KYORITSU 8113 Manuel utilisateur

INSTRUCTIONMANUAL
MODEL8113
KEWSNAP SERIES
KYORITSUELECTRICALINSTRUMENTSWORKS,LTD.,
TOKYO,JAPAN
AC/DCCLAMPADAPTOR

Page
1.Features................................................................................. 1
2.Specifications......................................................................... 2
3.SafetyNotes.......................................................................... 4
4.InstrumentLayout.................................................................. 6
5.MeasuringMethod................................................................. 7
5-1.Preparation..................................................................... 7
5-2.Measurements................................................................ 7
5-2-1.DCCurrentMeasurements.......................................... 7
5-2-2.ACCurrentMeasurements......................................... 9
5-2-3.OtherApplications......................................................10
6.BatteryReplacement............................................................ 10
7.OptionalAccessories............................................................ 11
7-1.Model8021KEWENERGIZER.................................... 11
7-2.Model8008MULTI-TRAN.............................................12
Contents

●Model8113isaclamp-onAC/DCcurrent/voltageconverter
thatpermitsAC/DCcurrentmeasurementsfrom1mAto
200Ainconjunctionwithavoltmeterhaving0.1mV
resolution.
●Ithas5A,50Aand200AAC/DCmeasuringranges.Output
voltageis500mVfullscalefor5/50Arangesand200mVfor
200Arange.Forcurrentmeasurements,selecta200mV
multimeter(3-1/2digitdigitalreadout)rangeand1mAwillbe
indicatedas0.1mVand199.9Aas199.9mVfortheclamp
adaptor5Aand200Arangesrespectively.Therefore,Model
8113allowsyoutomakeawiderangeofcurrent
measurementsfromverylowcurrentequipmenttohigh
currentswitchbox.
●InconjunctionwithatrueRMSmultimeter,itisalsopossible
tomeasurearectifiedcurrentorevenAConDCcurrents.
●DCcoupled,inputwaveformappearsattheoutputjack
withoutdistortion.Thiswillpermitwaveformobservationon
theoscilloscope.
●AwidefrequencyrangeofDCto1kHz.
●Operatesontwo1.5VSUM-3batteriesorequivalentonly.
●HousingcaseismadeofULlistedfireretardantplastic
material.
―1―
1.Features

―2―
2.Specifications
Accuracy:
OutputVoltage:
(Aboveaccuraciesarespecifiedfor+ 18℃ to+ 23℃ at< 80%
relativehumidity)
Operating
Temperature
&Humidity: −10℃−+50℃ at<85%relativehumidity
Storage
Temperature
&Humidity: −20℃−+ 60℃ at<75%relativehumidity
Load
Resistance: >1kΩ
PowerSupply: 2×1.5VSUM-3batteryorequivalent
DCCurrent
ACCurrent
MeasuringRanges
0− 5ADC
0− 50ADC
0−200ADC
0− 5AAC
0− 50AAC
0−200AAC
Output
0−500.0mVDC(100mV/A)
0−500.0mVDC(10mV/A)
0−200.0mVDC(1mV/A)
0−500.0mVAC(100mV/A)
0−500.0mVAC(10mV/A)
0−200.0mVAC(1mV/A)
DCCurrent
ACCurrent
MeasuringRanges
0− 5ADC
0− 50ADC
0−150ADC
150−200ADC
0− 5AAC
0− 50AAC
0−150AAC
150−200ADC
Accuracy
±(1.5%rdg+5mA)
±(1.0%rdg+0.02A)
±(1.5%rdg+0.2A)
±3.0%rdg
±(2.0%rdg+5mA)DC-1kHz
±(1.5%rdg+0.04A)DC-1kHz
±(2.0%rdg+0.5A)DC-1kHz
±3.5%rdgDC-1kHz

―3―
LowBattery
Warning: Powerindicationlampflashesatalonger
intervalwhenbatteriesbecomeexhausted
andfinallystopsflashingatapprox.2.2V.
Current
Consumption: Approx.10mA
BatteryLife: 90hoursoncontinuoususe(formanganese
batteries)
Overload
Protection: <500AAC,DC
Withstand
Voltage: 1000VACforoneminutebetweenelectrical
circuitandhousingcaseormetalsectionof
transformerjaws.
Insulation
Resistance: >10MΩ at1000Vbetweenelectricalcircuit
andhousingcaseormetalsectionof
transformerjaws.
Conductor
Size: <approx.19mm
Dimensions: 180(L)×536(w)×31.4(D)mm
Weight: Approx.170g.(batteriesincluded)
Accessories: (included)
OutputcordModel7061,2×1.5VSUM-3(R-
6)batteryorequivalent,carryingcase,
instructionmanual
(optional)
KEWEnergizerModel8021
KEWMultitransModel8004&8008

―4―
Thisinstructionmanualcontainswarningsandsafetyrulesthat
mustbe observedby theuser toensure safeoperation ofthe
instrumentandretainitinsafecondition.
Therefore,readtheseoperatinginstructionsthoroughly and
completelybeforeusingtheinstrument.
Payparticular attentionto allWARNINGS and CAUTIONS in
thisinstructionmanual.
WARNING indicates warning to avoid an electrical shock and
CAUTION indicatescautionstoavoiddamagetotheinstrument
andmakeaccuratemeasurements.
WARNINGS
1.Iftheinstrumentshowsthefollowingconditions,donottryto
makemeasurementsandhavetheinstrumentcheckedfor
inspectionorrepair.
Instrumentshowsvisualdamage.
Testleadsaredamaged.
Instrumentcannotbeoperatedforintendedmeasurements.
Instrumenthasbeenstoredforlongperiodoftime
underabnormalconditions.
Instrumenthasbeensubjectedtosevereshocksand
vibrations.
2.Thetransformerjawsaremadeofmetalandtheirtipsarenot
insulated.
Beespeciallycarefulaboutthehazardofpossibleshorting
wheretheequipmentundertesthasexposedmetalparts.
3.Donotmakemeasurementsinanexplosiveatmosphere
(i.e.inthepresenceofflammablegassesorfumes,vaperor
dust)
4.Nevergroundyourselfwhenconductingelectricaltests.Do
nottouchexposedmetalpipes,outlets,fixtures,etc,which
3.SafetyNotes

―5―
mightbeatgroundpotential.Keepyourbodyisolatedfrom
groundbyusingdryclothing,rubbershoes,rubbermats,or
anyapprovedinsulatingmaterial.
5.DONOTmakebatteryreplacementwithpowerswitchsetto
ONpositionandtestleadsconnectedtoinstrument.
6.DONOTuseModel8113onacircuitwithavoltageofhigher
than600VAC.
7.DONOTexposeittothedirectsun,extremesoftemperature
andhumidityordewfall.
CAUTIONS
1.Whennotinuseforalongperiodoftime,placethe
instrumentinstorageafterremovingthebatteriesfromit.
2.Nevermakemeasurementsiftheinstrument,thetestleadsor
yourhandiswet.
3.Calibrationandrepairofanyinstrumentshouldonlybe
performedbyqualifiedandtrainedservicetechnicians.
4.Donotinstallsubstitutepartsorperformanyunauthorized
modificationoftheinstrument.ReturntheunittoKyoritsuor
yourdistributerforrepair.
5.Useadampclothanddetergentforcleaningthecaseofthe
ontheinstrument.Donotuseabrasivesorsolvents.
6.AlwayssettheRangeSelectorSwitchtotheOFFposition
afteruse.
7.Theinstrumentmustbeusedbyacompetent,trainedperson
andoperatedinstrictaccordancewiththeinstructions.
Kyoritsuwillnotacceptliabilityforanydamageorinjury
causedbymisuseornoncompliancewiththeinstructionsor
safetyprocedures.Itisessentialtoreadandunderstandthe
safetyrulescontainedintheinstructions.Theymustbe
observedwhenusingtheinstrument.

―6―
4.InstrumentLayout
1 TransformerJaws
Pickupcurrentflowingthroughtheconductor.
2 JawTrigger
Operatesthetransformerjaws.Presstoopenthejaws.
3DCZeroAdj.Knob
ZeroadjustsonDCcurrentranges.
4PowerSwitch
Turnspoweronandoff.
5RangeSelectorSwitch
Selects2A,20Aor200Arangeposition.
6BatteryIndicator
LEDstartsflashingwhenthepowerswitchisturnedon.It
flashesatalongerintervalwhenthebatteryvoltagebecomes
exhaustedandfinallystopsflashingatapprox.2.2V.Also,
servesasapoweronandlowbatteryindicator.
7OutputTerminals
Outputvoltageproportionaltoinputcurrentisavailablefrom
theseterminals.
8HandStrap
Preventstheinstrumentfromslippingoffthehandduringuse.

―7――7―
BEFOREUSINGTHEADAPTOR,READTHESAFETY
NOTESCAREFULLYANDTHOROUGHLY.
5-1.Preparation
(1)Removethebatterycompartmentcoverandinstallthe
batteries.Refertosection6forbatteryreplacement.
(2)ADCvoltmeterorACvoltmeterisrequiredforDCorAC
currentmeasurement.Connecttheoutputcordstotheinput
terminalsofthevoltmeter.IncaseofDCcurrent
measurements,alwaysinserttheredbananaplugintothe
plus(+)terminalandtheblackbananaplugintothe
minus(−)terminalofanymultimeterwithaninputimpedance
ofatleast1kΩ asshowninFig.1.
5-2Measurements
Notes: ● Themaximumdiameterofaconductorunder
testis19mm.
Donotclampontoaconductorlargerthanthis
size.Accuratemeasurementscannotbemade
wherethe8113isclampedontoalarger
conductorwithitstransformerjawsnotfully
closed.
●Toobtainmaximumaccuracy,thetransformer
jawsarepreciselyadjusted.Exerciseextreme
caretoavoidshock,vibrationorexcessive
forcewhenusingtheclampadaptor.
5-2-1DCCurrentMeasurements(2A,20A&200Aranges)
(1)WiththepowerswitchsettoONposition,selectthedesiredrange.
5.MeasuringMethod

―8――8―
(2)RefertoTable1foroutputvoltageagainstinput.Setthe
multimetertotheappropriaterange.Whenusingadigital
multimeter(3-1/2digit),selectthe200mVDCor2VDCrange.
Table1
Note: Measurementscanbemadeupto5Aatthe2Arange
positionandupto50Aatthe20Arangeposition.
(3)Beforemakingmeasurements,adjustDCA0Adj.control
knobtozerooutput.Foraccuratemeasurements,itisa
recommendedpracticetomakethiszeroadjustmentfor
everymeasurement.Itisalsonecessarytomakethezero
adjustmentaftereveryrangechangingasitmaycausethe
instrumenttobeoffzero.
(4)Pressthetriggertoopenthetransformerjawsandclamponto
oneconductoronly.Sincethecurrentmeasuredwillbedisplayed
asvoltage,readitinamperesaccordingtoeachrange.Forexample,
takethereadingasfollowsusinga3-2/1digitdigitalmultimeter.
Table2
Note: Thedisplaymayslightlyflickeronthe2Arangebecause
ofhighsensitivitywhenthe200mVrangeisused.
RangeSwitch
Position
2ADC
20ADC
200ADC
OutputVoltage
0−500.0mV(100mV/A)against0−5ADCinput
0−500.0mV(10mV/A)against0−50ADCinput
0−200.0mV(1mV/A)against0−200ADCinput
RangeSwitch
Position
2A
20A
200A
DigitalTesterReadings
200mVRange
50.0mV
100.0mV
−
50.0mV
100.0mV
−
50.0mV
100.0mV
2VRange
0.050V
0.100V
0.500V
0.050V
0.100V
0.500V
0.050V
0.100V
0.50A
1.00A
5.00A
5.0A
10.0A
50.0A
50A
100A
Currents
measured