IM-7800 Rev G 2021-05 4 of 16
WARNING: Portable RF communications equipment should be used no closer than 30 cm (12 inches) away
from any part of the REF 7800 Sternum Saw. Otherwise degradation of the performance of the
REF 7800 could result.
WARNING: Use of Battery Packs other than those specied in the List of Compatible Accessories could result in
increased electromagnetic emissions or decreased electromagnetic immunity of this equipment and
result in improper operation.
WARNING: REF 7800 Sternum Saw has been tested for use in the vicinity of High Frequency Electrosurgical /
Electro coagulation equipment. Use of such equipment in close proximity to the REF 7800 Sternum
Saw may cause unintended interruptions or unintended motion of durations up to 1 second. Place
the Instrument on a safe surface when not in use. Do not place powered Instruments on the patient
when not in use. Avoid entwining HF surgical cables with powered instruments.
WARNING: The following items should be periodically inspected for signs of damage and repaired or replaced as
needed to ensure continued safety with regard to electromagnetic disturbances over the life of the
system:
• Check that the Battery Connection base, the rear end cap and the front of the instrument
remain securely connected to the main instrument body.
• Check for damage to the battery pack housings.
CAUTION: Do not run the handpiece without the saw blade.
CAUTION: Disconnect battery from handpiece before loading or unloading the saw blade.
CAUTION: Leaving a battery pack attached to a handpiece when not in use may accelerate battery drain.
CAUTION: Make sure the tip of the sternum saw blade is inside the foot of the sternum guard and does not
strike any part of the sternum guard. Improper alignment of the sternum guard with the saw blade
can cause damage to the sternum saw or saw blade.
CAUTION: If the handpiece will not hold the saw blade or sternum guard securely, discontinue use and return to
the factory for service.
CAUTION: Check Safety Lock and Trigger before operating the handpiece. If the trigger should stick in the
depressed “ON” position, do not use the handpiece. Immediately remove the battery. Return the
instrument to the factory for service. Always use the Safety Lock when the instrument is not running.
CAUTION: Check that the battery is fully seated and locked.
CAUTION: Federal Law (U.S.A.) restricts this device to sale by or on the order of a physician (or properly licensed
practitioner).
CAUTION: Disconnect sternum guard, battery, and saw blade prior to sterilization. Batteries must be sterilized
separate from the handpiece and sternum guard.
CAUTION: DO NOT run instruments while warm. Cool by exposure to room temperature.
CAUTION: DO NOT immerse in liquid to cool.
CAUTION: Repairs or alterations to MicroAire products made by anyone other than MicroAire or an Authorized
MicroAire Repair Facility will void that product’s warranty, and the customer will be responsible for
any costs related to returning the product to working condition
NOTE: Disconnect sternum guard, battery, and saw blade prior to sterilization. Batteries must be
sterilized separately.
NOTE: The REF 7800 Sternum Saw is suitable for use in hospitals and surgery centers. These systems should
not be used near Magnetic Resonance Imaging equipment.
NOTE: The emissions characteristics of this product make it suitable for use in industrial areas and hospitals
(CISPR 11 Class A). If it is used in a residential environment (for which CISPR 11 Class B is normally
required) this equipment might not offer adequate protection to radio frequency communication
services. The user might be required to take mitigation measures, such as relocating or re-orienting
the equipment.
NOTE: Mailing address is located on back cover.