Use a lower setting for patients new to the InterX 1000. On
the first day of a new treatment program, start on Preset
2, the lowest frequency setting. Use this setting for the
first 1-2 days then switch to Preset 1 for the next 1-2 days
of treatment.
As the body gets used to the stimulation, go to Preset 3,
then Preset 4. You’ll want to rotate through Presets
settings 1-4 throughout the treatment program.
It helps to make a note of which Preset is used every day
and what areas are treated.
Do not progress to the next highest preset setting if
symptoms increase at a lower preset setting. Keep treating
at the lower settings until positive results are seen.
Preset 5 is for acute or emergency treatments or very
strong pain that needs to be worked through. Please note,
setting 5 can either work very well on strong pain or irritate
it. By design, setting 5 does not work on the flex arrays.
If you need additional information on the device or
its use, please refer to the supplied Operations
Manual or call our Support Team at (972) 665-1810.
Preset 1 = 60hz
Preset 2 = 15-60hz
Preset 3 = 30-120hz
Preset 4 = 240hz
Preset 5 = 480hz