CellPath JAY-0100-00A Manuel utilisateur

Slimline Slide Drying Hotplate
User Manual
www.CellPath.com |T: +44 (0)1686 611333

2
User Manual
C
ellTec Slimline Slide Drying Hotplate
Thank you for purchasing this piece of CellTec laboratory equipment.
To get the best performance from your equipment and for your own safety
please read these instructions carefully before use.
www.CellPath.com | T: +44 (0)1686 611333
2
User Manual
Contents
3 General Notes
3 Specimen Safety
4 User Safety
4 Power Lead and Connection
to Electrical Supply
5 Specification
5 Location
6 Operating Instructions
7 Cleaning Instructions
8 Calibration and Oset Instructions
9 Setting Controller Oset Parameters
10 Routine Inspection Recommendations
10 Spare Parts
11 Wiring Diagram
12 Troubleshooting Guide
14 Warranty Terms and Conditions
15 Non Warranty Information
19 EC Declaration of Conformity

3
www.cellpath.com | T: +44 (0)1686 611333 3
Slimline Slide Drying Hotplate
Symbols
This symbol appears in documents
and on equipment to warn the user
that there are hot surfaces on the
equipment.
This symbol appears in documents
and on equipment to warn the user
that instructions must be followed
to ensure correct or safe operation.
General Notes
This product is designed for laboratory
use only. Always follow good laboratory
practice.
If this product is not used in accordance
with these instructions then basic
safety protection may be aected.
If damaged or in case of failure the
power supply unit supplied with this
product should be replaced with an
equivalent power supply unit.
Before using any cleaning or
decontamination method please
refer to the Maintenance and Cleaning
section to ensure the proposed method
will not damage the unit.
Connect only to a power supply with
the corresponding voltage to that
specified on the rating label positioned
on the rear of the unit.
Ensure that the power supply
has an earth (ground) terminal.
Specimen Safety
It is the users responsibility, to ensure that
the temperature set on the instrument,
is at a level where no damage is caused
to diagnostic specimens used with the
equipment. In the event of this instrument
malfunctioning, all specimens within the
device should be checked to ensure no harm
or damage to the specimen has been caused.
Amendments
Issue 4: May 2021
3
www.CellPath.com | T: +44 (0)1686 611333 3

4
User Manual
User Safety
The equipment you have purchased complies
with the following European Directives EMC
Directive 2004/108/CE Electromagnetic
Compatibility and Low Tension 2006/95/CE as
indicated in the EC Declaration of Conformity
included in the document. This instrument
has been designed and constructed in a
manner which minimises the risk of electrical
shock to the operator, oers maximum
protection from overheating and provides
clear and adequate labelling of instrument
controls. The instrument requires no regular
servicing, but CellPath Ltd do recommend an
annual inspection, as detailed in the manual
which will prolong the life of the instrument
to ensure continued safety.
Do not touch any electrical
contacts or open any closure
plates. RISK OF ELECTRIC SHOCK!!
DO NOT:
Allow molten wax to accumulate on the
surface of the hotplate.
Use metal instruments or scouring
agents to clean the surface of the
hotplate.
Place fluid containers on the surface
of the hotplate without an adequate
spillage tray.
Do not immerse in water.
DO:
Position the unit so it can be
disconnected from the power
supply with ease.
www.CellPath.com | T: +44 (0)1686 611333
4
User Manual
Power Lead
and Connection
to Electrical Supply
Check the electrical supply is
compatible with the rating label.
IF IN DOUBT CONSULT AND
ELECTRICIAN. THE PRODUCT
MUST BE EARTHED!
Where the mains supply or
plug connection diers refer to
local regulations or consult an
electrician.

5
www.cellpath.com | T: +44 (0)1686 611333 5
Slimline Slide Drying Hotplate
5
www.CellPath.com | T: +44 (0)1686 611333 5
Specication
The CellTec Slimline Slide Drying Hotplate
is designed to dry parain tissue sections
eiciently whilst taking up a minimum of
bench space in the laboratory. The hotplate
features a matt black heated surface which is
digitally controlled at a constant temperature
and is combined with four movable rails
which allow heated air to pass underneath
the slides to accelerate drying. The movable
rails can be removed to allow slides to
be placed directly on the heated surface
according to customer preference.
The Slimline Slide Drying Hotplate
features a digital temperature control,
over temperature cut out and miniature
circuit breakers for safety.
Dimensions: Width 183mm
Depth 450mm
Height 85.5mm
Capacity: 28 Slides
(2 rows of 14 slides)
Weight: 2.2Kg (4.4 lbs)
Display: Digital Display
with 0.5° accuracy
Power Supply: 110V/230V a.c. 50-60Hz
Safety: Class 1 cut out
Heater Power: 400 watts
Location
The product must be placed on a smooth,
level and sturdy work surface. Suitable for use
in ambient temperatures 5°C to 40°C with a
maximum humidity 80% (temperature 31°C)
decreasing to 50% (temperature 40°C).
Environment
This instrument is required to comply
with the European Union’s Waste Electrical
and Electronic Instrument (WEEE)
Directive 2002/96/EC.
It is marked with the following symbol:
This product should be returned to
CellPath Ltd for disposal and recycling.

www.cellpath.com | T: +44 (0)1686 611333
6
User Manual
Operating Instructions
Place the hotplate on a smooth and level
work surface.
Place the four microscope slide support
rods in position according to user
preference.
Connect the mains plug to the electrical
supply and switch on (Ensure the power
supply is properly earthed).
Turn on the hotplate.
Select the desired temperature.
A. Press button Pthen release it (do not
hold down button Pfor 5 seconds).
B. The display will show SP alternating
with the current set temperature.
C. To change the set temperature
press the UP key to increase the value
or DOWN to decrease it. These keys
increase or decrease the value one
digit at a time, but if the button is
pressed for more than one second
the value increases/decreases rapidly,
and aer two seconds pressed, the
speed increases even more to allow the
desired values to be reached rapidly.
D. Exiting the Set mode is achieved by
pressing the Pkey or automatically if
no key is pressed for 15 seconds.
Aer that time the display returns
to the normal function mode.
The heater indicator will illuminate to
show heater activity.
The instrument will then warm up to the
desired temperature, you will observe
the temperature rise on the display.
The instrument is designed to warm
up quickly (roughly 3 to 5 minutes).
This means that when initially warming
up, the instruments will slightly over
shoot the set target temperature
(by 2-5
°
C). The heater will then stop
and allow the instrument to cool to the
desired working temperature and then
maintain the set working temperature
at +/-1
°
C, by alternating heating and
cooling.

7
www.cellpath.com | T: +44 (0)1686 611333 7
Slimline Slide Drying Hotplate
Cleaning Instructions
The lower case work of the slimline
slide dryer hotplate, including the
control panel, may be wiped using small
quantities of mild detergent or polishes
applied with a so cloth.
The heated surface of the hotplate will
require cleaning at regular intervals,
using a minimal quantity of mild
domestic detergent applied with a so
synthetic sponge.
The microscope slide support rods can
be removed to facilitate easier cleaning.
SCOURING PADS OR DE-SCALING
AGENTS MUST NOT BE USED TO
CLEAN THIS INSTRUMENT.
Miniature Circuit
Breakers
Located at the rear of the instrument.
In the event of a fault, push back in to reset.
If fault situation continues, please contact
your Service Engineer or CellPath Ltd.
Portable Appliance
Testing
Portable appliance testing should be carried
out by a qualified person.
THIS EQUIPMENT MUST
NOT BE FLASH TESTED!

www.cellpath.com | T: +44 (0)1686 611333
8
User Manual
Calibration and
Offset Instructions
Turn on the Slide Drying Hotplate.
Set the temperature of the Slide Drying
Hotplate to the desired temperature.
Let the instrument reach working
temperature and leave for 30 minutes
before carrying out a calibration check,
this ensures that the temperature
of the hotplate has equilibrated.
Check temperature with a calibrated
digital thermometer fitted with a surface
probe. The temperature should be
measured in the centre of the hotplate
(back to front and side to side), as shown
in the image.
Compare the temperature shown on
the Slide Drying Hotplate display with
the temperature shown on your external
digital probe.
If the two numbers are in close
agreement i.e. within ± 1°C (as shown in
the image above), then the instrument
is within calibration. If the temperatures
are not in agreement i.e. a dierence
greater than ± 1°C, then the instrument
will need to be re-calibrated.

9
www.cellpath.com | T: +44 (0)1686 611333 9
Slimline Slide Drying Hotplate
Setting Controller
Offset Parameters
Press the Pbutton and keep it pressed
for 5 seconds until the display flashes
showing SP1.
Then press the down arrow,
the display will change to r.P.
Then press the Pbutton again
and the display will show 0.
Press the UP arrow until the number
shows 146 then press P.
Once the controller is unlocked
the display will show SPL.
Press the DOWN arrow 5 times
until the display shows CA.
CA = calibration and is where
adjustments can be made to the
temperature on the display so that the
temperature shown is the same as the
temperature shown on your reference
digital thermometer.
Press the Pbutton. The display will
now show a number, this number is
the current adjustment (oset) applied
to the temperature display. This was
the adjustment (oset) which was
used when the slide drying hotplate
was manufactured to calibrate the
instrument.
Use the up and down arrows to select
the correct adjustment to match the
temperature shown on your reference
thermometer. Then press the Pbutton
to set the adjustment (oset).
The screen should now show CA.
Press the Up arrow and keep it pressed
(for about 5 seconds) until the display
shows the measured temperature on
the display again.
The temperature on the display should
be in close agreement i.e. within ± 1°C
of the temperature shown on your
reference thermometer.
It is recommended to leave the
instrument heat for a further 30 minutes
before re-checking the calibration of the
instrument. To make sure you do not
need to make a further small change
to the adjustment (oset).

www.cellpath.com | T: +44 (0)1686 611333
10
User Manual
Part Number Description Quantity Required
JBA-EE0718-00A Silicone Heater 200W 2 O
JBA-ES0272-00A Mains Switch 1 O
JBA-EL0100-00A Neon Amber and Tags 1 O
JBA-EX1151-00A Power Entry Module 1 O
JBA-EX1212-00A 5A Circuit Breaker 2 O
JBA-EX1236-00A Digital Temperature Controller PID 1 O
JBA-EX1035-00A Solid State Relay 1 O
JBA-EH1243-00A Thermal Cut Out 120°C 1 O
JBA-ET0835-00A Temperature Sensor K-Type Thermocouple PTFE Insulated 1 O
Routine Inspection
Recommendations
CellPath Ltd recommend that a simple annual
inspection be made for all CellTec laboratory
equipment in order that any malfunction can
be identified and rectified as early as possible.
This is to ensure user safety and prolong
instrument life span.
Recommended checks to be made:
Condition of Power Lead.
A visual inspection to ensure the
insulation is not damaged and that
the correct fuse is fitted.
Functioning of Heater On Lamp.
Heater lamp should be on when the
instrument is warming up.
Condition of the Hotplate Surface.
Surface of the hotplate should
be free of scratches and dents.
Spare Parts List
Table des matières
Autres manuels CellPath Séchoir
Manuels Séchoir populaires d'autres marques

Indesit
Indesit IDV 75 Manuel utilisateur

Infiniton
Infiniton SD-DG85C Manuel utilisateur

BOMANN
BOMANN WT 5019 Manuel utilisateur

Alliance Laundry Systems
Alliance Laundry Systems TMB795C Manuel utilisateur

Asko
Asko T793C Manuel utilisateur

Kenmore
Kenmore 8041 - 5.8 cu. Ft. Capacity Electric Dryer Manuel utilisateur