deconta aircontrol S Series Manuel utilisateur

Instruction manual
Negative Pressure Monitor
aircontrol S 110V
Manufacturer: deconta GmbH
Im Geer 20, D - 46419 Isselburg
Description / Type No.: aircontrol S 1 110V Type 608
aircontrol S 2 110V Type 603
aircontrol S 3 110V Type 609
aircontrol S 1
aircontrol S 2
aircontrol S 3

Instruction manual
aircontrol S 110V
page 2
Table of contents
on page
1 Introduction 3
2 Basic safety advice 4
2.1 Intended use 4
2.2 Operation 4
3 Transport 5
3.1 Delivery 5
3.2 Transport 5
4 Scope of delivery 5
5 Technical description 6
6 Technical data 7
6.1 Technical data 7
6.2 Basic equipment and possible upgrades 7
6.3 Terminal assignment potential-free alarm contact 8
7 Initial operation 8
7.1 Telealarm GSM, inserting a Sim-Card 9
7.2 Standby function 10
8 Display-indication 11
8.1 aircontrol 11
8.2 Telealarm 12
9 Settings / operation 13
9.1 Possible settings / Code chart aircontrol 14
9.2 Possible settings / Code chart Telealarm 15
10 Extension modules 16
10.1 Additional measuring channels 16
10.2 Printer module 16
10.3 Emergency battery 17
10.4 Measurement data memory module 18
11 Declaration of conformity 20

Instruction manual
aircontrol S 110V
page 3
1 Introduction
Thank you for selecting a deconta product.
With this device you obtain a practical solution with simple operation, which was
completed in a compact and functional way.
The deconta products guarantee:
Stability, long life and serviceability on site
Mechanics with „kick“
Pleasing design
The copyright of this instruction manual remains with deconta. This manual is intended
for assembly, operation and maintenance personnel. It contains instructions and drafts of
technical nature which may neither be distributed nor used in any unauthorised way for
competitive purposes or passed on to others.
For more information, please visit our website www.deconta.com

Instruction manual
aircontrol S 110V
page 4
2 Basic safety advice
The handling of the appliance technology is only allowed for instructed staff. The exact
knowledge of the instruction manual is an important condition for your staff in regard to
the handling of the machine.
2.1 Intended use
As an operator, you are obliged by deconta to follow the instruction manual and to use
this engineered technology equipment only in accordance with the regulations and its
suitability! In the event of non-observance, deconta assumes no liability.
2.2 Operation
In order to ensure the safety during the operation of the device, the following must be
observed:
Do not place in an explosive area
Necessary repairs, maintenance and cleaning, in particular in the field of electrical
equipment has to be realized only by qualified staff
The safety and protective equipment has to be kept in perfect functioning.
Attached safety instructions have to be kept in a readable state and have to be
followed.
In order to ensure safety, any changes to the machine are prohibited.
ATTENTION!
The device is not suitable for the use in a condensed, corrosive, flammable and
explosive compartment air. The ambient and medium air temperature must be
situated in a sector from -10 up to+50°C
Protect against moisture!

Instruction manual
aircontrol S 110V
page 5
3 Transport
3.1. Delivery
Unless agreed otherwise, the lock system is packed completely and securely for
transport by deconta. Transport damages have to be documented at once during the
handing over of the carrier or another supplier. Please note the possible damages
additionally on the way bill. To avoid damages caused by improper handling or
carelessness, it is common practice to handle the Transport with care.
3.2 Transport
In order to protect the device during the transport, it is located in a transport box.
Care must be taken to ensure the device is not exposed to impacts or blows, because
otherwise the function and safety of the device cannot be guaranteed.
4 Scope of delivery
If no other agreements have been made, the scope of delivery of a negative pressure
control device consists of:
Portable set
Power cable (only aircontrol S 1)
Short description
Instruction manual

Instruction manual
aircontrol S 110V
page 6
5 Technical description
The aircontrol S is designed to measure, document and react to alarm conditions in
regard to the Negative Pressure Monitoring.
The devices can be equipped with individual extension modules according to customer
requirements.
There are 3 different versions available. aircontrol S 1 offers space for 4 extension
modules, aircontrol S 2 offers space for 10 extension modules, aircontrol S 3 offers
space for 5 extension modules and is provided with an additional automatic telephone
device (Telealarm GSM), to initiate an alarm via the mobile phone network, as well as
with integrated connectible sockets for Standby-units.
For the calibration or service it is sufficient to change the concerned modules.
Thanks to the combination of an easy operation and high flexibility the aircontrol S
represents a powerful and future-proof measuring device for your site.

Instruction manual
aircontrol S 110V
page 7
6 Technical data
6.1 Technical data
S 1 S 2 S 3
Measurements in mm (L x W x H) 305 x 350 x 177 475 x 350 x 177 495 x 350 x 187
Weight 5,5 kg 7,9 kg 8,4 kg
Power connection 100-120 V / 16 A 100-120 V / 16 A 100-120 V / 16 A
Range of measurement 0 - 100 Pa 0 - 100 Pa 0 - 100 Pa
Pressure resistance Load cell max. 0,25 bar max. 0,25 bar max. 0,25 bar
Occupied extension slots 1 1 1
Free extension slots 4 10 5
6.2 Basic equipment and upgrades
S 1 S 2 S 3
Powder coated lockable housing
Large display with light
Keyboard vandal-proof
Dry alarm contact (e.g. for external Telealarm
GSM)
Alarm plug 100-120 V / 16 A
Connectible socket 100-120 V / 16 A for
Standby-units
Measuring connec tion for tube 8 x 1 mm
1 piece measuring channel
Telealarm GSM integrated Art.-No.
Additional measuring channels O
(max. 4) O
(max. 7) O
(max. 5) BE2324
Printer module (occupies 3 extension slots) O O O BE2326
Emergency-battery (occupies no extension
slot) O BE2327
Measuring data memory module O O O BE2328
= standard O = optional = not possible

Instruction manual
aircontrol S 110V
page 8
6.3 Terminal assignment dry alarm contact (only aircontrol S 1 und S 2)
PE
1
3
2
In normal condition: contact 1 and 2 open, contact 1 and 3 closed
In alarm condition: contact 1 und 2 closed, contact 1 and 3 open
Without main voltage: contact 1 and 2 closed, contact 1 and 3 open
7 Initial operation
Important instructions:
Before every use, please check the device, cable and plug for damages.
In the event of malfunction, you have to stop the device at once and secure it.
Have the malfunctions repaired immediately.
The device and their electrical connections may not get wet or be operated in a
humid environment
The screw caps of the measuring and reference connections have to be screwed
down tightly.
The measuring tubes may not be flexed or damaged.
Do not place the measuring tube in an area where they can be pressed together
Do not expose the reference connection to dynamic pressure ratios (e.g.: wind)
The device has to stand horizontally on an even surface.
Do not expose the device in operational state to any vibrations or shocks.
Short tube lengths shorten the response time of the measuring device.
By measuring tube lengths over 150 m, tubes with bigger diameter should be
used.
The pressure on the measuring connections must not exceed 0,1 bar (10.000 Pa),
because otherwise the device can be damaged. Never blow inside of it, suck or
impinge it with other pressures!
Starting position:
Connect negative pressure areas with measuring tubes to the device
„connection -“
Choose the reference measuring point and connect through measuring tube to the
device „connection +“.
Turn on the instrument

Instruction manual
aircontrol S 110V
page 9
Important: Rooms adjoining to the area of operations have to be chosen as reference
measuring points. A configuration of the reference measuring point can be cancelled, if
the Negative Pressure Monitor is placed at the location of the reference measuring point.
Attention: Never use the Negative Pressure Monitor in the
contaminated work area!
7.1 Telealarm GSM, inserting of Sim-Card (only aircontrol S 3)
If the Sim-Card is protected by a PIN-Number, this Number must be saved in the
Telealarm before the Sim-Card is inserted. (Enter #20, see Code-chart).
Switch-off unit and pull power plug.
Loosen screws of protective cover and take off cover.
Press on the yellow button with a sharp-pointed object, e.g.
ballpoint pen. Push out the Sim-Card-Holder and remove from
slot.
Insert Sim-Card into Holder. Ensure that the chamfered
corner is in the correct position.
Reinsert Sim-Card-Holder. Ensure the correct installation
position of the Sim-Card-Holder.
Replace protective cover.

Instruction manual
aircontrol S 110V
page 10
7.2 Standby Function (only aircontrol S 3)
The Standby Function is activated via a Code-Entry (Code-chart aircontrol, Code 60).
As soon as the Negative Pressure Monitor aircontrol notices a negative pressure in the
removal area that is too low (constant alarm via X seconds, adjustable at aircontrol), an
integrated Standby-mode is activated.
The Standby-Negative pressure unit is supplied with power by its connection to one of
the blue sockets.
After the completion of the alarm the Standby-Negative pressure unit keeps on running
and is switched off via a code-entry (Code-chart, aircontrol, Code 61) only after the cause
of alarm is clarified and resolved.