4
Elastocon Manual – Cell Oven for Stress Relaxation EB 21, EB 22 and EB 21 LTHTP
2.2.2 Test name
Fromthemainscreen,taponthenameeldthat
correspondstotheovencellchannel(seegure2.2.2.a).
Type the test name from the touch screen keyboard (see
gure2.2.2.b). One test name row on the screen can hold up
to 40 characters.
1. Use the white arrow buttons on the upper left corner to
swapbetweendierentkeyboardcharacterlayouts.
2. Tap on the Enterbuttontonalizethetestnameinput.
3. To re-enter any character on the same test name row,
it is possible to correct any character in the test name with
a tap on any of the white left or right arrow buttons in the
lower left and right corner on the screen.
4. To cancel any input changes, tap on the ESC button and
return to the previous screen.
2.2.3 Power failure
Power failure during a test
If the test time is running when a power failure occurs, the
oven temperature and test time counters will continue from
the time when the power failure occurred, when the power is
resumed.
Power failure indicator
When the power is resumed after a power failure during
a test with temperature control at set point and test timer
running, the screen will lit up a power failure button.
Tap on the Power Failure button to see more information
andtoresetthepowerfailure,seegure2.2.3.a.
Anewscreenwillshow,seegure2.2.3.b.
This feature will give the operator information about
duration of the power failure and temperature deviation
caused by the power failure. The power failure button will
remain until the button Clear & Exit is tapped.
2.2.4 Set up screen (user)
Password
A password is needed to enter the set up mode.
Tap on the Setup button on the main screen to enter the
passwordscreenforsetup,seegure2.2.4.a.
Default password is: 1111
Figure 2.2.3.a – Power failure.
Figure 2.2.3b – Power failure.
Figure 2.2.4.a – Password.
Figure 2.2.2.a – Test name.
Figure 2.2.2.b – Test name.