HC-3
1-1
SECTION 1: TECHNICAL SPECIFICATIONS
INPUT: 220-240 Volts, 50 Amps –Delay Circuit Breaker
50 Hz or 60 Hz (must be preset)
OUTPUT: 0-15 Volts, 0-333 Amps
0-10 Volts, 0-500 Amps
0-5 Volts, 0-1000 Amps
OVERLOAD: Short duration overloads are possible on each tap. The test set is capable of up to
5000+ amps, depending on the impedance of the test circuit.
DUTY CYCLE: Continuous at 100%
5 minutes ON/15 minutes OFF at 200%
1 minute ON/10 minutes OFF at 300%
10 seconds ON/5 minutes OFF at 400%
3 seconds ON/5 minutes OFF at 500%
INSTRUMENTATION: Currentmeter: 4 1/2 digit LCD
Ranges: 0-19.999/199.99/1999.9/5000+ amperes
Accuracy: +/- .5% F.S. to 2000 Amps, 1% F.S. 2000-5000 Amps
Timer: 6 digit LCD, in cycles or seconds
Ranges: 0-999999 cycles or 0-9999.99 seconds
Accuracy: +/- 0.1% reading
DIMENSIONS: 22 1/4" W x 22 1/4" D x 46" H; (266 lbs. with cables)
(565 mm W x 565 mm D x 1168 mm H; 121 kg)
OUTPUT LEADS: 8 ft. sense leads
2 ft. high current leads (Parallel 4/0)
6 ft. low current leads (10 Ga)
4 ft medium current leads (4 Ga)
ENVIRONMENTAL
AND ENCLOSURE
RATINGS: Not intended for wet locations
Indoor use only
IP Rating: IPX0
Temperature Rating: 10 –40 degrees Celsius
Humidity Rating: < 95%
Altitude Rating: <3300 ft. (1000m)
Pollution Degree: 2
Overvoltage Category: II