Emery Winslow 7600E Guide rapide

1
Configuration and Setup
V4.31
Model 7600E
Digital Weight Display

2
TABLE OF CONTENTS
SPECIFICATIONS....................................................4
DIMENSIONS……………………………………………5
INSTALLATION.........................................................6
SETUP ACCESS ......................................................7
MENU LAYOUT........................................................8
CONFIGURATION....................................................9
CALIBRATION........................................................11
REMOTE SERIAL DISPLAY...................................12
SERIAL PORTS ....................................................14
SERIAL COMMANDS.............................................15
ANALOG OUTPUT .................................................17
DIGITAL INPUT/OUTPUT.......................................18
SETPOINTS, BATCHING .......................................19
SERIAL COMMUNICATION (Setpoint/Accumulator).38
TIME & DATE..........................................................39
WEIGH IN/OUT.......................................................40
TEN POINT LINEARIZATION ……………………..43
TOTALIZER (Accumulator) ……………………….…43
SMART SERIAL......................................................44
DISPLAY MESSAGES............................................52
115/220 VAC...........................................................53
SPARE PARTS.......................................................54
DIGITAL CONFIGURATION RECORD...................55

3
Alternate Packaging for
extreme wash-down and with
Totalizer in common enclosure
Emery Winslow Scale Company
73 Cogwheel Lane 4530 North 25th Street
Seymour, CT 06483 Terre Haute, IN 47805
(203) 881-9333 (812) 466-5265
(203) 881-9477 FAX (812) 466-1046 FAX

4
SPECIFICATIONS
LOAD CELL A/D CONVERTER
TYPE: 24 bit delta sigma (1:16,777,216)
EXCITATION: 5 VDC, 120 mA max.
SIGNAL INPUT: 16 mv
SENSITIVITY: 0.1 uV/grad
UPDATE RATE: 30 update/second
DISPLAY: Six (6) Decades, 0.6 inch LED
KEYPAD: Full numeric plus controls
POWER INPUT: 117/217 VAC, 50‑60 HZ, 20 watts, fuse 0.50 A Slo-Blow.
SERIAL PORTS:
Port 1: RS232C or 20ma
Port 2: RS485, RS232C or 20mA.
Smart Serial Setup: 8 custom print files plus 8 macro files, 30 characters each.
Batch Start/Stop: Control from front panel or remote input.
Setpoint Operation: 4 output relays configurable for normal setpoints,
over/under or manual/auto batch modes.
TIME & DATE: 12/24 hr, battery backed.
ENCLOSURE: Stainless Steel, NEMA 4X, Tilt-Stand Base, 7lbs.
CASE: 8.82” (w) x 6.31” (h) x 4” (d) Tilt or panel mount.
NRTL: Met Labs file E112367 certified to UL60950-1, CAN/CSA-C22.2 60950-1-03
NTEP: Class III/IIIL, 10,000 divisions CoC 97-009A4
OPTIONS:
ANALOG OUTPUT: 0-10v, 4-20ma (16 bit D/A).
DIO: 4 AC/DC –inputs, 4 AC/DC outputs (SS Relays, 0.5 amp)
ETHERNET TCP/IP
DUAL CHANNEL

5
DIMENSIONS

6
CalSwitch SW-4
TB-1
———-1——–Signal +
———-2——–Signal -
———-3-——Sense -
———-4—— Excitation -
———-5——-Sense +
——-—6——-Excitation +
SenseJumpers
Option Header
Analog Output
A/DRFIShield
Software Version#
VX.XX
INSTALLATION
POWER WIRING: The indicator is designed to be operated from 117/217 VAC,
50‑60 Hz. The unit power cord must be plugged into a grounded 3 ‑wire polarized
AC wall socket. All normal wiring and grounding precautions should be observed,
including use of a "clean" AC power line.
SCALE WIRING: The unit is equipped with cable gland entry for load cell cable
insertion and internal (pluggable) terminal strip for 4 / 6 wire connection.
Remove sense jumpers P11-8/7, P11-6/5 for six wire.

7
SETUP ACCESS: To access instrument configuration, calibration or to enable op-
tions, depress the “ZERO” key for five seconds.
The Audit Trail counters (“Pxxxx” and “Cxxxx”) are displayed first followed by access
code request (“AC?”). The initial factory setting is “0000” which is entered with four
numeric “0”’s and enter (“ENT”). If no entry is made, instrument returns to operate
mode. The “Master” access code is “1042”.
The access code can be changed to any four digit combination during setup exit
when display again shows “AC?” or skip with ENT to continue with no change,
The “CHECK” key provides the software version “V X.XX” followed by the display
test routine. Use the “ENT” key to advance to the keyboard test and to exit tests.
After entry, use the “TARE RECALL” key to select a main menu; Configuration
(“SEL.CFG”), Calibration (“SEL.CAL” or “SEL.CL1”), or Options (“SEL.OPX”)
and “START/STOP” to back step.
The “GROSS/NET” key enters selected menu and is used to step through sub
categories. Individual parameter selection is made with the “TARE RECALL” key,
which then steps through the parameter choices (“ZERO” key back steps within the
menu).
The “ENT” key is used at any point to “back” up from categories to menus and to
“save?” and “AC?” and exit.
During the exit steps, if changes were made, the display is “save?” with alternate
“no”. To save changes, use the “UNITS” key to select “yes” and “ENT” to exit.
Calibration numeric entries are entered directly followed with the “ENT” key.
Front panel access is inhibited if conventional “sealing” is applied by setting jumper
J1-1 in the B position. The board mounted “CAL” button is then used for access.

8
MENU LAYOUT (SEL.XXX)
CFG
Configuration: Divisions, count by, decimal, over range, filter, AZM,
zero range, motion, lb/kg, tare select, serial port selection, digital
input enable, remote display, test mode, dual channel sum.
CL1
Calibration: Zero, Span (Earlier units have ‘CAL’ prompt)
OP1
Analog Output
OP2
Setpoint - Digital I/O
OP3
Time and date
OP4
Weigh-IN / Weigh-OUT Truck mode
OP5
10 point linearity, Filter select, Range select
OP6
Accumulator
OP7
Smart Serial
OP8
AC/DC Battery
OP9
Consult Factory

9
CONFIGURATION: “SEL.CFG”
1. Press and hold the ‘ZERO’ button for 10 seconds until a ‘P XX’ is displayed,
then release. [Note: Limited configuration parameters (non-metrological) can
be accessed by pressing and holding ‘PRINT/7’ key for 4 seconds.]
2. Wait for ‘C XX’ prompt, then the Access Code ‘AC ?’ prompt. Enter four (4)
zeros ‘0000’ using the numeric keypad, press ‘ENTER’.
3. Press the ‘TARE RECALL’ button repeatedly until ‘SEL.CFG’ is displayed.
4. Press the ‘GROSS/NET’ button to enter into the Configuration mode. (1_
XXXX will be displayed indicating you are viewing Parameter 1.)
5. Use the ‘GROSS/NET’ button to scroll through Parameters 1 through 50. (All
parameters may not be displayed, depending on setup.)
6. Change any given Parameter with the ‘TARE RECALL’ button. (Scrolling
function, selections wrap around if you have gone past the intended value.)
7. Once all Parameter values have been set, press the ‘ENT’ button to exit Edit
mode. (SEL.CFG’ will be displayed.)
•To exit and save: Press the ‘ENT’ button. ‘Save Y/N’ is displayed – press the
‘UNITS/4’ button to change the selection to ‘Y’, press ‘ENT’ to save, then ‘ENT’
again at the ‘AC ?’ prompt to exit, keeping the Access Code the same. (‘AC’
prompt is not displayed if no changes were made.)
Note: for firmware versions 4.29.0 and earlier, Capacity is the combination of “1”, “2”
and “3”. (Press the CHECK key to view version, ENTER 3 times to escape)
Parameter
Settings
Definition
1
5, 10, 15, 20...100, 120...1000
Number of divisions x100 ,100 =
10,000 divisions
2
1, 2, 5, 10, 20, 50, and 100.
Count by selection10,000 divisions,
count by 2 = 20,000
3
0, 0.0, 0.00, 0.000, and 0.0000
Decimal point selection
Firmware versions 4.30 and later:
Parameter
Settings
Definition
1
1000, 2000, 5000, 10000, etc
Scale Capacity
2
0.001, 0.002, 0.005 …100, 200, 500
Scale Resolution
3
N/A
N/A
4
105P, 9 d (105% or 9 divisions)
Overrange selection

10
5
1, 2, 3, 4, 5, 6, 8, 10, 12, 15...90
Digital filter selection
(averaging)
6
off, 0.5, 1, 3, 5, 10 (divisions)
Auto Zero Maintenance (AZM)
7
1.9, 5, 10, 20, FS (% of capacity)
Zero range selection:1.9% of 2,000 x
0.2 = 38.0 lb
7.1
off, on (ISM)
Zero’s scale on power-up
8
off, 1, 3, 5,10 (divisions)
Motion Band selection
9
lb, kg, con
Units selection and convert
9.1
Of, At, Tr, On
Off, Auto Tare (At), Key Tare (Tr), Both
10
nt, Gtn, n.nt, n.Gtn.
Port 1 serial output selection: nt display
only, Gtn is Gross Tare Net and
n.nt/n.Gtn inhibit negative gross printing
11
off, co, de
Off, Continuous, or Demand
12
7o, 7E, 8n
7- odd, 7- even or 8- none
13
12, 24, 48, 96, 19, 38
Baud rate selection
14
off, 1, 2, 3, 5, 10, 15 (seconds)
Delay between lines or continuous output.
19
A, b
A : adds “STX” in continuous, b : No
“STX” in continuous
20
nt, Gtn, n.nt, n.Gtn
Port 2 serial output selection
21
off, co, de, Ln, IP
Off, Continuous, Demand, Network
22
7o, 7E, 8n
7- odd, 7- even or 8- none
23
12, 24, 48, 96, 19, 38
Baud rate selection
24
off, 1, 2, 3, 5, 10, 15 (seconds)
Delay between lines or continuous output.
25
1 –32 (RS485/RS422)
Network address selection
30
off, on
DIO Inputs
40-46
See page 12
RSD functions
50
nor
Expanded Test resolution (X10)
51
no
Sum, dual channel
Autres manuels pour 7600E
1