HRRS SERIES
High Resistance
Decade Substituter
User and Service Manual
Copyright © 2013 IET Labs, Inc.
Visit www.ietlabs.com for manual revision updates
HRRS im/January 2013

Email: info@ietlabs.com
TEL: (516) 334-5959 • FAX: (516) 334-5988
www.ietlabs.com
IET LABS, INC.

Email: info@ietlabs.com
TEL: (516) 334-5959 • FAX: (516) 334-5988
www.ietlabs.com
IET LABS, INC.
i
HRRS Series
WARRANTY
We warrant that this product is free from defects in material and workmanship and, when properly used,
will perform in accordance with applicable IET speci cations. If within one year after original shipment,
it is found not to meet this standard, it will be repaired or, at the option of IET, replaced at no charge when
returned to IET. Changes in this product not approved by IET or application of voltages or currents greater
than those allowed by the speci cations shall void this warranty. IET shall not be liable for any indirect,
special, or consequential damages, even if notice has been given to the possibility of such damages.
THIS WARRANTY IS IN LIEU OF ALL OTHER WARRANTIES, EXPRESSED OR IMPLIED,
INCLUDING BUT NOT LIMITED TO, ANY IMPLIED WARRANTY OF MERCHANTABILITY OR
FITNESS FOR ANY PARTICULAR PURPOSE.

ii
HRRS Series
WARNING
OBSERVE ALL SAFETY RULES
WHEN WORKING WITH HIGH VOLTAGES OR LINE VOLTAGES.
Dangerous voltages may be present inside this instrument. Do not open the case
Refer servicing to qualifi ed personnel
HIGH VOLTAGES MAY BE PRESENT AT THE TERMINALS OF THIS INSTRUMENT
WHENEVER HAZARDOUS VOLTAGES (> 45 V) ARE USED, TAKE ALL MEASURES TO
AVOID ACCIDENTAL CONTACT WITH ANY LIVE COMPONENTS.
USE MAXIMUM INSULATION AND MINIMIZE THE USE OF BARE
CONDUCTORS WHEN USING THIS INSTRUMENT.
Use extreme caution when working with bare conductors or bus bars.
WHEN WORKING WITH HIGH VOLTAGES, POST WARNING SIGNS AND
KEEP UNREQUIRED PERSONNEL SAFELY AWAY.
CAUTION
DO NOT APPLY ANY VOLTAGES OR CURRENTS TO THE TERMINALS OF THIS
INSTRUMENT IN EXCESS OF THE MAXIMUM LIMITS INDICATED ON
THE FRONT PANEL OR THE OPERATING GUIDE LABEL.

iii
HRRS Series
Table of Contents
Table of Contents
Chapter 1 Introduction
1.1 Introduction ........................................................................................................... 1
Chapter 2 Specications
2.1 Specications ........................................................................................................ 2
2.2 Ordering Information ............................................................................................ 3
Chapter 3 Operation
3.1 Initial Inspection and Setup .................................................................................. 5
3.2 Dial Setting ........................................................................................................... 5
3.3 Environmental Conditions .................................................................................... 5
3.4 Storage .................................................................................................................. 5
Chapter 4 Maintenance
4.1 Maintainability and Reliability ............................................................................. 6
4.2 Preventive Maintenance ........................................................................................ 6
4.3 Verication of Performance .................................................................................. 6
4.3.1 Calibration Interval ..................................................................................... 6
4.3.2 General Considerations ............................................................................... 6
4.3.3 Calibration Procedure ................................................................................. 7
4.4 Schematic .............................................................................................................. 7
4.5 Replaceable Parts .................................................................................................. 8
Figures and Tables
Figure 1-1: HRRS Series High Resistance Decade Substituter ...............1
Figure 2-1: HRRS sample label ................................................................4
Figure 4-1: HRRS Series Schematic Diagram .........................................7
Table 4-1: Replaceable Parts ...................................................................8
Figure 4-2: Replaceable Parts .................................................................8

iv
HRRS Series
Table of Contents
This page is intentionally left blank.

1
HRRS Series
1Introduction
Chapter 1
INTRODUCTION
1.1 Introduction
The High Resistance Decade Substituter (HRRS
Series) is a family of instruments offering a broad
choice of high range, excellent performance resis-
tance sources (Figure 1.1). High-resistance design
is implemented with special resistors, switches,
binding posts, and construction-design to allow high
resistance performance without sacrificing other
electrical properties.
The HRRS Series employs state-of-the-art precision
resistors of various types for high accuracy, high sta-
bility, and low temperature and voltage coefcients.
The standard models offer a choice of one to seven
decades. For custom units, models up to 11 decades
are available . The panels are clearly labeled showing
the step size and maximum allowable applied input
for each decade. Higher voltage models up to >10 kV
are also available from IET Labs.
The binding posts are insulated with Kel-F high-
resistance, non-moisture-absorbing material ensure
low leakage.
With a resolution as low as 10 Ω and a maximum
available resistance of over 1 TΩ (= 1000 GΩ;
= 1000 kMΩ), the HRRS Series may be used for
exacting precision high-value resistance applications.
Steps up to 10 TΩ are also available from IET Labs,
Applications include calibration of meters and
megohmmeters, and checking of electrochemical
and biomedical sensors and instruments. These
instruments are useful wherever small currents and
high resistances are required, such as in testing high-
impedance ampliers and the insulation of low-power
circuits.
The HRRS series complements the HARS series
which provides resistance steps as low as 1 mΩ. The
units may be rack-mounted to serve as components
in measurement and control systems.
Figure 1-1: HRRS Series High Resistance
Decade Substituter

2
HRRS Series
2 Specifications
Chapter 2
SPECIFICATIONS
For convenience to the user, the pertinent specications are given in an OPERATING GUIDE afxed to the
case of the instrument. Figure 2.1 shows a typical example.
2.1 Specifications
Resistance
per step
Total
decade
resistance
Accuracy
18-28°C; <50% RH Max voltage
per step
(V)
Maximum
voltage
(V)
Temperature
coefficient
(±ppm/°C)
Voltage
coefficient
(±ppm/V)
Stability
(±ppm/yr)
Resistor
type
Q B F
10 Ω100 Ω±(0.01%+2 mΩ) ±(0.03%+2 mΩ) ±(0.10%+2 mΩ) 2.5 25 15 - 10
Wirewound,
non-inductive
100 Ω1 kΩ±(0.01%+2 mΩ±(0.03%+2 mΩ) ±(0.10%+2 mΩ) 8 80 5 - 10
1 kΩ10 kΩ±0.01% ±0.03% ±0.10% 23 230 5 - 10
10 kΩ100 kΩ±0.01% ±0.03% ±0.10% 70 700 5 - 10
100 kΩ1 MΩ±0.01% ±0.03% ±0.10% 230 2000 5 - 10
1 MΩ10 MΩ±0.01% ±0.03% ±0.10% 1000** 2000 15 <1 25
Metal-oxide
film
10 MΩ100 MΩ±0.03% ±0.10% ±1% 1000** 2000 15 <1 50
100 MΩ1 GΩ±0.10% ±0.20% ±1% 1000** 2000 15 1 100
1 GΩ10 GΩ±0.20% ±0.50% ±1% 1000** 2000 50 1 500 High-voltage
film
10 GΩ100 GΩ±0.50% ±1% ±1% 1000** 2000 50 2 500
100 GΩ1 TΩ
See HRRS-5kV Series if these values are required.
1 TΩ10 TΩ
**To apply up to 2000 V for the rst step, use the preceding decade at its “10” position. e.g. to obtain 1 MΩ with 2000 V max, set 100 kΩ switch to position “10.”
Zero resistance
<3 mΩ per decade
Environmental conditions
Operating Conditions: 10°C to 40°C; <50% RH
Storage conditions: -40°C to 70°C
Terminals
Two ve-way binding posts on 2 special, low-
leakage, Kel-F insulating sockets, and one metal
ground binding post electrically connected to the
case
Max voltage to case
2000 V peak
Mechanical:
Model Dimensions Weight
3 Decade 31.2 cm W x 8.9 cm H x 10.2 cm D
(12.3" x 3.5" x 4.0")
1.4 kg
(3.0 lb)
4-5 Decade 37.5 cm W x 8.9 cm H x 10.2 cm D
(14.8" x 3.5" x 4.0")
1.6 kg
(4.0 lb)
6-7 Decade 43.9 cm W x 8.9 cm H x 10.2 cm D
(17.3" x 3.5" x 4.0")
2 kg
(4.5 lb)
8-9 Decade 48.3 cm W x 17.8 cm H x 19.7 cm D
(19.0" x 7.0" x 7.8")
5.1 kg
(11 lb)

3
HRRS Series
3Specifications
Model Total resistance Number of
decades Resolution
HRRS-*-7-100 1.111 111 GΩ7 100 Ω
HRRS-*-7-1k 11.111 11 GΩ7 1 kΩ
HRRS-*-7-10k 111.111 1 GΩ7 10 kΩ
HRRS-*-8-10 1,111.111 1 MΩ8 10 Ω
HRRS-*-8-100 11.111 111 GΩ8 100 Ω
HRRS-*-8-1k 111.111 11 GΩ8 1 kΩ
HRRS-*-9-10 11,111.111 1 MΩ9 10 Ω
HRRS-*-9-100 111.111 111 GΩ9 100 Ω
To specify accuracy, replace * with “Q,” “B,” or “F” as required
Options
-RM: Rack mountable case for standard 19” rack
-K: Kelvin-type 4-terminal posts
-RO: Rear output
2.2 Ordering Information
Model Total resistance Number of
decades Resolution
HRRS-*-3-1M 1.11 GΩ3 1 MΩ
HRRS-*-3-10M 11.1 GΩ3 10 MΩ
HRRS-*-3-100M 111 GΩ3 100 MΩ
HRRS-*-4-100k 1.111 GΩ4 100 kΩ
HRRS-*-4-1M 11.11 GΩ4 1 MΩ
HRRS-*-4-10M 111.1 GΩ4 10 MΩ
HRRS-*-5-10k 1.111 1 GΩ5 10 kΩ
HRRS-*-5-100k 11.111 GΩ5 100 kΩ
HRRS-*-5-1M 111.11 GΩ5 1 MΩ
HRRS-*-6-1k 1.111 11 GΩ6 1 kΩ
HRRS-*-6-10k 11.111 1 GΩ6 10 kΩ
HRRS-*-6-100k 111.111 GΩ6 100 kΩ

4
HRRS Series
4 Specifications
Figure 2-1: HRRS sample label
IET LABS, INC. • Long Island, NY 11590 • Email: info@ietlabs.com • Tel: (516) 334-5959
HRRS BLBL/01-16-2013
Observe all safety rules when working with high voltages or line voltages. Connect the (G) terminal to earth ground in order to maintain the case at a safe volt-
age. Whenever hazardous voltages (> 45 V) are used, take all measures to avoid accidental contact with any live components: a) Use maximum insulation and
minimize the use of bare conductors. b) Remove power when adjusting switches. c) Post warning signs and keep personnel safely away.
HRRS SERIES HIGH-RESISTANCE DECADE SUBSTITUTER OPERATING GUIDE
CONSULT INSTRUCTION MANUAL FOR PROPER INSTRUMENT OPERATION
Keep unit in a sealed environment when not in use.
Maintain binding post area clean for minimum electrical leakage.
CAGE CODE: 62015 • www.ietlabs.com
Resistance
per step
Total
decade
resistance
Accuracy
18-28°C; <50% RH Max voltage
per step
(V)
Max
Voltage
(V)
TC
(±ppm/ºC)
VC
(±ppm/V)
Stability
(±ppm/
yr)
Resistor
Type
Q B F
10 Ω 100 Ω ±(0.01%+2 mΩ) ±(0.03%+2 mΩ) ±(0.10%+2 mΩ) 2.5 25 15 - 10
Wirewound,
non-inductive
100 Ω 1 kΩ ±(0.01%+2 mΩ ±(0.03%+2 mΩ) ±(0.10%+2 mΩ) 8 80 5 - 10
1 kΩ 10 kΩ ±0.01% ±0.03% ±0.10% 23 230 5 - 10
10 kΩ 100 kΩ ±0.01% ±0.03% ±0.10% 70 700 5 - 10
100 kΩ 1 MΩ ±0.01% ±0.03% ±0.10% 230 2000 5 - 10
1 MΩ 10 MΩ ±0.01% ±0.03% ±0.10% 1000** 2000 15 <1 25
10 MΩ 100 MΩ ±0.03% ±0.10% ±1% 1000** 2000 15 <1 50 Metal-oxide
lm
100 MΩ 1 GΩ ±0.10% ±0.20% ±1% 1000** 2000 15 1 100
1 GΩ 10 GΩ ±0.20% ±0.50% ±1% 1000** 2000 50 1 500 High-voltage
lm
10 GΩ 100 GΩ ±0.50% ±1% ±1% 1000** 2000 50 2 500
**To apply up to 2000 V for the first step, use the preceding decade at its "10" position. e.g. to obtain 1 MΩ with 2000 V max, set
100 kΩ switch to position "10."
Zero resistance
<3 mΩ per decade
Terminals
Two ve-way binding posts on 2 special, low-leakage,
Kel-F insulating sockets, and one metal ground binding
post electrically connected to the case
Environmental conditions
Operating Conditions: 10°C to 40°C; <50% RH
Storage conditions: -40°C to 70°C
Max voltage to case
2000 V peak
WARNING
MODEL: HRRS-Q-2-100M
SN: B2-1306578
Autres manuels pour HRRS SERIES
2
Ce manuel convient aux modèles suivants
57
Table des matières
Autres manuels IET Labs Équipement de test
Manuels Équipement de test populaires d'autres marques

SMART
SMART KANAAD SBT XTREME 3G Series Manuel utilisateur

Agilent Technologies
Agilent Technologies BERT Serial Manuel utilisateur

Agilent Technologies
Agilent Technologies N3280A Manuel utilisateur

Vernier
Vernier Go Direct Voltage Manuel utilisateur

Lifeloc
Lifeloc R.A.D.A.R. Manuel utilisateur

Fluke
Fluke T5-600 Instructions d'utilisation et d'installation