1. Press ppropri te ISOTOPE SELECTION KEY
2. Press SETUP
3. The c libr tion v lue for selected isotope is displ yed on ETD (#1)
4. Use the ⇑⇓ keys to set new c libr tion v lue on ETD (#1)
5. Press ENTER
6. Press SETUP
++#&" .*'+'( $, *
1. Press OTHER KEY
2. Press SETUP
3. Use ⇑⇓ keys to set new c libr tion v lue on ETD (#1)
4. Press ENTER
5. Press SETUP
6. System returns to norm l oper ting mode – the new c libr tion v lue is stored
*'+'( !,$+$#)+#'&$, *
1. Press ny st nd rd ISOTOPE SELECTION KEY
2. Press SETUP
3. Current def ult v lue for selected isotope is displ yed on ETD (#1).
Comp re this v lue to “Appendix D” in Oper tion M nu l
4. Use the ⇑⇓ keys to djust isotope def ult v lue s needed on ETD (#1)
5. Press ENTER
6. Press SETUP
7. System returns to norm l oper ting mode – the new c libr tion v lues re stored
The following method c n lso be used to determine the di l setting for c libr ted st nd rd.
.+#-#+. *,) % &+
1. Press OTHER KEY
2. ETD (#1) displ ys di l setting for OTHER key
3. Use the ⇑⇓ keys to djust isotope def ult v lue s needed on ETD (#1)
4. Put source in well nd t ke re ding
5. Press nother isotope key
6. Press OTHER nd di l setting is b ck to def ult setting