Greyline Instruments SLT 5.0 Manuel utilisateur

www.greyline.com
USER'S GUIDE
Installation & Operation
Instructions
Level & Flow Monitor
Model SLT 5.0
Manual Series B.1.4

Note: This page has been left blank intentionally.

Page 3
SLT 5.0 Level and Flow Monitor
INDEX
CONNECTIONS: ...............................................................................................4
KEYPAD SYSTEM............................................................................................6
CALIBRATION MENU.....................................................................................7
RUN....................................................................................................................8
MESSAGE..........................................................................................................8
STATUS..............................................................................................................8
PASSWORD.......................................................................................................9
UNITS/MODE..................................................................................................10
CALIBRATION................................................................................................11
CHANNEL SETUP ..........................................................................................14
RELAY PARAMETERS..................................................................................15
SPECIAL FUNCTIONS...................................................................................16
SENSOR MOUNTING/LOCATION...............................................................18
SENSOR MOUNTING METHODS ................................................................20
APPLICATIONS HOTLINE............................................................................27
PRODUCT RETURN PROCEDURE ..............................................................28
APPENDIX A - OPTIONS...............................................................................30
DATA LOGGING (Optional)...........................................................................33
CONVERSION GUIDE ...................................................................................35
SPECIFICATIONS...........................................................................................36
IMPORTANT NOTE: This instrument is manufactured and calibrated to meet product specifications.
Please read this manual carefully before installation and operation. Any unauthorized repairs or
modifications may result in a suspension of the warranty.
Available in Adobe Acrobat pdf format

Page 4
SLT 5.0 Level and Flow Monitor
CONNECTIONS:
POWER INPUT: The standard model requires AC power input between 100 to 240 VAC 50/60Hz . No
adjustments are necessary for voltages within this range.
Optional DC input model requires 9-32 VDC/9 Watts. Connect to + and - terminals.
Optional Thermostat and Heater modules are available rated for 115 VAC or 230 VAC.
IMPORTANT NOTE: To comply with CE electrical safety standards, AC power input and relay
connection wires must have conduit entry to the instrument enclosure. Installation requires a switch,
overcurrent fuse or circuit breaker in the building (in close proximity to the equipment) that is marked
as the disconnect switch.
Risk of electric shock. Loosen cover screw to access connections. Only qualified personnel
should access connections.
Note: Use of instrumentation over 40°C ambient requires special field wiring.
Note: User replaceable fuse is 2 Amp 250V (T2AL250V).

Page 5
SLT 5.0 Level and Flow Monitor
CONNECTIONS

Page 6
SLT 5.0 Level and Flow Monitor
KEYPAD SYSTEM
The following diagram shows the SLT 5.0 menu system. Arrows show the four directions to leave a
menu box. Pressing a corresponding keypad arrow will move to the next item in the direction shown.
Move the cursor (underline) under numerals and increase or decrease numerals with the and keys.
To store calibration values permanently (even through power interruptions), press the .

Page 7
SLT 5.0 Level and Flow Monitor
CALIBRATION MENU

Page 8
SLT 5.0 Level and Flow Monitor
RUN
The main display shows the units selected from the Units/Mode menu, Level,
Range, Flow, HRT, and Volume rate being measured, TOTALIZER and RELAY
states. The SLT 5.0 will start-up with this display and will return to this screen
after a timeout if keys are not pressed in other menus.
MESSAGE
Press from the RUN display to view status of the data logger and error/warning
messages provided by the instrument. The word Message will appear on the RUN
display if error messages are being generated by the instrument. Refer to the
manual section Error/Warning Messages for a description. Press to return to the
main display.
STATUS
Press from the RUN display to view instrument status. Range will be displayed
in linear units.
Tot Displays the current totalizer reading.
EC Displays echo confidence.
Relays Energized relays will display as a white character on a black
background.
24 HR LOG (Data Logging option only)
Press from the RUN display to view a formatted flow report from instruments
with a built-in data logger. Press to scroll down one day or repeatedly to scroll
to a specific date. Up to 365 days can be stored. Newest date will overwrite the
oldest. Press to return to the main display.

Page 9
SLT 5.0 Level and Flow Monitor
PASSWORD
The password (a number from 0000 to 9999) prevents unauthorized access to the
Calibration menu.
From the Run display press the key to get to Password. Factory default
password is 0000 and if it has not been changed press the to proceed to the
Menu Selections screen.
If a password is required, press to place the cursor under the first digit and or
to set the number, then to the second digit, etc. Press or to proceed to
the Menu Selections screen.
A new password can be stored by going to Special Functions/New Password.

Page 10
SLT 5.0 Level and Flow Monitor
UNITS/MODE
From Mode press the and then the or to select Level,Range, Flow,
HRT or Volume.
Range mode displays distance from the sensor to the target or liquid surface
like a tape measure. Range mode is useful to measure the exact
distance from the sensor to the zero level during calibration, or to
monitor "outage" or free space in a tank.
Level mode can be used to measure tank level in linear units, or "Head"
(height of flow) in an open channel for comparison with a flume
manufacturer's flow tables.
Volume Volume mode displays tank inventory in engineering units like
gallons or liters.
HRT Horizontal Round Tank mode sets the SLT 5.0 to calculate and
display volume units in a horizontal round tank.
Flow mode is for open channel flow through a flume or weir.
From Volume HRT or Flow press to make your selection. (Range or
Level mode will bypass the Volume units selection menu). Volume Hrt or Flow
modes will give you the additional choice of volumetric units:
ft3 - cubic feet
USG - US gallons
USMG - US million gallons (FLOW only)
IG - Imperial gallons
IMG - Imperial million gallons (FLOW only)
m3 - cubic meters
L - liters
bbl - U.S. oil barrel
Press the to store your selection then the to the next menu item and to
enter.
Linear
press the key and then the or to select your units
of measurement. Press the to store your selection.
Time
press and then or to select day, hr, min or
sec.
Temperature
press and then or to select Cor F
(Centigrade or Fahrenheit).
Press or to return to the Menu Selections screen.
Table des matières