EYELA KCL-2000W Manuel utilisateur

Environmental Chamber
R00
取扱説明書
KCL-2000W
This manual is designed to use this unit safely with the best performance.
Be sure to read “Safety precautions carefully
before use.
Please keep this manual in a place easily accessible to every user.
Important
Tokyo Rikakikai Co., Ltd.
Instruction Manual

Safety precautions
Though we are trying to look into conceivable risk of using the product, it is very
difficult for us to expect all. It means that all the instructions in this manual do not
cover all the potential risks that may be caused by the product. However, if you follow
the instructions, you surely can handle and operate the products safely. Please use
extreme care when handling this product and try to prevent all the potential accidents
and mechanical failures.
Alert Mark
Signal word Definition
1. Signal word for warning
Any flammable material such as organic
solvent cannot be contained in this unit.
Also, due to its functions and characteristics,
some parts of this unit subject to high
temperature. If you touch them carelessly,
you may get burned unexpectedly.
This manual shows precautions for your
safety to prevent careless injuries.
They are classified and defined according to
their risk, and indicated with an alert mark
and a signal word.
Please follow these instructions.
Warning Mishandling the product may cause users serious personal
injury or loss of life.
Caution Mishandling the product may injure users or cause
property damage.
Dangerous Mishandling the product will cause users serious personal
injury or loss of life.

2. Warning display on the product
For highly priority danger, warning label is attached
on the machine body. The position of the label is as below.
Please read the instructions carefully whenever using the machine.
* If you have any trouble reading the label because of
damage and etc., please replace with new one.
Please contact us if you require new label.
Do not touch the
cooling fin with
your bare hands.
Edgy fin may cause
you physical unjury.
Caution Dangerous
Do not put any
explosive
materials such as
organic solvent .

Check the type and quantity of each part.
Contents
Items contained in
your packing
1. For safety use ........................................
2. Outline
2-1 Application ............................................
2-2 Specifications ........................................
2-3 Heating and cooling data .......................
2-4 Optional accessories ...............................
2-5 Name of each part....................................
3. Name and functions of the control panel
3-1 Control panel ...........................................
3-2 Operation keys and operating key indicator ...
3-3 LCD panel and LED indicator ...............
3-4 Overview of operation .....……................
3-5 Safety and alarm functions …................
4. Installation
4-1 Installed environment...............................
4-2 Environmental conditions .......................
4-3 Installation ...............................................
4-4 Installation .........................................…..
4-5 Utility connection ...............................
1
1
2
5
6
7
8
9
11
12
15
18
18
19
20
21
Shelf
Contents KCL-
2000W
2
3
1
4
6
Main unit
7
Model
1
2
1
1
Shelf
Bracket Right
Silicon stopper
1
8
External water tank
5Plate for fixing rack for external
water tank
Rack for external water tank
9
1
This instruction manual describes the procedure
of installation, operation, troubleshooting,
maintenance / check-up, and disposal for
Environmental chamber
KCL-2000W.
Please read this manual carefully before use.
Instruction manual
10
11
12
14
16
17
1
5
1
Wet-bulb wick
Drain hose
Dropper
15
Grounding adapter
Washing brush 1
Screw
Water-supply hose set
Water-supply hose
Alarm output connector
13
Leaflet on panel key operation
18
1
1
1
1
Introduction
Thank you for purchase of
Besides this instruction manual, we have
enclosed the leaflet on panel key operation,
which explains the procedure for making
operation program with panel key.
Please refer to it if needed.
5. Operation
5-1 Preparation ....................................
5-2 Operation ................................…...
6. Troubleshooting ....................................
7. Maintenance and check-up
7-1 Operation test for residual correct device....................
7-2 Cleaning and caring the product......
8. Disposal of unit ....................................
9. After-sales service ...............................
22
31
35
42
42
46
46
Bracket
Left Right
Connect to ground
Grounding adaptor
100V only
2
2
Contents KCL-
2000W
Model
1
2
Left
• Shelves and brackets are fixed to
the chamber with adhesive tape
19
20
CD-ROM Communication
software
1
1
1
RS-232C communication cable

- 1 -
1
This unit is not designed with explosion-proof structure.
Use extreme care when handling it.
For safety use
Do not put any inflammable materials such as
organic solvent into the chamber.
While operating the unit in high temperature,
sample in the chamber may vaporize, catch
fire and explode. Inflammable materials are
nitrates, nitro compounds, etc. and explosive
materials are chloride peroxides, inorganic
peroxides, salt nitrates, organic solvents etc.
This unit is not designed with explosion-
proof structure.
Dangerous
Caution
Do not touch the cooling fin with
your bare hands.
The cooling fin is very sharp.
Do not touch it with your bare hand on the
occasion of maintenance of the unit to avoid
injury of your hand.
2
2-1 Use application
Outline of the product
Warning
Do not remodel the product.
Do not use it out of specified applications.
Remodeling or using the product out of specified
applications may cause electric shock or malfunction.
This unit is an environmental chamber used
for various tests such as humidity test and
insulation performance test of electric parts,
corrosion test of metal parts or drugs and
medicines field etc under controlled
temperature and humidity.
Cooling fin

- 2 -
2-2 Specifications
Product
Model
Air flow system
Dry-bulb temperature sensor
Heater
Environmental chamber
Forced air flow
-15~+85℃/25~98%RH Minimum unit 0.1℃/0.1%RH
P.I.D. control by a microprocessor
Platinum temperature measuring resistor Pt100Ω
Digital setting through the membrane switch
KCL-2000W
・Normal mode:Automatic control by microprocessor (Interval defrosting mode)
・Manual mode (Manual start, Automatic stop/Manual stop)
1.Auto-start program (1 min.~99 days 23hr. 59min.)
2.Auto-stop program (1 min.~99 days 23hr. 59min.)
3.User’s program (7 patterns)
・1 pattern:contains Max. 10 segments, 1 min.~99 days 23hr. 59min./1 segment
・Repetition time of program:1~999 times or endless
・Program type:Target precedence control of temperature and humidity, Step control,
Gradient control
Stainless steel sheath heater 650W(SUS304)
Refrigeration unit, Coolant Air-cooled type 200W Coolant : R-134a
Within ±0.5℃/±3%RH *2
Within 4℃/12%RH *2
Digital readout Minimum unit 0.1℃/0.1%RH
・Set/Measured temperature, Set/Measured relative humidity, Total time (1 min.~999 days
23 hrs. and 59 min.)
・Graphic display (Switch-over)
Residual current device and excess current breaker, Door switch
Stand-alone over heat protector (Variable type +30~+93℃)
Stand-alone over heat protector for boil-dry protection
Self-diagnosis system of temperature controller
(Dry-bulb temperature sensor failure, Wet-bulb temperature sensor failure, Heater fault,
Refrigeration unit fault (High pressure, Overload), Temperature curve failure, Power failure,
Overheat, Overcool, Cooling failure, Level sensor fault, Humidity control failure, Humidity
upper limit fault, Water supply failure, Low level of water supply tank, Door open, Watch dog,
SSR fault, Full drain tank (when using an optional drain tank))
・RS-232C interface
・Recorder output terminal for temperature and humidity
・Alarm output terminal (Contact voltage type)
・Calibration of temperature indication ・Maintenance mode
・External water supply ・Forced draining ・Input terminal for drain tank alarm
Wet-bulb temperature sensor Platinum temperature measuring resistor Pt100Ω
Heater for humidifying Stainless steel sheath heater 200W×2(SUS316)
Interior Antibacterial stainless steel sheet Others: SUS304
Temperature controller P.I.D. control by microprocessor, Noncontact zero cross
Chamber capacity
Shelf
Shelf size
Chamber dimensions *3 Approx. 140L
500W×400D×700H (mm)
Max.15kg/shelf under uniform loading Supplied shelves : 2 pce. as standard with 2 brackets
468W×375D (mm)
Pitch of bracket holder /
No. of columns 40mm pitch 5 levels/2 columns
Door With observation window (without inner door)・Quadplex glass・
Equipped with heater for reducing dew condensation, no double door
Temp./Humidity
control range *1
Temp./Humidity
control accuracy *1
Temp./Humidity
uniformity *1
Temp./Humidity control
Temp./Humidity setting
Temp./Humidity indication
Defrosting system *2
Program
Other functions
Safety features
Overload relay holding circuit, Protection timer for ref. unit, High pressure switch for ref. unit.
Cable port 2 ports at both sides of unit ID40mm
Supplying water Distilled water or ion-exchange water through a polyethylene tank or a cartridge water purifier
Electric conductivity : 10µs/cm or less
Tank capacity Inside : Approx. 10L (Removable) Outside : Approx. 20L (connected by one-touch socket)

- 3 -
Model
Overall dimensions *3
Power input
Rated power source
Net weight *4
5 ~35℃
15A 1.5kVA
AC100V 50/60Hz
Range of available temperature
635W×755D×1695H mm
153㎏
KCL-2000W
Recorder output Measured temperature : DC1mV/ ℃Measured humidity : DC1mV1%RH
Alarm output Contact capacity AC250V 5A Normal : OFF Abnormal : ON
Drain port Nozzle OD 11×ID7 mm
*1
*2
*3
*4
*
The above performance has been obtained under conditions of 20℃of room temperature, rated power
source voltage, and without load.
The performance is indicated in conformity with JTM KO1-1998 (Testing method and indication form
for constant temperature and humidity bath by Japan Testing Machinery Association).
If the unit is operated at 30~40℃or lower temperature for a long time (one day~one week or longer),
please set “interval defrosting operation” for the operation.
The chamber temperature will go up by about 3℃, and the chamber humidity increase by about
20%RH, when the defrosting mode works under room temperature 20℃.
Actual changes of such temperature and humidity differ depend on actual operating conditions.
The chamber dimensions and overall dimensions without external tank do not include any projections.
The net weight includes external tank part and rack frame.
If the rack is not used, the net weight is about 127kg.
The required time to heat and cool, and to humidify, temperature and humidity control accuracy, and
temperature and humidity uniformity doffer depend on the environmental temperature, power source
voltage, with load or without load.

- 4 -
Humidification control range
0 20 40 60 80 100
0
20
40
60
80
100
Temperature (℃)
Relative humidity (%) RH
(1) 20℃room temperature, unloaded condition
*1
*2 The accuracy of temperature control differs depends on the environmental conditions such as load condition.
If the unit is operated at 30~40℃or lower temperature for a long time (one day~one week or longer), interval
defrosting operation must work. The temperature in the chamber is raised about 3℃, and the humidity in the
chamber is raised about 20%RH, when the defrosting mode works under 20℃of room temperature.
These variation differs depends on actual operating conditions.

- 5 -
2-3 Heating and cooling data
・AC-100V 50Hz
・Unloaded condition
Note; The maximum temperature and required time to heating differs depend on the
environmental temperature, power source voltage, with load or without load.
-20
-10
0
10
0 10 20 30 40 50 60 70 80 90
20
35℃RT
20℃RT
30
40
50
60
70
80
0 5 10 15 25 50
Time (min.)
Temp. (℃)
30 35 40 45
90
20
20
* Without humidification control
(1) 20℃room temperature, Cooling curve (85→20℃)
(2) 20/35℃room temperature, Cooling curve
Temp. (℃)
* Without humidification control
Time (min.)

- 6 -
2-4 Optional accessories
●Drain tank
Description
Cat. N0.
Drain tank
204960
30
40
50
60
70
80
0 10 30 40 60 70
90
20
20
10
0
-10
-20
50
(3) 20
℃
room temperature
Time (min.)
Temp. (℃)
* Without humidification control
(Refrigeration unit is OFF.)
If 10L capacity drain tank is full (option), the
unit indicates alarm message and stops
temperature and humidity control.
The tank cannot be used when the unit is placed
on the floor without using a rack.
Use the drain tank in a place where is 310mm
lower than the unit.