•The instrument can e used only under conditions that are specified in
Technical Specification, see Chapter 5.
•Do not expose the instrument to aggressive gases, vapours, liquids and dust.
•If you have transferred the unit from cold to hot environment, it can cause the
condensation. We recommend a short acclimatization.
•If the device will e out of operation for a longer time, it is recommended to
remove the atteries. This prevents the possi ility of leakage into the device.
Leakage can cause serious damage or to destroy the instrument.
•RCD under test can trip-out even if it should not trip-out during particular
measurement(s). This may e due, inter alia, faulty RCD or ecause of a
leakage current flow via RCD under test which adds up the differential current
generated y the instrument. This may result in interruption of operation of
various equipment(s) and cause damage (e.g. loss of data in computers) and
/ or threats, including threats to life or health (e.g. health facilities). Therefore,
we strongly recommend that measurements are carried out in agreement with
a person who is responsi le for the operation of the o ject under test and who
implement measures to prevent any damage. The simplest such measure (if it
is possi le) is to turn off such equipment(s).
•The instrument contains two fairly strong magnets. Do not leave them near
the equipment and items that could e damaged y the magnetic field - such
as watches, credit cards with magnetic strips, etc.
•Images in this manual are illustrative and may vary slightly from the actual
state.
1.2. Gene al desc iption of the inst ument
The FITESTpro is a compact instrument with patent-protected storage system of
the test tips in the transport position – sharp tips are safely hidden.
High contrast right multicolour graphic OLED display ensures excellent
legi ility. When measured under low light conditions it is possi le to illuminate
the measured o ject y a right white LED light positioned on the front side of
the housing.
The FITESTpro can measure:
- RCD trip-out time
- RCD trip-out current
- RCD contact voltage
- Loop resistance (no-trip of RCD)
- AC voltage
- Phase (live) conductor test
1.3. Standa ds applied
Measurements: EMC: Safety:
EN 61557-1 EN 55022, class B EN 61010-1
EN 61557-6 EN 61326-1 EN 61010-2-031
EN 61000-4-2,3,4,5,6