KAM Simple Precision OFD Manuel utilisateur

API COMPLIANT
TEL +1 713 784 0000
FAX +1 713 784 0001
3939 Ann Arbor Drive
Houston, Texas 77063 USA
www.KAM.com
User Manual
Kam® OFD™
Optical Fluorescence Detector PTB 08 ATEX 1026
OFDMANUAL-0816
An ISO 9001 certified company

1
OFDMANUAL 0816
TABLE OF CONTENTS
SECTION TITLE PAGE
1Available Models and Mounting Options 2
2Specifications 3
∙ Specifications 3
∙ Dimensional Drawings 4
3Installation 6
∙ Sensor location 6
∙ Do's and Dont's 7
∙ Main Line 9
∙ Removal 12
∙ Analyzer Loop 13
∙ Wiring 14
∙ Connecting the OFD 15
4Field Calibration 16
5Maintenance 18
Appendix A Modbus Register 22
CAUTION:
When installing the OFD™ sensor in a pipeline containing petroleum products,
petrochemicals, waste waters with the presence of pressure & temperature, and high-pressure
steam refer to the Pipeline Operators’ “Health, Safety and Environmental Policy Procedures” to
ensure safe installation.
KAM CONTROLS, INC. reserves the right to make changes to this document without notice.

2
OFDMANUAL 0816
INTRODUCTION
AVAILABLE MODELS and MOUNTING OPTIONS
FIG. 1-1 FIG. 1-2 FIG. 1-3
Rectractable OFD™
on a main pipe, with
2", 3", or 4" flanged
seal housing
Rectractable OFD™
on a main pipe, with
2" MNPT seal housing
OFD™ FT Flow Through
on a densitometer loop
with ¾" or 1" MNPT
Full-
opening
Ball
Valve
Full-
opening
Ball
Valve

3
OFDMANUAL 0816
SPECIFICATIONS
Media: Refined products
Material: Wetted parts–316 stainless steel
Power: 12–24 VDC 15 Watts max (Max current is 1275 mA @ 12V and 637 mA @ 24V)
Output: 4-20 mA
Communication Port: RS485 (MODBUS RTU)
RS232 (Diagnostics)
Fluid temperature: -40º to 160º F (-40º to 70º C)
Electronics temp.: -40º to 160º F (-40º to 70º C)
Pressure ratings: ANSI 150, 300, 600, 900 (flow-through models only)
Mounting: 3/4" and 1" MNPT Flow Through
2" MNPT Seal Housing Insertable/Retractable
2", 3", or 4" Flanged Seal Housing Insertable/Retractable
EX enclosure: 3" x 6" x 3" (76 mm x 152 mm x 76 mm)
Shaft length: 12" to 60" – Off-the-shelf lengths are 24", 30", 36", 48", and 60"
(610 mm to 1524 mm)
(Off-the-shelf 610 mm, 762 mm, 914.4 mm, 1219 mm, 1524 mm)
Pipe size: 3/4" to 48" (20 mm to 1200 mm)
Weight: from 10 lbs. (4.5 kg)

L= 24" 30" 36" or 48" other sizes can be
4
OFDMANUAL 0816
SPECIFICATIONS CONTINUED
DIMENSIONAL DRAWINGS
"
2
3
4
INCHES
8.40
8.60
8.60
INCHES
8.50
8.75
8.90
INCHES
8.90
9.15
9.40
INCHES
9.40
9.40
9.65
MM
213
218
218
MM
216
222
226
MM
226
232
240
MM
240
240
245
150 300 600 900
TABLE 2-2 Flange Size and Class (SL)
A
B
C
D
E
INCHES
1.443
1.25
1
4.53
7
MM
37
32
25
115
178
TABLE 2-3 DIMENSIONS
Shaft Lengths are available in .5" (12.7
mm) increments.
Standard sizes are 24", 30", 36", 48",
and 60". (610 mm, 762 mm, 914.4 mm,
1219 mm, 1524 mm)
FIG. 2-1 OFD™ Sensor with Flanged Seal Housing
A
Seal Housing
¾" FNPT
B
Shaft Length ± .5"
SL
C (Shaft Diameter)
D
E

5
OFDMANUAL 0816
SPECIFICATIONS CONTINUED
FIG. 2-4 FLOW THROUGH OFD™ for ANALYZER LOOP
INCHES
.3
.5
1. 4
2.25
6.75
7
4.53
1. 2
MM
8
13
36
57
171
178
115
30
A
B
C
D
E
F
G
H
TABLE 2-5 ¾" MNPT DIMENSIONS
INCHES
.3
.5
1.45
2.4
6.75
7
4.53
1. 3
MM
8
13
37
61
171
178
115
33
A
B
C
D
E
F
G
H
TABLE 2-6 1" MNPT DIMENSIONS
G
F
A
B
H
C
D
E
¾" FNPT
¾" or 1"
MNPT

6
OFDMANUAL 0816
INSTALLATION
LOCATION
For optimal batch detection, KAM CONTROLS recommends that you install the in-station OFD™ sensor at the
first accessible pipeline location inside the terminal fence-line — upstream of the interface cut valve(s). This al-
lows the operator ample time to open/close the cut valves prior to the arrival of the product interface.
KAM CONTROLS also strongly recommends that you utilize a preview (or out-station) OFD™ sensor. This lets the
operator decide how to optimize each batch cut prior to actually making the batch cut at the in-station and gives
the operator more confidence in their decisions as well as the time to identify and resolve any issues that may
arise during a critical interface.
OUT-STATION/PREVIEW
OFD
INCOMING
OFD
TYPICALLY
1 - 2 MILES
(1.6–3.2km)
TO TANK
TO TANK
TO TANK
IN-STATION/CUT
OFD
TERMINAL FENCE LINE
CUT VALVES
TERMINAL FENCE
FIG. 3-1 RECOMMENDED OFD™ LOCATIONS
PRIORTO INSTALLATION
Remove all the protective packaging materials including the red cap placed at the tip of the sensor probe. En-
sure that the OFD™ sensor was not damaged during transit.
CAUTION: Do not stand the OFD™ sensor on the probe end or allow the probe to hit the ground or any hard
surface under any circumstances. This will damage the lens, and the OFD™ sensor will not work properly.

7
OFDMANUAL 0816
I
P
INSTALLATION CONTINUED
INSTALLATION DO’S AND DON’TS
P
FLOW
FLOW
P
FLOW
FLOW
P
FLOW
FLOW
P
FLOW
FLOW
P
FLOW
FLOW
P
FLOW
FLOW
DO NOT install the
fast loop OFD™
sensor in a straight
portion of pipe.
It needs to be
mounted off the
bend opposite the
pump.
DO NOT install the
OFD™
sensor with the
lens facing directly
into the flow. If
the product has
particulate
matter in the fluid,
like sand, this will
sandblast the lens
and could cause
premature failure.
DO NOT attempt to
screw the OFD™
sensor either in
or out by hand.
Always use a
1 1/4" or 1 3/8"
wrench on the
wrench flat below
the electronics
enclosure.
I

8
OFDMANUAL 0816
P
FLOW
FLOW
P
INSTALLATION CONTINUED
INSTALLATION DO’S AND DON’TS
Minimum 8"
(20.3 cm)
DO NOT use teflon
tape on the OFD™
sensor threads. DO
use liquid thread
sealant.
DO install the
OFD™ sensor with
a minimum of 8"
or 200mm be-
tween the lens and
the nearest flat
surface.
DO install the
OFD™ sensor with
a sun shade if the
electronics are
directly exposed to
sunlight.

9
OFDMANUAL 0816
Remove all the protective packaging materials, and ensure that the OFD™ sensor was not damaged during transit.
PRIORTO INSTALLATION
MAIN LINE INSTALLATION
The KAM®OFD™ sensor should be installed
according to FIG. 3-2. KAM CONTROLS
recommends installing the OFD™ sensor at a 2 or
10 o’clock position to ensure the tip of the probe
remains in the fluid. A full opening ball valve is
used to isolate the OFD™ sensor from the pipeline
during installation or removal. The seal housing of
the OFD™ sensor allows the optical probe to be
inserted and removed from the pipe under pressure
and flow conditions. It is the user’s responsibility to
ensure that the OFD™ sensor is placed at the most
representative point within the flow profile. The
OFD™ sensor should be inserted so that the tip of
the probe is located 1/4" above the inner wall of the
pipeline. This ensures that the probe is not
damaged when pigging the pipeline.
NOTE: If line pressure exceeds 100 psi, use a KAM®
IT Insertion Tool when installing/removing
the KAM®OFD™ sensor.
Lay the OFD™ sensor on the ground or a table.
Loosen Socket Cap Screws on the locking
collar. This will allow the OFD™ shaft to slide
through the seal housing.
Push the OFD™ shaft though the seal housing
until the OFD™ probe sits flush with the end of
the seal housing or seal housing flange. FIG.
3-3 and 3-4. (Remove red protection cap on the
tip of the probe if it has not been removed.)
Place a mark with a sharpie or a permanent
marker on the shaft at the edge of the locking
collar. (Do not use anything sharp to mark the
shaft as this will create grooves that will
damage the O-rings in the seal housing.)
1.
2.
3.
4.
Locking Collar
Seal Housing
Full-opening
Ball Valve
Socket Cap
Screw
FIG. 3-2
FIG. 3-4
INSTALLATION CONTINUED
Prior to mounting the OFD™ sensor on the
Full-opening Ball Valve, you must determine
the insertion length required.
Mark here
Mark here
FIG. 3-3