DTE Lux E Manuel utilisateur


content
1. Introduction ....................................................................1
2. Principle and Application ..............................................1
3. Product Performance Structure and Components .....1
4. Basic Technical Specications .......................................3
5. Installation and Demounting ........................................3
6. Operation ........................................................................4
7. Light intensity measurement .........................................5
8. Precaution .......................................................................6
9. Contraindication ............................................................6
10. Maintenance .................................................................6
11. Troubleshooting ............................................................7
12. Packing list ...................................................................8
13. Transportation ..............................................................8
14. Storage and transportation .........................................8
15. After service ..................................................................9
16. European authorized representative...........................9
17. Symbol instruction .......................................................9
18. Environmental protection .........................................10
19. EMC - Declaration of conformity..............................10
20. Statement ....................................................................13

1
1. Introduction
Guilin Woodpecker Medical Instrument Co., Ltd. is a high-tech enterprise
in researching, developing, and producing dental equipment, and has
a perfect quality assurance system, main products including ultrasonic
scaler, curing light, micro motor, apex locator and ultrasurgery etc.
2. Principle and Application
2.1 LUX E adopts the principle of ray radiation to solidify the light-
sensitive resin by shooting at it in a short time.
2.2 This product is used for dentistry. It has the function of accelerating
dental restoration and solidifying the material of dental whitening.
3. Product Performance Structure and Components
LUX E curing light (dentistry) is mainly composed by high power LED,
optical ber, main unit and charge device.

2

3
4. Basic Technical Specications
4.1 Adapter:
a) Rechargeable Lithium battery:
Battery voltage and capacity: 3.7V, 1400mAh
Battery model: ICR18490
b) Adapter input: AC100~240V 50Hz/60Hz
Output: DC5V 1A
4.2 Applied part: Optical ber
4.3 Light source:
a) 5W high power blue light LED
b) Wave length: 420nm~480nm
c) Light intensity: 1600mW/cm² ~ 1800mW/cm²
4.4 Work condition:
Environment temperature: +5℃to +40℃
Relative humidity: 30%~75%
Atmosphere pressure: 70kPa to106kPa
4.5 Size: 40×50×260mm
4.6 Net weight: 147g
4.7 Power consume: ≤8W
4.8 Protection type against electrical shock: ClassⅡ
4.9 Protection degree against electrical shock: Type B
4.10 Protection against harmful ingress of water or particular
matter: ordinary equipment (IPX0)
4.11 Safety in the presence of ammable anesthetic mixture with
air, oxygen or nitrous oxide: not suitable under this condition.
5. Installation and Demounting
5.1 Take o the red cap from the optical ber, then insert the metal part
into the front of main unit (make sure to screw when installing, oblique
insert should be prevented).

4
5.2 Fix the light hood on the bottom of the optical ber.
5.3 Uninstall the LED, just reverse the procedure above.
5.4 Battery replacement method: open the battery cover of the main
unit, take the battery out, than disconnect the plug slightly. Connect the
plug of the new battery correctly, put the new battery in, and then x the
battery cover.
5.5 After using the machine or it is in need of charging, make the main
unit headon inserted in the pedestal and compress it to make sure the
main unit and pedestal chucking. When there is no need of charging,
please pull out the pedestal Adapter or put main unit into the pedestal
inversely.
6. Operation
6.1 Press the mode button to set the working mode, the corresponding
indicator will on when a mode set.
6.1.1 Full-power: the blue light radiates in full power.
6.1.2 Ramping: the blue light power increases stronger continually,
after 5 seconds reaches to the highest power.
6.1.3 Pulse: the blue light works on the pulse condition.
6.2 Press the time button to set the solidifying time, 4 working time is
available: 5, 10, 15, 20seconds.
6.3 When operating, aim the optical ber at the correct position, press the
power button to start or stop to emit of the blue light.
6.4 During operation, the blue light can be stopped by press the power
button at any time.
6.5 A battery detect circuit is xed inside the main unit, when low power
is detected, the indicator of the main unit will wink, please charge in
time.
6.6 Connect the Adapter and the charge device correctly, put the main

5
unit into the charge device, the yellow indicator will enlightened, when
charge finish, the green indicator will enlightened and the yellow one
will o. The green light is the power indicator of the charging nish. If a
battery error is detected, the indicator of the charge device will wink.
6.7 When operating nish, please clean the optical ber with calico to
avoid infecting the light intensity.
6.8 This equipment will turn o automatically if don't any action within
2 minutes, turn it on by press any button.
6.9 The eective light intensity of this equipment is much more higher
than Halogen Lamp, The solidied depth of the curing light composites
resin for 10 seconds will not less than 4mm.
6.10 This equipment has the function of light intensity measuring.
Connect the Adapter of the light intensity measure device correctly,
aim the top of optical ber at the sensor, turn on the main unit to emit
blue light, the green indicator of the light intensity measure device will
indicate the value of the light intensity.
6.11 The optical ber should be sterilized for 4 minutes with 134℃and
2.0bar~2.3bar (0.20MPa~0.23MPa) before each use.
6.12 The curing light is equipped with over-heat protection system. It can
continuously work 200s, For example, continuously operate the curing
light for 10 times under 20s working mode (even the curing light works
less than 20s, it is counted as a full operation), then it will come into
over-heat protection status. And only after 2-minute sleep, it can restart
working 200s continuously.
7. Light intensity measurement
7.1 Connect the output plug of power adapter to the plug of DC5.0V in
the pedestal.
7.2 Choose general mode and aim the optical ber at the measurement
point, press on / o button, the present light intensity is displayed on the
indicator of pedestal.

6
8. Precaution
8.1 Please recharge the battery at least 4 hours before rst time usage.
8.2 During operation, the light should be aimed straightly at the resin to
ensure the eect of solidication.
8.3 Avoid aiming the blue light at eyes directly.
8.4 Only the original pedestal charger, adapter and Lithium battery could
be used, because other brand pedestal charger, adapter and Lithium
battery are likely to damage the circuit.
8.5 It is forbidden to touch the charging connector with metal or other
conductor, to avoiding damage the circuit of charge or the battery.
8.6 Charging the battery in the condition of cool and ventilated. Please
make sure of pressing out the buckle between the main unit and the
pedestal, otherwise the battery charging might be failed because of the
poor contact.
8.7 Do not disassemble the Lithium battery, it will lead to the circuit
short or the electrolyte leakage.
8.8 Do not squeeze, shake and short the battery, do not store the battery
with metal material.
①WARNING: If the curing light works for 40s continously, the
temperature of the top of optical ber may reach 56℃.”
②WARNING: Do not modify this equipment without authorization
of the manufacturer.
9. Contraindication
The heart disease patients, pregnant women and children should be
cautious to use this equipment.
10. Maintenance
10.1 The optical ber should be sterilized for 4 minutes with 134℃and

7
2.0bar~2.3bar (0.20MPa~0.23MPa) before each use, other parts should
be cleaned by clean water or neutral sterilized liquid, but do not soak
the equipment in the water. Do not clean by volatile or soluble liquid,
otherwise the marks of the control panel will fade.
10.2 Please clean the optical fiber to avoid the remaining resin on the
surface and infect the life-span and the eectiveness of solidication.
11. Troubleshooting
Faults Possible causes Solutions
No indication
No response.
1. Battery is out of
power.
2. Faulty of battery.
3. Battery is protected.
1.Change a new
battery.
2. Change a new
battery.
3. Charge.
The screen shows
“Er”. Faulty of main unit. Repair.
The screen shows
“E1”. Low battery. Charge.
Light intensity is
weak.
1. The optical fiber is
not installed correctly.
2. There is crevice on
the optical ber.
3. There is resin on
the top of the optical
ber.
1. Reinstall the
optical ber.
2. Change a new
optical ber.
3. Clean the resin.
The equipment
doesn’t charge
when the adapter
is connected.
1. The adapter is not
connected well
2. Faulty of adapter or
incompatible.
3. The charging point
is impurity.
1. Reconnect.
2. Change the adapter.
3. Clean by the
alcohol.

8
Faults Possible causes Solutions
Eective duration
of the battery
become short.
The capacity of the
battery decreased.
Change a new battery.
The mode
indicator twinkles
when charging.
1. low voltage.
2. Short-circuit of the
battery.
1. Back to normal
after 15 minuets
charging.
2. Change a new
battery.
If all the above solutions have been completed, the machine still can not
work normally. Please contact our special repair shop or us.
12. Packing list
The components of the machine are listed in the packing list.
13. Transportation
13.1 Excessive impact and shake should be prevented in transportation.
Lay it carefully and lightly and don't invert it.
13.2 Don't put it together with dangerous goods during transportation.
13.3 Avoid solarization and getting wet in rain or snow during
transportation.
14. Storage and transportation
14.1 The equipment should be handled carefully and lightly, kept away
from the shaking source, installed or stored at shadowy, dry, cool and
ventilated places.
14.2 Don't store the equipment together with articles that are combustible,
poisonous, caustic, and explosive.
Table des matières
Manuels Équipement dentaire populaires d'autres marques

Vatech
Vatech EzRay Air VEX-P300 Manuel utilisateur

KaVo
KaVo GENTLEpower LUX Contra-angle 25 LP Manuel utilisateur

DENTSPLY
DENTSPLY SmartLite Focus Manuel utilisateur

LM
LM ProPower CombiLED Manuel utilisateur

Owandy Radiology
Owandy Radiology RX-AC Manuel utilisateur

mectron
mectron Piezosurgery Manuel utilisateur