3. SPECIFICATIONS
Function: Insulation resistance, AC voltage
Display: 3 1/2 digit LCD with bar graph, max. reading of 3200 count and annunciators
Response time: Less than 5 sec. (auto ranging)
Data hold indication: “DH” mark on LCD readout
Infinity indication: “OL (∞)” mark on LCD readout (over 3200 count)
Safety standard: IEC 61010-1, CAT. II 600V phase to earth
EMC standard: EN 61326
Constructional standard: IEC 61557-2
Insulation resistance: DC 500V-50MΩor more (MIS-2D)
DC 1000V-50MΩor more (MIS-3D, MIS-4D)
Withstanding voltage:AC 3700V/1minute, between input terminal and outer case
Low battery indication: “B” mark on LCD readout
Temperature characteristics (0~40℃): ±5%rdg of specified accuracy
Operation temperature: 0℃to 40℃, less than 80% RH without condensation
Storage temperature: -10℃to 60℃, less than 80% RH without condensation
Power supply: 1.5V alkaline batteries (AA size, LR-6) x 6
Dimension: 170(W) x 105(H) x 54(D)mm, approx. 365g (excluding batteries)
Accessories: Line test lead .......... 1
Earth test lead ........ 1 Line test lead
Batteries ................. 6
Test leads case ...... 1 Earth test lead
Belt ......................... 1
Instruction manual .. 1
Optional accessory: Test lead with remote switch
4. MEASURING RANGES AND TECHNICAL DATA
●Insulation resistance measurement