The PeriWave Pump Product Line:
PeriMicro, PeriMilli, and PeriNano OPERATOR’S MANUAL
Congratulations on your purchase from CorSolutions! Please read this Operator’s Manual which explains
proper operation of the fluid delivery pump.
Safety Warning: - Not approved or intended for human use.
- Always wear protective eyewear when pump is in use.
Liquid Outlet
Liquid Inlet
Pump
Rollers
Flow
Sensor

2
CorSolutions Version V2.003
SETTING UP THE PUMP
Hardware and Cable Connections. The pump has been shipped with a Power Supply and appropriate
interface cables. The power supply and USB cable connect to the receptacles located on the rear panel of the
pump. The power supply connects to a 110-220V AC power source. The USB cable should be connected into
the host computer USB port. The hardware and cable connection set-up is now complete. The power switch
is also located on the rear panel of the pump.
If you are using two pumps with one computer, repeat the process.
IMPORTANT POINTS:
1) Pump is designed and calibrated for aqueous-based solutions. If other solution types are used,
the flow rates will not be accurate.
2) For proper operation, 1/32-inch tubing must be attached to the flow-sensor outlet. This tubing
is part of the dampening system and is needed for proper closed loop feedback for the flow
control. If a different outlet tubing type is desired, please employ a union at the end of the
1/32-inch outlet tube to connect to the desired tubing type. CorSolutions has adapters
available to convert from 1/32-inch to 1/16-inch (part number: Union-188-P)
3) The lines should be rinsed with 5% bleach before turning off the pump or after use with salty
or viscous solutions. This will ensure the flow sensor is maintained and will prevent it from
clogging and avoid the growth of bacteria.
4) Only use water-based solutions or 20/80 ration of Isopropynol/Water to rinse the lines. Use of
solvents such as acetone and Ethanol will cause damage to the pump tubing.

3
CorSolutions Version V2.003
Software. Two software versions have been shipped with your pump. One version is for computers that do
not have LabView installed and the other version is for computers with LabView installed. Please select the
appropriate version for your computer.
Non-LabVIEW Computers:
1. Install the software version for computers without LabVIEW, click on the Setup icon and then
installation will be completed automatically.
or
LAbVIEW Software Computers:
1B. To install the software version for computers currently running the program LabVIEW use the
“CorSolutions Pump App Build” . Copy the entire directory “CorSolutions Pump App Build” with all
its contents, to the computer hard drive. Then click on the application CorSolutionsPump.EXE, and
installation will begin.
Installation of the USB to Serial Driver: For both LabVIEW and non-LabVIEW computers. Please note that the
application requires Administrative Rights so that the communication port (COM) may be configured
correctly for the pump.
Installing the USB to Serial Driver (PC Requires Administrative Rights):
2. Be sure to leave the USB to Serial adapter cable plugged into the USB port with device powered on
while making this change.
3. Run the program “CDM20824_Setup.exe” (You will need administrative rights)
4. Open the PC’s “Device Manager”
5. Expand the "Ports (COM & LPT)" entry (click on the + in the box).
IMPORTANT: If the version for computers not having LabView is accidently installed on a
computer with LabView, corruption of the LabView program is possible.

4
CorSolutions Version V2.003
6. Right click on the "USB Serial Port (COMx)" entry, a drop down will appear:
7. Click on the "Port Settings" tab at the top, then click the "Advanced" button.

5
CorSolutions Version V2.003
8. The final dialog will appear:
9. The "Latency Timer (msec)" setting is to be changed. Click on it then select and change the value to 2.
10. Click on the OK button, and follow back through the dialogs clicking OK in turn.
2

6
CorSolutions Version V2.003
USING YOUR PUMP
Starting the Software. Start the software by clicking on the “CorSolutions” icon shown below.
A box, similar to the one below will then open.
Then select “Scan All Ports”, as indicated below.
A CorSolutions’pump will be detected if present, and will be indicated in the “Device Found” field.

7
CorSolutions Version V2.003
Alternativly to the “Scan all ports” method, if the COM port number is known in advance, one may check the
“Check for Device?” box in order to avoid the scanning process.
Once a pump(s) is found, select the “Finished” button at the bottom right of the screen. (If “No Device on
port” occurs thus indicating there is no pump found, check to see the communication cable is connected and
that there is power to the pump.)
The software will then open as indicated below for a single pump (Left) or two pumps (right).

8
CorSolutions Version V2.003
Purging the Pump. To purge the pump, place the pump inlet into a suitable vessel containing the aqueous-
based solution, as shown here.
Next remove the fitting to the inlet side of the flow sensor, as shown below. This will relieve the backpressure
in the system during the purge process.
Remove Inlet Line
To Flow Sensor When Purging
At high Flow Rates

9
CorSolutions Version V2.003
Then click on the “Mode” button to put the pump in “Purge Mode”. The “Flow Control” light should not be
illuminated and the Purge Mode button will illuminate. In Purge Mode the flow sensor is disabled, and the
pump operates as a traditional peristaltic pump.
Enter a flow rate value of 500 and then click “Set”. This will have the pump run at a fixed rate of 500 µL/min.
Liquid will soon be observed exiting the Flow Sensor inlet fitting, as shown below. Allow for liquid to continue
exiting until all major air bubbles are purged.
Purge Until Liquid Exits
the tube and no air
bubbles are observed

10
CorSolutions Version V2.003
Next click the “Stop” button in the software.
Then reconnect the fitting into the flow sensor inlet, ensuring that the tubing is seated past the ferrule’s end
for the PeriNano & PeriMicro pump models. (For the PeriMilli model, insure the tube is flush with the bottom
of the flat bottom ferrule.) The image below shows how the tubing needs to protrude from the fitting end in
order to make a proper connection to the flow sensor.
Ce manuel convient aux modèles suivants
2
Table des matières
Manuels Pompe à eau populaires d'autres marques

Sykes AmeriPumps
Sykes AmeriPumps GP100M Guide de dépannage

DUROMAX
DUROMAX XP WX Series Manuel utilisateur

BRINKMANN PUMPS
BRINKMANN PUMPS SBF550 Manuel utilisateur

Franklin Electric
Franklin Electric IPS Manuel utilisateur

Xylem
Xylem e-1532 Series Manuel utilisateur

Milton Roy
Milton Roy PRIMEROYAL Manuel utilisateur