CONFIGURATION: “SEL.CFG” Use Gross/Net to enter the menu and step to each
category, Tare Recall to select parameters. ENT to return to menu selection.
Capacity is the combination of “1”, “2” and “3”.
Step Parameters Definition
1 5, 10, 15, 20…100, 120…1000 Number of divisions x100 ,100 = 10,000
divisions
2 1, 2, 5, 10, 20, 50, and 100. Count by selection10,000 divisions, count
by 2 = 20,000
3 0, 0.0, 0.00, 0.000, and 0.0000 Decimal point selection
4 105P, 9 d (105% or 9 divisions) Overrange selection
5 1, 2, 3, 4, 5, 6, 8, 10, 12, 15……90 Digital filter selection (averaging)
6 off, 0.5, 1, 3, 5, 10 (divisions) Auto Zero Maintenance (AZM)
7 1.9, 5, 10, 20, FS (% of capacity) Zero range selection:1.9% of 2,000 x 0.2 =
38.0 lb
7.1 off, on (ISM) Zero’s scale on power-up
8 off, 1, 3, 5,10 (divisions) Motion Band selection
9 lb, kg, con Units selection and convert
10 nt, Gtn, n.nt, n.Gtn. Port 1 serial output selection: nt display
only, Gtn is Gross Tare Net and n.nt/n.Gtn
inhibit negative gross printing
11 off, co, de Off, Continuous, or Demand
12 7o, 7E, 8n 7- odd, 7- even or 8- none
13 12, 24, 48, 96 Baud rate selection
14 off, 1, 2, 3, 5, 10, 15 (seconds) Delay between lines or continuous output.
19 A, b A : adds “STX” in continuous, b : No “STX”
in continuous
20 nt, Gtn, n.nt, n.Gtn Port 2 serial output selection
21 off, co, de, Ln Off, Continuous, Demand, Network
22 7o, 7E, 8n 7- odd, 7- even or 8- none
23 12, 24, 48, 96 Baud rate selection
24 off, 1, 2, 3, 5, 10, 15 (seconds) Delay between lines or continuous output.
25 1 – 16 (RS485/RS422) Network address selection
30 off, on DIO Inputs
9.1 Of, At, Tr, On Off, Auto Tare (At), Key Tare (Tr), Both