4DC-TH 201x Installation Guide
1. Installation Method
1.1 Notes on Installation
1.2 Required Settings of Camera
- This camera is designed for indoor use only and should be installed where constant temperature
can be maintained. Outdoor usage is strongly not recommended.
ㆍDo not point the camera towards the entrance or the outside of the building.
ㆍDeploy the solution indoors only in an environment with a constant temperature and no wind.
ㆍIt is recommended to set up a queuing area to manage the ow of people.
- Thermal sensor should measure human skin temperature. Therefore, for accurate scanning,
individuals must face the camera directly straight without wearing any hats, hoods, glasses, etc.
those might cover the facial skin.
- IPInstaller.exe is necessary for modifying the camera settings. This .exe le can be downloaded
from below link. (ID: admin / PW: Thermal#)
IPInstaller.exe
- Thermal sensor detects and alerts all the temperature that is higher than pre set-up ‘Event Max
Temperature’. In thermal image, there must not be any electronic equipment or devices possessing
heat. (e.g. smart phones, light, lamp, etc.)
- Scanned temperature varies with distance from the camera. Below chart is our examination in
ofce. As shown from the chart, at least 0.5°C is decreased by a meter by meter. To minimize the
temperature difference, it is strong and highly recommended to install the camera and measure the
temperature within less than 2 meters from the moving subject.
Before registering the camera to NVR or VMS, all the required settings should be modied by using
IPInstaller.exe. Settings that must be modied are as below.
- IP address (the default IP address is 192.168.0.100)
- ID / PW : admin / Thermal#
- Setting the proper temperature seen in the image
The temperature can be calibrated in SETUP – Video - Thermal Parameter Setup - Offset.
Dis.(m) Skin Temperature test (°C) Remark
1st 2nd 3rd 4th 5th
1 36.61 37.03 38.70 38.80 37.8 -The measured temperature is not xed
and varies by up to ± 0 to 1°C for each
distance.
- As the distance increases from the
cam, the temperature is generally lower.
- DC-TH2011 model
2 35.45 36.26 37.70 38.10 36.8
3 34.99 35.58 37.30 37.50 36.50
4 35.06 35.13 36.60 36.50 35.50
※The skin temperature varies according to environment and person. So calibration is required based on each user. When installing, the
user must run several tests and gain data like the above chart and then determine proper ‘Event Max Temperature’. For example, refer-
ring to above chart data, it would be proper to set ‘Event Max Temperature’ 38°C . If so, the event alarm will mostly be active around
one meter or closer. After alarm, a user must measure the temperature once more by a thermometer for accuracy.