Instructions for use: FlexPulse G2 Item No.: 451600-0512. Version: 01
Page 3
Operating the FlexPulse G2
The battery of new FlexPulse G2 unit needs to be charged before first use. Please read the
chapter “charging the battery” on page 5.
Connecting the magnetic field coil
Connect the cable(s) of one or two coils to the FlexPulse G2. Make sure that the plugs line up
with the ports from the unit. It may be best to line up the bottom edge of the coil plug with the
bottom edge of the port and then “hinge” it into the port before fully inserting the plug.
Switching the unit ON
Push the ON/OFF button .
The program, which was used during the previous session, is shown on the display. The intensity
value is flashing. If no program is running and no key is pressed for 120 seconds, the device
turns off automatically.
Selecting a program
Select a program by pressing the key. To scroll through the programs,
press the key as often as necessary until you have selected the program
of your choice. Programs may be reset during operation by pressing the
key again, but the timer will have to be reset too. The programs are: P01-
3Hz, P02-7.83Hz, P03-10Hz, P04-14Hz, P05-23Hz, P06-40Hz, P7-50Hz, P8-
60Hz, P9-10Hz/100Hz (alternating every minute), P10-999 (=1000) Hz.
Setting the intensity
Increase and decrease the intensity with left and keys. The intensity
is displayed on the screen. The default is 10. The intensity can be set
between level 1 and level 10. By keeping the intensity keys pressed, the
intensity will increase/decrease more rapidly.
Note: Duration of battery operation depends on intensity. Higher intensities will drain the battery
faster, especially with the higher frequency programs.
Setting the program time
All 10 programs have a 30 minute default time. To change the program time,
use the right / keys. You can change it from 10 min to 20 min, 30 min,
40 min, 50 min, 1 h, 2 h, 4 h, 6 h or continually. In continuous mode the
display will show “cont”.
Starting the program
Push the key to start the program. The intensity display will stop flashing. If the program is
not running, the intensity value is flashing. When the stimulation time is completed the
stimulation terminates automatically and the device will beep three times.
Stopping the program
You can stop the program by pressing the key for 4 seconds. The unit will reset to the default