DORC 1100 Manuel utilisateur

MICRO
DIATHERMY
SYSTEM
Type 1100
SERVICE
MANUAL
D.O.R.C. International b.v.
Scheijdelveweg 2
3214 VN Zuidland
The Netherlands
Phone : (+31) (0) 181 45 80 80
Fax : (+31) (0) 181 45 80 90
E-mail : [email protected]
30200300E

D.O.R.C. MICRO DIATHERMY SYSTEM - Service Manual
30200300E
Contents Page
Section 1 General ..............................................................1
Section 2 Specifications .........................................................2
Section 3 Warnings ...........................................................4
Section 4 Cautions .............................................................5
Section 5 Short instructions .....................................................6
Section 6 Block diagram ........................................................7
Section 7 Schematic diagrams ...................................................8
Section 8 Adjustment manual ..................................................12
Section 9 Servicing and adjustment .............................................14
Section 10 Spare parts manual ..................................................16
Appendix A ...................................................................17
Figure List
Fig.1 Diathermy output ...............................................2
Fig.2 Diathermy unit - Front panel ......................................6
Fig.3 Diathermy unit - Rear panel .......................................6
Fig.4 Diathermy Block diagram .........................................7
Fig.5 Logic PCB schematic ............................................8
Fig.6 Logic PCB Layout ..............................................9
Fig.7 RF-Power PCB schematic .......................................10
Fig.8 RF-Power PCB Layout ..........................................11
Fig.9 RF output at maximum preset setting ...............................15
Fig.10 Diathermy unit - Inside ..........................................16
The contents of this document are confidential and the sole property of D.O.R.C. International b.v.
This document may not be reproduced in whole or in part, by photography or print or any other
means, without written permission from D.O.R.C. International b.v.

D.O.R.C. MICRO DIATHERMY SYSTEM - Service Manual
1/17
30200300E
Section 1
GENERAL
The Dutch Ophthalmic Research Center (D.O.R.C.) is an innovative company that is involved in
research, development and manufacturing of high quality instruments and equipment for ophthalmic
surgery.
All D.O.R.C. products have been clinically tested before they were released to the ophthalmic
surgeon.
D.O.R.C. plans to be able to be of service to the ophthalmic profession with even more innovative
helpful products, for many years to come.
This Service manual should be helpful to you, while working with your D.O.R.C. MICRO
DIATHERMY SYSTEM. Of course we are willing to listen to any possible suggestions that make
the Service Manual even more useful.

D.O.R.C. MICRO DIATHERMY SYSTEM - Service Manual
2/17
30200300E
0
1
2
3
4
5
6
7
8
9
Power P (W)
0 20 40 60 80 100
Preset (%)
300Ω
600Ω
1200Ω
Diathermy Output
Load
Figure 1 - Diathermy output
0
2
4
6
8
10
12
Power P (W)
200 400 600 800 1000 1200
Load R
100
50
20
Preset
(%)
(Ω)
Section 2
SPECIFICATIONS
GENERAL:
Apparatus :MICRO DIATHERMY SYSTEM
Type : Type 1100
Weight :5,5 Kg
Dimensions :13 x 22,5 x 29 cm.
MAINS INPUT:
Mains supply :100-110 VAC (50/60 Hz)
230-240 VAC (50/60 Hz)
Fuses :2x at 100/110VAC 1 AT
2x at 230/240VAC 500 mAT
Power Consumption :45 VA
OUTPUT (values at room temperature)
RF output power :12 Watt (MAX.), see fig.1
RF Frequency :13.56 Mhz, Cristal controlled.

D.O.R.C. MICRO DIATHERMY SYSTEM - Service Manual
3/17
30200300E
AMBIENT CONDITIONS:
Operating temperature :15-40EC
Humidity :85% Max.
The D.O.R.C. Diathermy system complies with the safety standards as described in the international
norm IEC601-1, Type BF, Class 1

D.O.R.C. MICRO DIATHERMY SYSTEM - Service Manual
4/17
30200300E
Section 3
WARNINGS
A warning:
Indicates a potentially harmful situation to yourself or others.
Electric shock hazard
The Diathermy system contains high voltage circuits.
After performing any repair, calibration procedure performs a final electrical safety check
and leakage current test.
Unplug the power cord before cleaning or servicing the unit. Should the power cord or plug
become cracked, frayed broken or otherwise damaged, it should be replaced immediately.
Do not touch any exposed wiring or conductive surface, whilst the cover is off and the unit
energized. The voltages present, when electric power is connected to the Diathermy system,
can cause injury or death.
Never wear a grounding wrist strap when working on a energized unit.
The operator should now perform any servicing except as specifically stated in the
Instruction Manual.
Do not, under any circumstances, perform any testing or maintenance on medical
instruments, while they are being used to monitor a patient. Always turn the Diathermy
system off before cleaning.
Explosion hazard
Never use this Diathermy system in the presence of flammable anaesthetics.

D.O.R.C. MICRO DIATHERMY SYSTEM - Service Manual
5/17
30200300E
Section 4
CAUTIONS
A caution:
Indicates a condition that may result in equipment damage or malfunction.
Electrostatic discharge through the printed circuit boards will damage the components of the
Diathermy system. Handle all circuit boards (replacements and defective) by their non-
conductive edges and use anti-static containers when transporting them.
Before servicing the equipment, ground yourself and the tool to discharge any accumulated static
charge by wearing a static control wrist strap.Use hospital-grade grounded receptacle only.
Servicing of this product, in accordance with the Service Manual, should never be undertaken in
the absence of proper tools, test equipment and the most recent revision of the Service Manual,
which must be clearly and thoroughly understood.
Do NOT apply tension on the line cord.
Check rear panel voltage setting before connecting the Diathermy system to AC main power.
NEVER IMMERSE THIS UNIT IN LIQUID !!

D.O.R.C. MICRO DIATHERMY SYSTEM - Service Manual
6/17
30200300E
Section 5
SHORT INSTRUCTIONS
NOMENCLATURE OF FRONT PANEL.
1. Power button.
2. Preset regulator.
3. RF Output connector.
4. RF Output indicator
Fig. 2 Diathermy unit - Front Panel
NOMENCLATURE OF REAR PANEL.
1. A/C Outlet with fuse
holder.
2. Footswitch
connector.
3. Earth Potential
Equalization.
Fig. 3 Diathermy unit - Rear Panel

D.O.R.C. MICRO DIATHERMY SYSTEM - Service Manual
7/17
30200300E
6. BLOCKDIAGRAMS
Fig. 4 Diathermy unit - Blockdiagram

D.O.R.C. MICRO DIATHERMY SYSTEM - Service Manual
8/17
30200300E
Section 7
SCHEMATIC DIAGRAMS
Fig.5 LOGIC PCB SCHEMATIC.