6. Configuration Parameters Setup
1. When the scale is off, press and hold ZERO/ON/OFF and UNIT/HOLD keys until ‘SEtUP’ is shown, which
indicates that the scale is in Configuration parameter setup mode.
2. During setup mode, press the UNIT/HOLD key to change the flashed digits. Press the TARE key to confirm the
flashed digits. Press the ZERO/ON/OFF key to exit this mode.
3. Summary of Configuration Parameters Setting:
P1.xy 5 Auto-off time: no auto-off;
01-15 minutes auto-off time 5
P2.xy 02
0=no hold function; 1=hold lager weight reading; 2-
when the variety is within +
2~50d ,auto release hold function when weight is
below 10d and auto-hold new stable weight (more than 10d);
1
P3.x 3
0=no RS232 function;
1=continuously output display data;
2= continuously output gross, TARE and net weight;
3=output display data one time when scale is stable;
4=output gross, TARE and net weight one time when scale become stable;
5=Bio-RS232;
6= continuously output display data use five numbers;
7= output display data use five numbers when scale is stable;
1
P4.x 3
Baud rate for RS232: 0=1200bps, 1=2400bps, 2=4800bps, 3=9600bps,
4=19200bps 3
P5.x 0 RS232 format: 0=8N1, 1=7O1, 2=7E1 0
P6.xy 5
Resolution select:
00=500, 08=2400, 16=7500, 24=35000,
01=600, 09=2500, 17=8000, 25=40000,
02=750, 10=3000, 18=10000, 26=50000,
03=800, 11=3500, 19=12000, 27=60000,
04=1000, 12=4000, 20=15000, 28=70000,
05=1200, 13=5000, 21=20000, 29=75000,
06=1500, 14=6000, 22=25000, 30=80000,
07=2000, 15=7000, 23=30000, 31=100000
14
P7.x 0 Division select: 0=1, 1=2, 2=5 1