Digital Scale Technical Manual
Load1 Detection
shows.
Load standard weight, press . shows.
Press to change digit. Long press to right scroll digit.
Long press to input decimal point. Input weight value.
Keep load stable. Press to display weight code .
Wait until weight code is stable. Press to start weight detection.
Scale automatically enters Load2 Detection.
Load2 Detection
shows.
If one weight calibration is enough, press to exit Calibration.
Load standard weight, press . shows.
Press to change digit. Long press to right scroll digit.
Long press to input decimal point. Input weight value.
Keep load stable. Press to display weight code .
Wait until weight code is stable. Press to start weight detection.
Scale automatically enters Load3 Detection.
Load3 Detection
shows.
If two weights calibration is enough, press to exit Calibration.
Load standard weight, press . shows.
Press to change digit. Long press to right scroll digit.
Long press to input decimal point. Input weight value.
Keep load stable. Press to display weight code .
Wait until weight code is stable. Press to start weight detection.
Scale automatically exits Calibration.