Gate EASY HighBack Manuel utilisateur

A DIRECT HEALTHCARE GROUP COMPANY
GATERD.SE
EASY HighBack
Teknisk manual

EASY HIGHBACK
GATERD.SE2

TECHNICAL MANUAL
3
GATERD.SE
Contents
1. Parts list 04
2. Spare parts 04
3. Accessories 06
4. Cleaning and maintenance 07
5. Replacement of remote control 09
6. Calibration of actuators 09
7. Reset service indicator 09
8. Troubleshooting when chair cannot be operated 10
9. Safety instructions 10
10. ID label 11
11. Warranty 11
12. Technical specication 12
13. Information on EMC 14
14. Further information om EMC, advanced 14
15. Dimensions 17
EASY HighBack
EASY in the high-back version (HB) is a development of the conventional concept using the back strap/neck
support already available. This hygiene chair is tted with a high, padded, water-repellent back complete
with neck support (available in two versions). The high backrest provides soft yet stable support and its lower
section is tted with a pliant lumbar support for additional comfort.
EASY HB is designed with telescoping pillars to create as much space and exibility as possible for care
personnel in the performance of their work.
EASY is available with our new, soft Comfort seat with an opening at the rear for a more natural, hygienic
working method.

EASY HIGHBACK
GATERD.SE4
1. Detaljförteckning
1a 58-248-M Headrest M
1a 58-248-L Headrest L
1b 58-247-FLEX-1 Back/Foam-Flex
258-295-1 Backrest HB
357-288-2 Armrest lock
458-294-L/R Armrest HB (left/right), alloy only
558-209 Handrest
658-284-1 Seat frame HB
757-250-1 Plastic seat
756-233 Plastics/keyhole, green
858-280-L/R Pillars, complete (left/right)
958-277 Holder/footrest
10 58-283-1-L/R Footrest tube (left/right)
11 58-296-1 Footrest
12 58-281-1 Bottom frame
13 57-049-RF Wheel with bolt - fully braked
13 57-040-RF Wheel with bolt - directional lock and fully braked
14 58-360 Cover
15 57-793 End plug
16 58-253 Battery Pb 5,0 Ah
17 58-291 Cover, control box
18 58-293 Bellows (silicone)
19 58-282 Cap/tilt actuator
20 58-252 Actuator TA16
21 58-290 Cover plate, left
21 58-289 Cover plate, right, incl. holder tilt actuator
22 57-235 Cover/armrest
23 58-243-1 Connectors
2. Reservdelar
Part no. Description
58-255 Charger, universal volt
58-255-1/2/3/4 Cable for 58-255 (EU; UK; US; AUS)
58-254 Remote control for Easy

TECHNICAL MANUAL
5
GATERD.SE

EASY HIGHBACK
GATERD.SE6
3. Tillbehör
Part no. Description
56-205 Shower seat around hole
56-206 Soft PU seat
56-212-EASY Barrier (52cm), Easy
56-213 Splashback
56-214 Armrest lock (M8)
56-216 Inll
56-217 Waist belt/thigh belt
56-218 Slides
56-220-EASY Heel strap kit (pair), Easy
56-221-HB Width kit (+50mm) Easy HB
56-227-L Calf rest, left
56-227-R Calf rest, right
56-229 Chest belt
56-232 Plug, soft seat
56-234 Comfort seat/back opening, green, Easy
56-260-L Footrest, left
56-260-R Footrest, right

TECHNICAL MANUAL
7
GATERD.SE
4. Cleaning and maintenance
The chair is made of stainless steel pipes (powder coated) and aluminium. Exposed details that have not been
painted are made of aluminium or stainless steel. Be sure to check that the product feels stable / slip-free - no
loose screw joints may be present. Damaged or defective parts should be replaced as soon as possible.
After performing maintenance below, reset the service indicator (see page 7).
Cleaning, general
• Clean the chair by wiping with a common detergent (NOT based on Pentanol) or washing machine / auto-
clave (max 85 degrees).
NOTE: Remote control / control box or coupling (female) (if remote control is removed before washing) must be
protected from high pressure water. This can be done with a plastic bag or the like that can withstand the wa-
ter pressure of the washing machine. Place remote control / control box or connector (female) so that running
water cannot penetrate.
• Seat - can be removed and cleaned separately or washed in machine / autoclave (max 85 degrees).
• The pillars - cleaned with commonly used cleaning / disinfectants in health care (alcohol-based). The pillars
should be cleaned with agents that partly dissolve fat and partly remove lime stains. When cleaning the
pillars, the chair is set at maximum height. Apply detergent and allow to work as directed (make sure the
agent also ows into the pillar), rinse.
NOTE: At regular intervals, the pillar bottom plate should also be cleaned - this is done by removing the bottom
plug (see Figure 18) and cleaning agent applied directly to the pillar. Let water work / rinse out. Mechanical
machining may be needed.
NOTE: To avoid pinch damage - be sure to always reposition the protective cap.

EASY HIGHBACK
GATERD.SE8
Check thoroughly (at least once a year)
1. Armrests - so that the slip guard / wrist guard is intact and xed.
2. Pillars - Calibrate the chair and then check its functionality with the remote control. Make sure the pillars
(up / down) are synchronous and feel loose. Also, check the bellows so that they are whole - these must
be replaced if they are damaged.
3. The seat / soft seat - is complete and without cracks. No sharp edges may be present.
4. Footrest plates - are whole and possible to adjust in the intended way. Make sure clips are fully and
securely fastened. The heel straps should be located on the back of each footrest plate.
5. Wheels - Do not roll slowly or sit loose. Also check that the brake is working.
6. Back / Headrest should be complete and have no holes or other damage to the straps / locking brackets.
1
6
2
3
4
5

TECHNICAL MANUAL
9
GATERD.SE
5. Replacement of remote control
Incorrect remote control is easily changed by releasing the contact / remote control on the left pillar - new
remote control is mounted in the same way.
NOTE: It is important that contact is locked with the clamp lock to ensure IP X5.
6. Calibration of actuators
The calibration of the actuators (height / lowering only) is normally done fully automatically each time the chair
reaches its end positions (min height, max tilt).
If the chair still does not go in parallel or if the tilt function does not seem to work, an ACTIVE calibration can be
done. This is done by holding down <button-slow / up> + <button-slow / down> simultaneously.
After about 15 seconds the chair will be moved to its end positions ie. my height and maximum slope
backwards. Hold down the buttons until the end positions are reached. Release the buttons and wait 5
seconds. The chair is now calibrated and can be used again.
Calibration should NOT be done with user in chair. Note that all 3 actuators move during the calibration
phase. Otherwise, cancel the calibration immediately and contact Customer Service / GATE (or your
local dealer).
7. Reset service indicator
The remote control has a service indicator for periodic maintenance reminder (min once a year). The service
indicator measures voltage and is activated after 11 months (ashes every 4 seconds). After 12 months the
indicator will glow steadily.
The indicator is reset by holding down buttons on the remote control (top line, <fast up>, <fast down>) for 30
seconds.

EASY HIGHBACK
GATERD.SE10
8. Troubleshooting when chair cannot be operated
NOTE: Chair cannot be operated over seat height 65cm if chair is tilted forward. Seat can also not be tilted
forward if seat height is higher than 65cm.
Before starting troubleshooting - always restart the battery (see below - hold the battery OFF)
button for about 5 seconds. Wait a minute. Activate the battery through its ON button).
1. Check the battery charge via the re-
mote control’s battery indicator (acti-
vate the remote control if necessary by
pressing any button).
2 Switch off the battery manually using
the Battery Off button (see below).
Wait a minute. To activate the battery,
press its On button (see below).
Make sure the battery indicator light is
on or ashing.
1a. Indicator shows green ==> Calibrate
actuator (see page 7).
The indicator light is on / ashing ==>
Replace the remote control and charge
the battery as instructed.
1b. Indicator shows red ==> Charge the
battery.
The indicator light is not on / ashing
==> Change battery.
1c. Indicator is off ==> Connect the
charger to the remote control. If charg-
ing is indicated in the remote control,
charge the battery as instructed.
If charging is not indicated - continue
to 2.
If the above has been done and the chair cannot be operated - contact Customer Service / GATE.
9. Safety instructions
Shower chair should only be used:
• As a shower / hygiene tool
• Indoors, on at and hard surfaces
Table des matières
Autres manuels Gate Fauteuil roulant