S.Cod.
d.C.sc.
Uo.sc.
D.Ca.s.
S.Cod.
a.CAL.
1.CA.
10.CA.
If key is pressed and held for 5
seconds programming mode is entered
U.opt. d.Cal. SECU.
d.pnt.
Cal.t.
l.sCl.
H.sCl.
l.inp.
H.inp.
CAL.t.=U.ınP.
Yes
No
d.Cnf.
i.typ.
ratE
Hold
dsp.C.
Unit
If first and then keys are
pressed together , measurement
unit appears. See Unit parameter
for programming.
If first and then keys are
pressed together, the maximum and the
minimum measurement values become
equal to the measured value at that time
and the message res. appears on
display.
Pressing this key the minimum
measured value up to that time
appears.
Pressing this key the maximum
measured value up to that time
appears
The minimum measured
value
The maximum
measured value
Measurement unit 870 1453
Bar.
&
Programming Mode
3/3
Running Mode
Measurement
value
1000
Res.
EI141-E-09-160217
SET
SET
SET
SET SET
SET
SET
CAL.t. = Calibration type.
Selectable as S.inP. or .
If S.inp. is selected, input type is one the
four standard input types. If U.inP is
selected, input types can be modified.
See NOTE 1 for modification.
U.inP
i.typ. = Input type.
Input type can be selected as 0-20mA,
4-20mA, 0-1V or 0-10V.
See NOTE 1 for programming.
rAte = Sampling rate.
Measurement is performed at each 200ms.
However,
for FASt rate, each measurement is displayed.
for SLo.1 rate, the average of 4 successive
measurements is displayed.
for SLo.2 rate, the average of 8 successive
measurements is displayed.
for SLo.3 rate, the average of 16 successive
measurements is displayed.
See NOTE 1 for programming.
Hold = Display holding parameter.
Selecting NonE , this parameter becomes
inactive.
Selecting Lo. , always the minimum
measured value is displayed.
Selecting Hi. , always the maximum
measured value is displayed.
dSP.C. = Display configuration.
Selectable as PrcS. or Pr.Un. . If PrcS. is
selected, process value appears. If Pr.Un.
is selected, process value and then
measurement unit are displayed 4 and 2
seconds successively.
See NOTE 1 for programming.
Unit = Measurement unit.
A constant, a message etc. to be displayed
can be entered. If a decimal point is desired,
it should be included before entering the
character.
For including decimal point first, then,
keys are pressed and held together. And
then, by using key decimal point can be
adjusted.
d.Pnt. = Decimal point.
Decimal point can be adjusted between
1. and 3. digits.
See NOTE 1 for programming.
L.SCL. = Lower limit for the scale.
It can be adjusted between -1999 and
(H.SCL. -100).
See NOTE 1 for programming.
H.SCL. = Upper limit for scale.
It can be adjusted between (L.SCL. +100)
and 4000.
See NOTE 1 for programming.
SET
At this state, the reference voltage or
current that corresponds to L.SCL. Parameter
is applied to the input.
To initialize calibration, first then
keys are pressed together and held until ‘CAL'
message appears. See NOTE 2.
SET
At this state, the reference voltage or
current that corresponds to H.SCL.
parameter is applied to the input.
To initialize calibration, first then
keys are pressed together and held until
‘CAL' message appears. See NOTE 2.
SET
SET
SET
s.Cod. = Access code for calibration menu.
This parameter should be 222.
See NOTE 1 for programming.
A.CAL. = Current calibration.
At this state, 20 mA current is applied to the
input of the device.
10.CA. = 10V
At this state, 10V is applied to the input of
the device.
input calibration.
1.CA. = 1V input calibration.
At this state, 1V is applied to the input of
the device.
For initializing calibration, first then
keys are pressed together and held until ‘CAL'
message appears. See NOTE 3.
To initialize calibration, first then
keys are pressed together and held until ‘CAL'
message appears. See NOTE 3.
To initialize calibration, first then
keys are pressed together and held until ‘CAL'
message appears. See NOTE 3.
s.Cod. = Access code for safety menu.
This parameter should be 333.
See NOTE 1 for programming.
d.C.Sc. = ConF. menu protection level
parameter.
nonE = No menu is seen.
P. no = Menu is seen but can not be programmed.
P.yES. = Menu is seen and programming is possible.
See NOTE 1 for programming.
Uo.Sc. = U.oPt. menu protection level
parameter.
nonE = No menu is seen.
P. no = Menu is seen but can not be programmed.
P.yES. = Menu is seen and programming is possible.
See NOTE 1 for programming.
d.Ca.S. = d.CAL. menu protection level
parameter.
nonE = No menu is seen.
P. no = Menu is seen but can not be programmed.
P.yES. = Menu is seen and programming is possible.
See NOTE 1 for programming.
CAL. CAL.
C.End C.End
s.Err.
S.Err.
C.Err.
C.Err.
The message on the left flashes approximately 5
seconds and calibration is completed.
The message on the left flashes approximately 5
seconds and calibration is completed.
If calibration is success, the C.End
message appears for 1 second.
If calibration is unsuccess, the S.Err
message appears for 1 second and
passed to the next step.
.
If the difference between the reference voltages or currents applied for the calibration of H.inP. and L.inP. is lower than one
half of the full scale, this error message appears on the display.
For example: Assume that the selected input type is 0-1V. In this case, if the difference between the reference voltages
applied for calibration of H.inP. and L.inP. is lower than 0.5V, this error message appears.
If the reference voltage or current applied to the input for calibration is too high or too low, this error message appears.
ERROR MESSAGES
CAL okey?CAL okey?
Yes Yes
No No
NOTE 2 NOTE 3
SET
For adjusting a selected parameter first press and hold key. Then, by using keys adjustment
can be made.
Parameter adjustment method
NOTE 1
If increment key is pressed and held 0.6 seconds, the value of the selected parameter changes rapidly. If
waited enough, the value increases 100 at each step. After 1 second following the release of the key, initial
condition is returned. The same procedure is valid for the decrement key.
Entering from programming mode to run mode:
If no key is pressed within 20 seconds during programming mode, data is stored automatically and the run mode is entered.
Alternatively the same function occurs by pressing key and holding for 5 seconds after pressing key and
entering main menu.
SET
If calibration is success, the C.End
message appears for 1 second.
If calibration is unsuccess, the .Err
message appears for 1 second and
passed to the next step.
C.