Knick SE605H Manuel utilisateur

User Manual SE605H
Conductivity Sensor
www.knick.de
Read before installation.
Keep for future use.

SE605H
2
Supplemental Directives
READ AND SAVE THIS DOCUMENT FOR FUTURE REFERENCE. BEFORE ATTEMPTING TO
ASSEMBLE, INSTALL, OPERATE OR MAINTAIN THE PRODUCT, PLEASE ENSURE A COMPLETE
UNDERSTANDING OF THE INSTRUCTIONS AND RISKS DESCRIBED HEREIN. ALWAYS OBSERVE
ALL SAFETY INFORMATION. FAILURE TO COMPLY WITH INSTRUCTIONS IN THIS DOCUMENT
COULD RESULT IN SERIOUS INJURY AND/OR PROPERTY DAMAGE. THIS DOCUMENT IS
SUBJECT TO CHANGE WITHOUT NOTICE.
These supplemental directives explain how safety information is laid out in this
document and what content it covers.
Safety Chapter
This document's safety chapter is designed to give the reader a basic understanding of
safety. It illustrates general hazards and gives strategies on how to avoid them.
Warnings
This document uses the following warnings to indicate hazardous situations:
Symbol Category Meaning Remark
WARNING Designates a situation that can lead to death or
serious (irreversible) injury.
The warnings con-
tain information on
how to avoid the
hazard.
CAUTION Designates a situation that can lead to slight or
moderate (reversible) injury.
None NOTICE Designates a situation that can lead to property
or environmental damage.
Symbols Used in this Document
Symbol Meaning
Reference to additional information
Interim or final result in instructions for action
Sequence of figures attached to an instruction for action
Item number in a figure
Item number in text

SE605H
3
Table of Contents
1 Safety .................................................................................................................... 5
1.1 Intended Use.......................................................................................................................... 5
1.2 Personnel Requirements.................................................................................................... 5
1.3 Residual Risks......................................................................................................................... 6
1.4 Hazardous Substances........................................................................................................ 6
1.5 Operation in Hazardous Locations................................................................................. 7
1.6 Electrical and Thermal Parameters in Hazardous Locations ................................. 7
2 Product ................................................................................................................. 9
2.1 Package Contents................................................................................................................. 9
2.2 Product Identification......................................................................................................... 10
2.3 Nameplates............................................................................................................................. 11
2.4 Symbols and Markings ....................................................................................................... 12
2.5 Design and Function ........................................................................................................... 13
2.6 Measuring Principle............................................................................................................. 14
2.7 CondCheck Special Version Design and Function.................................................... 14
3 Installation............................................................................................................ 15
3.1 General Installation Instructions..................................................................................... 15
3.2 Installation in ARF203 Flow-Through Fitting.............................................................. 17
3.3 Installation in Piping............................................................................................................ 18
3.4 Electrical Installation ........................................................................................................... 19
3.4.1 General Information on Electrical Installation ............................................ 19
3.4.2 Sensor Cable............................................................................................................ 19
4 Operation.............................................................................................................. 20
4.1 CondCheck Special Version: Checking the Measured Value................................. 20
5 Maintenance, Cleaning, Calibration ................................................................... 21
5.1 Maintenance .......................................................................................................................... 21
5.2 Cleaning................................................................................................................................... 21
5.3 Calibration............................................................................................................................... 22

SE605H
4
6 Maintenance......................................................................................................... 23
6.1 Replacing the O-Rings........................................................................................................ 23
7 Troubleshooting................................................................................................... 26
8 Decommissioning ................................................................................................ 27
8.1 Removing the Sensor.......................................................................................................... 27
8.2 Disposal.................................................................................................................................... 27
9 Accessories ........................................................................................................... 28
10 Dimension Drawings ........................................................................................... 31
11 Specifications ....................................................................................................... 32

SE605H
5
1 Safety
This document contains important instructions for the use of the product. Always fol-
low all instructions and operate the product with caution. If you have any questions,
please contact Knick Elektronische Messgeräte GmbH & Co. KG (sometimes hereafter
referred to as “Knick”) using the information provided on the back page of this docu-
ment.
1.1 Intended Use
The SE605H (the “product”) is a coaxial 2-electrode sensor with a measuring range of
0…1000µS/cm. The sensor body and sensing electrodes are made of stainless steel.
The sensor is ideal for use in the hygienic measuring and monitoring of ultrapure water.
The measurement data is output via a suitable industrial transmitter.
The defined operating conditions must be observed when using this product.
➜Specifications, p. 32
Sealing from the process is only ensured if the SE605H sensor is operated with Knick
weld-in sockets. ➜Accessories, p. 28
If weld-in sockets from other manufacturers are used, they must be tested and evalu-
ated by the operating company.
1.2 Personnel Requirements
The operating company shall ensure that any personnel using or otherwise interacting
with the product is adequately trained and has been properly instructed.
The operating company shall comply and cause its personnel to comply with all appli-
cable laws, regulations, codes, ordinances and relevant industry qualification standards
related to product. Failure to comply with the foregoing shall constitute a violation of
operating company’s obligations concerning the product, including but not limited to
an unintended use as described in this document.

SE605H
6
1.3 Residual Risks
The product has been developed and manufactured in accordance with generally ac-
cepted safety rules and regulations. The SE605H sensor was subjected to a risk assess-
ment. Nevertheless, not all risks can be sufficiently reduced; the following residual risks
remain:
Environmental Influences
The effects of pressure, moisture, corrosion, chemicals, and ambient temperature can
negatively impact on safe operation of the product.
Observe the following instructions:
• Only operate the SE605H sensor in compliance with the stated operating condi-
tions. ➜Specifications, p. 32
• If using chemically aggressive process media, regularly check the SE605H sensor for
damage.
• Adhering and sticky process media may interfere with measurements. Regularly
remove adherents. As required, recalibrate the SE605H sensor. ➜Cleaning, p. 21
➜Calibration, p. 22
1.4 Hazardous Substances
In certain situations (e.g., sensor replacement), personnel may come into contact with
the following hazardous substances:
• Process medium
• Cleaning medium
This document specifies the personal protective equipment that needs to be used
when carrying out handling instructions.
The operating company is responsible for conducting a hazard assessment.
See the relevant manufacturers’ safety data sheets for hazard and safety instructions on
handling hazardous substances.

SE605H
7
1.5 Operation in Hazardous Locations
The SE605H-X is certified for operation in explosive atmospheres.
Observe all applicable local and national codes and standards for the installation of
equipment in explosive atmospheres. For further guidance, consult the following:
• IEC60079-14
• EU directives 2014/34/EU and 1999/92/EC(ATEX)
• NFPA70(NEC)
• ANSI/ISA-RP12.06.01
The electrical and thermal parameters of the sensors must be adhered to.
➜Electrical and Thermal Parameters in Hazardous Locations, p. 7
Memosens Ex sensors are marked by an orange-red ring.
Combined with a model CA/MS‑***X**, model CA/MS‑***X**‑L measuring cable, or a
certified measuring cable that is identical in hardware and function, the sensor may be
connected to a suitable measuring device, as specified in the Certificates BVS 15 ATEX
E141 X and IECEx BVS 15.0114 X.
1.6 Electrical and Thermal Parameters in Hazardous Locations
Certificate Number Marking
BVS 16 ATEX E 037 X II 1G
IECEx BVS 16.0030X Ex ia IIC T4 Ga
JPEx DEK19.0046X Ex ia IIC T4 Ga
Thermal Parameters
Temperature class Ambient Temperature
Range Ta
Maximum Permissible
Process Temperature
T4 -20 °C < Ta < 115 °C
(-4°F<Ta<239°F)
115 °C
(239°F)
Special Conditions
• The measuring cable and sensor may only be used within the ambient temperature
range specified for the temperature class.
• The measuring cable, including its connecting head, must be protected from elec-
trostatic charging if it passes through areas of Zone 0 (category 1G).

SE605H
8
• The Memosens sensors may not be operated in electrostatically critical process
conditions. Intense vapor or dust flows directly impacting on the connection
system must be avoided.
• Metallic process connection parts must be mounted at the installation site so that
they are electrostatically conductive (<1MΩ).

SE605H
9
2 Product
2.1 Package Contents
• SE605H in the version ordered
• User Manual
• Quality Certificate
• Control Drawing1)
• EU Declaration of Conformity1)
1) Part of package contents only for Ex-approved products.

SE605H
10
2.2 Product Identification
The different versions of the SE605H are encoded in a product code.
2.2.1 Product Code
Hygienic 2-Electrode Conductivity Sensor SE605H - _ _ _ _ _ / _ _ _ - _ _ _
Explosion
protection
None N
ATEX X
Communication Memosens M
S
Process connection Ingold socket, 25 mm (G 1¼″)H 0
Ingold socket, 25 mm (G 1¼″), 50 mm long H Z
Clamp 1.5″J 1
Clamp 2″J 2
Gasket material FKM FDA F
EPDM FDA E
FFKM FDA H
FKM FDA USP VI V
EPDM FDA USP VI U
FFKM FDA USP VI W
Special version 1) None / 0 0 0
Customer-specific special datasheet / 0 0 F
CondCheck 2) / 0 0 P
Certificates 3) None - 0 0 0
Inspection Certificate 3.1 according to
EN 10204 - 0 0 3
FDA - USP VI - 0 0 U
Surface Ra < 0.4 µm - 0 0 4
Surface Ra < 0.8 µm - 0 0 8
FDA - 0 0 F
1) Can be combined.
2) Only available without explosion protection (N).
3) Can be combined. Max. three certificates possible.