DEWESOFT DS-CAL Manuel utilisateur

DS-CAL
SOFTWARE USER MANUAL
DS-CAL V20-1
1

DS-CAL
SOFTWARE USER MANUAL
1. Table of contents
1. Table of contents 2
2. About this document 5
2.1. Legend 5
3. Introduction 6
3.1. Calibration system equipment support 6
3.1.1. Multifunction calibrator support 6
3.1.2.Digital multimeter support 7
3.2. Basic principle 7
3.3. Scope of supply 8
3.3.1. DS-CAL-BOX 8
3.4. Optional accessory packages 9
3.4.1. CHG-CAL-ADAPTER 9
3.4.2. KRYPTON-TC-CAL-SET 9
3.4.3. DSI-CAL-SET 10
3.4.4. DSI-5A-CAL-CABLE 10
4. Calibration system equipment configuration 11
4.1. Digital multimeter setup 11
4.2. Agilent 34401A 11
4.3. Multifunction calibrator setup 12
4.3.1. FLUKE 57XX setup 12
4.3.2. FLUKE 55XX setup 13
5. Connecting the calibration system together 14
5.1. Digital multimeter 14
5.2. Multifunction calibrator 15
5.3. DS-CAL test board 16
5.3.1. DS-CAL1 v1 test board 17
5.3.2. DS-CAL1 v3 test board 17
5.3.3. DS-CAL2 test board 18
6. Software and driver installation 19
6.1. Installation of the USB-to-serial adapter driver 19
6.2. Installation of DSCalibrator software 19
7. Configuring the DSCalibrator software 21
7.1. General settings 21
7.2. Users 21
7.3. External devices 21
7.4.Database 23
7.5. Tolerance 24
7.6. Custom fields 25
8. Using the DSCalibrator software 27
DS-CAL V20-1 2/67

DS-CAL
SOFTWARE USER MANUAL
8.1. Adjustment 27
8.2. Calibration check 28
8.3. Adjust and Check 29
8.4. Calibration procedure 29
8.5. Special considerations 29
8.5.1. Omission of tests 30
8.5.2. Different module grounding setting 30
8.5.3. Dewesoft UUT without CAN port 31
9. Connecting the UUT 32
9.1. Sirius USB 32
9.2. Sirius EtherCAT 33
9.3. Sirius Mini 35
9.4. Sirius modules 35
9.4.1. Sirius ACC and HS-ACC 36
9.4.2. Sirius CHG and HS-CHG 36
9.4.3. Sirius HV and HS-HV 37
9.4.4. Sirius LV and HS-LV 38
9.4.5. Sirius MULTI 39
9.4.6. Sirius STG and STGM 39
9.4. 7. Sirius HD-ACC 41
9.4.8. Sirius HD-LV 41
9.4.9. Sirius HD-STGS 42
9.4.10. Sirius AO 43
9.5. DEWE43 44
Table 17: DEWE43 UUT connection description 45
9.6. Krypton 45
9.6.1. Krypton 8xTH, 16xTH 46
9.6.2. Krypton 4xLV, 8xLV 48
9.6.3. Krypton STG 48
9.6.4. Krypton RTD 48
9.7. DSI adapters 49
9.7.1. DSI-ACC, DSI-ACC-0.16Hz, DSI-ACC-20mA 51
9.7.2. DSI-V-200 51
9.7.3. DSI-CHG-50 51
9.7.4. DSI-TH-K, DSI-TH-T, DSI-TH-J, DSI-TH-C 52
9.7.5. DSI-RTD 53
9.7.6. DSI-5A 54
10. Appendix A: Connectors of the DS-CAL test board 56
10.1. CAN connector 56
10.2. Sync connector 57
10.3. PPS connector 57
10.4. Power supply connector 58
10.5. GND connector 58
DS-CAL V20-1 3/67

DS-CAL
SOFTWARE USER MANUAL
11. Appendix B: Glossary and abbreviations 59
12. Warranty information 61
12.1. Calibration 61
12.2. Support 61
12.3. Service/repair 61
12.4. Restricted Rights 61
12.5. Printing History 62
12.6. Copyright 62
12.7. Trademarks 62
13. Safety instructions 63
13.1. Safety symbols in the manual 63
13.2. General Safety Instructions 63
13.2.1. Environmental Considerations 63
13.2.2. Product End-of-Life Handling 63
13.2.3. System and Components Recycling 63
13.2.4. General safety and hazard warnings for all Dewesoft systems 64
14. Documentation version history 67
DS-CAL V20-1 4/67

DS-CAL
SOFTWARE USER MANUAL
2. About this document
2.1. Legend
The following symbols and formats will be used throughout the document.
Important
It gives you important information about the subject.
Please read carefully!
Hint
It gives you a hint or provides additional information about a subject.
Example
Gives you an example of a specific subject.
DS-CAL V20-1 5/67

DS-CAL
SOFTWARE USER MANUAL
3. Introduction
The DS-CAL test board is a calibration device and DSCalibrator is the software used for calibration and
testing of DEWESoft UUT. The calibration system consists of:
● A computer or PC (running DSCalibrator software)
●Multifunction calibrator
●Digital multimeter
●DS-CAL test board
●UUT
●Proper set of cables according to the UUT
In order for a correct operation of your calibration system the procedures described in this manual must
be followed. If you should encounter any problems during the installation and use please feel free to
contact our support team.
3.1. Calibration system equipment support
Please read this chapter carefully! It contains important information about calibration system
equipment support and usage.
3.1.1. Multifunction calibrator support
The following multifunction calibrator models are supported:
●Fluke 5500A
●Fluke 5502E
●Fluke 5520A
●Fluke 5700A
●Fluke 5720A
●Fluke 5730A
Feature
FLUKE 55XX series
FLUKE 57XX series
DSI thermocouple calibration
(for example DSI-TH-K)
Supported
NOT POSSIBLE*
Krypton 8xTH / 16xTH
Supported
NOT POSSIBLE*
Extended filter tests on Sirius HS
modules
NOT POSSIBLE**
Supported
Current clamp calibration
Supported
NOT POSSIBLE***
Table 3: Multifunction calibrator calibration feature support matrix
DS-CAL V20-1 6/67

DS-CAL
SOFTWARE USER MANUAL
* - Thermocouple output is not available by the multifunction calibrator.
** - Extended frequency range for AC voltage is not supported by the multifunction calibrator.
*** - Supported AC / DC current output capabilities are too low.
3.1.2.Digital multimeter support
The following digital multimeter models are supported:
●Keysight 34465A
●Agilent 34410A
●Agilent 34401A
3.2. Basic principle
The multifunction calibrator is used for generating signals which are measured by the DEWESoft UUT.
The DSCalibrator software (in cooperation with DEWESoft) records these measurements and also
makes adjustments to the calibration factors stored in the UUT. The DS-CAL test board is used for
distributing the signals to and from the UUT and is controlled by the DSCalibrator software over the CAN
or USB interface. The digital multimeter is used for measuring output voltage or current (for example
the excitation voltage)
DS-CAL V20-1 7/67

DS-CAL
SOFTWARE USER MANUAL
3.3. Scope of supply
3.3.1. DS-CAL-BOX
Qty
Type
Usage
1x
DS-CAL-BOX
DS-CAL test board.
1x
DC Power supply 60W 12V including power
cord
Power supply for DS-CAL test board.
1x
USB‑A to USB‑B‑mini cable, 1.8m
Used for connecting DS-CAL test board to
the PC.
1x
Installation USB stick
Contains the software, drivers and the
manuals.
2x
USB‑to‑serial converter including RS232
cable, 1m (D9f-D9f‑1M‑RS232)
Used for connecting the PC, multifunction
calibrator and the digital multimeter
together.
1x
CAN Y‑cable, 1.2m (D9F‑2D9F‑1.2m)
Used for connecting the CAN port(s)
between UUT and the DS-CAL test board.
1x
PPS cable, 2m (L1B4M‑BNC‑2m)
Used for a precise PPS time source signal.
Connects to the DS-CAL test board.
2x
Sync cable, 3m (L00B4m‑L00B4m‑3m)
Used for connecting the Sync ports between
the UUT and the DS-CAL test board.
1x
Banana cable, 2m (CABLE‑GND) with
crocodile clip
Used for connecting GND between the UUT
and the DS-CAL test board.
12x
Safety test lead 4mm, red with gold‑plated
contacts, 1m
4x used for multifunction calibrator and
digital multimeter connection with the
DS-CAL test board.
11x
Safety test lead 4mm, black with gold‑plated
contacts, 1m
3x used for multifunction calibrator and
digital multimeter connection with the
DS-CAL test board.
8x
BNC‑to‑BNC cable, 0.5m
Used for connecting BNC UUT inputs and
DS-CAL test board.
8x
BNC‑to‑BNC cable, 1.2m
Used for connecting BNC UUT analog
DS-CAL V20-1 8/67

DS-CAL
SOFTWARE USER MANUAL
outputs and DS-CAL test board.
8x
MULTI cable, 0.4m (D15m‑D15m‑0.4m)
Used for connecting MULTI UUT inputs and
DS-CAL test board.
8x
DSUB9 cable, 0.4m (D9m‑D9m‑0.4m)
Used for connecting all DSUB9 UUT inputs
and DS-CAL test board.
8x
Counter cable, 0.4m (L1B7m‑L1B7m‑0.4m)
Used for connecting UUT counter inputs and
DS-CAL test board.
Table 4: Scope of supply for DS-CAL-BOX package
Illustration 2: CHG-CAL adapter accessory package
3.4. Optional accessory packages
3.4.1. CHG-CAL-ADAPTER
Accessory package CHG-CAL-ADAPTER is required for calibration of SIRIUS-CHG and SIRIUS-HS-CHG
modules and for DSI-CHG-50 adapter. CHG-CAL adapter is shown on the Illustration 2.
3.4.2. KRYPTON-TC-CAL-SET
Accessory package KRYPTON-TC-CAL-SET is required for calibration of Krypton 8xTH / 16xTH
instruments.
DS-CAL V20-1 9/67

DS-CAL
SOFTWARE USER MANUAL
Illustration 4: DSI-5A-CAL-CABLE accessory package
3.4.3. DSI-CAL-SET
Accessory package DSI-CAL-SET is required for calibration of DSI adapters. Contents of the package are
shown on the Illustration 3.
3.4.4. DSI-5A-CAL-CABLE
Accessory package DSI-5A-CAL-CABLE is required for calibration of DSI-5A adapters. Contents of the
package are shown on the Illustration 4.
DS-CAL V20-1 10/67
Table des matières
Manuels Équipement de test populaires d'autres marques

SMART
SMART KANAAD SBT XTREME 3G Series Manuel utilisateur

Agilent Technologies
Agilent Technologies BERT Serial Manuel utilisateur

Agilent Technologies
Agilent Technologies N3280A Manuel utilisateur

Vernier
Vernier Go Direct Voltage Manuel utilisateur

Lifeloc
Lifeloc R.A.D.A.R. Manuel utilisateur

Fluke
Fluke T5-600 Instructions d'utilisation et d'installation