(1) 6 Pack myerson DuraFlex x x
Medium Cartridges
(1) 6 Pack myerson DurAcetal x x
Medium Cartridges
(1) 6 Pack myerson VisiClear x x
Medium Cartridges VC-MD-6PK
Power Cord x x
Air hose x x
(1) Thermoplastic Model Separator MS-0200 x x
TMS Swabs (10) MS-0202 x x
TMS Trays (10) MS-0201 x x
Release Compound BS-0105 x x
Polishing Kit FN-0200 x
Bur Finishing Kit AP-0224 x
(1) Stone Removal Tool SC-0100 x
(1) 25lb. Universal Stone x
UNISTONE-25LB
Description Professional Kit Advance Kit
120v - FP-0101 120v - FP-0100
230v - FP-0103 230v - FP-0102
section 1~components
The following is a list of components that are included
in the myerson Flexpress™ Digital Injection Systems:
3

section 2 ~ storage & handling
The myerson Flexpress™ Digital Injection System should be stored in a
cool dry place. The unit should not be used in an environment exceeding
40°C (104°F). Keep all combustibles away from unit while in use. Never
place anything on the top of the unit.
section 3 ~ basic operation
Place the myerson Flexpress™ Digital Injection System on a clean dry
surface. Insert the power cord into the receptacle on the side of the unit
and plug the other end into a grounded 120V or 230v outlet.
To turn on the unit, press the rocker switch located on the left side of the
unit above the power cord receptacle. The control panel will illuminate
briey then turn back off. The myerson Flexpress™ is now in Standby
mode. To turn on the controller press the ON/STBY button (lower left
button). Below is what you should see in the blue LCD display.
From the Main Menu you can select any of 5 options by using the UP &
DOWN arrow keys. Below is the list of options & a description for each:
1. Run Stored Profile- myerson,LLC pre-programs the myerson
Flexpress™ with processing parameters for myerson DurAcetal®,
myerson DuraFlex and myerson VisiClear™. By selecting the pre-
programmed material, the Melt Temp, Melt Time and Hold Time are
already set for you. This is the recommended option to use when
processing cases using the myerson Flexpress™. There is also (3)
“Spare” locations for the customer to save their own settings.
2. Rerun Last Profile- By selecting this option the myerson
Flexpress™ will run the last set prole. This option is great when
processing several cases of the same material.
3. Edit User Profiles- This option is where you can create your own
processing proles and save them for future use. The myerson
FlexPress is capable of storing up to 7 unique processing proles.
4. Edit User Settings- This option allows you to change settings
such as temperature units (C/°F), turn off/on alarms, languages, etc.
6. Configure from PC- This option is used by repair personnel in
the event of a warranty repair of the myerson FlexPress.
4

5
section 4 ~ editing user profiles
The myerson FlexPress allows the user to save up to 7 custom
processing proles for easy switching between materials. The following
procedure outlines how to create and save proles in the myerson
FlexPress™ Digital Injection System. Note: myerson, LLC does not
provide process settings for materials other than myerson, LLC
supplied materials. Refer to the manufacturer supplied information
for processing settings.
1. From the main menu, use the UP/DOWN buttons until
“Edit User Profile” is displayed. Press the ENTER button
to select.
2. Using the UP/DOWN buttons until the “Spare” location to be
edited is displayed. Press the ENTER button to select.
3. The rst parameter to be modied is the “Name” of the spare
prole. Press the ENTER button and a ashing cursor should
appear. Using the LEFT/RIGHT buttons to move the cursor to
the character location you wish to modify. Once the cursor is
in position, use the UP/DOWN buttons to scroll through the
available characters. Press the ENTER button when nished.
4. Press the DOWN button and the “Melt Temperature” parameter
will be displayed. Press the ENTER button and a ashing cursor
will appear. Use the UP/DOWN buttons to change the Melt
temperature to the desired value. Once nished press the
ENTER button.
5. Repeat step 4 for the “Melt Time” and “Hold Time”parameters.
6. Once satised with the changes, press the LEFT button. The
screen should display “Saving Changes”. The prole you just
created will now appear in the “Run Stored Profile” option in
the main menu.

6
section 5 ~
Processing using the
“Run Stored Profile”
(Recommended Option)
1. From the main menu use the UP or Down buttons until the
“Run Stored Prole” is displayed then press the enter button.
2. Next use the UP/DOWN buttons until prole you wish to run
s displayed.
3. After selecting the prole the display will show the prole name.
Pressing ENTER will scroll between prole name, melt tempera-
ture, melt time and hold time. It is recommended to check the
setting before starting a prole.
4. Press the START button and the machine will begin running the
prole. If the machine is at room temperature, it will take approxi-
mately 15 minutes to reach the processing temperature.
5. Prepare the cartridge by scoring
a “+” using a utility knife.
See the illustration to the right. This will
create an even “Burst” of the cartridge
6. Next, coat the sides of the cartridge
with a light coat of release compound.
This will prevent the cartridge from sticking
to the walls of the heating chamber.
7. When the myerson Flexpress has calibrated itself at the Melt
Temp an alarm will sound and the upper display will read “LOAD”
signaling you to load the cartridge. At this time insert the cartridge
into the heating chamber, center the ask and tighten using the
ask strap and knobs. If you are not ready to start the prole at
this time and want to turn off the alarm, press the UP button.
This will silence the alarm, but the myerson Flexpress™ will
remain calibrated until you press START or put the unit in standby
by pressing the STBY button.
8. Press the START button to begin the prole. Now let the myerson
Flexpress™ do the rest. The unit will begin counting down the
Melt Time and once it expires the unit will automatically inject the
material into the ask. The piston will stay in the up position for
the duration of the Hold time then release.
9. The upper display will read “DONE”.
10. You can now remove the ask strap and knobs.
11. Grasp the ask using heat resistant gloves and press the
EXTEND/RETRACT button. This will extend the piston and
push out the remaining cartridge attached to the ask.

section 7 ~ myerson DuraFlex partial
injection procedure
The following procedure outlines the steps necessary to create
myerson DuraFlex Partials with the myerson Flexpress™ Digital
Injection System.
1. Evaluate the model for major undercuts and block out the
undercuts. Due to the exibility and adjustability of the material,
very little if any blocking out will be needed.
2. Paint the model with myerson Thermoplastic Model Separator
MS-0200.
7
section 6 ~
processing using the
“Re-run Last Profile”
1. From the main menu use the UP or DOWN buttons until the cursor
is to the left of “Rerun Last Prole” then press the enter button.
2. The display will show the prole name and prole settings of the
last injection. If this is not the prole you wish to run, press the
LEFT arrow button to return to the main menu.
3. The myerson Flexpress™ is now ready to process your case.
See steps 5-13 of Section 5 for a complete injection.
12. Press the EXTEND/RETRACT button again to retract the piston.
13. The myerson Flexpress™ is now ready to process another case.
Repeat steps 1-9 for your next injection.

3. Wax and set up the teeth just as you would for an acrylic partial.
4. Add the clasps to the partial. myerson,LLC recommends using
12 gauge sprue wax for creating clasps.
8

5. Coat the inside of the ask with a petroleum jelly. This will prevent
the stone from adhering to the ask.
6. Invest the partial using regular Buff stone. When stone sets, place
sprues using 3/16” (4.75mm) utility rope wax. Extend the sprue to
the injection port in the ask.
7. Coat the stone with stone separator and bolt the top half of the
ask in place. The utility wax should ll the injection port of the
ask. This will keep the stone from leaking out during the second
half of the investing process.
9

8. Pour up the ask using myerson, LLC’s Investing Stone
INVSTONE-25LB. This stone has been designed to prevent
the teeth from shifting during injection. Notice the stone is
not protruding from the widow in the ask.
9. When the last half of the stone is set, remove the bolts but leave
the ask together and place in boiling water.
10. Once the wax has been removed, remove and place holes in
the denture teeth. The holes in the teeth are very important.
They create the mechanical retention necessary to lock the teeth
in place as well as allow air to escape as the myerson DuraFlex
is injected. Once the holes have been created, the teeth need to
be placed back into the proper locations. A small dot of Super
Glue on the incisal area of the teeth will hold them in place as the
material is injected. Bolt the two halves together.
10
Table des matières
Autres manuels MYERSON Équipement de laboratoire
Manuels Équipement de laboratoire populaires d'autres marques

Agilent Technologies
Agilent Technologies 5800 ICP-OES Manuel utilisateur

Endress+Hauser
Endress+Hauser Cleanfit CPA875 Manuel utilisateur

NI
NI PXI-5422 Manuel

Collomix
Collomix Aqix Manuel utilisateur

SPEX SamplePrep
SPEX SamplePrep 6875 Freezer/Mill Series Manuel utilisateur

Ocean Insight
Ocean Insight FLAME-NIR+ Manuel utilisateur