I. Procedure for taking open circuit MILLIVOLT READINGS of the Thermocouple.
Disconnect Thermocouple from Baso Control and attach proper cable to
Thermocouple.
Turn selector switch to the MV range scale needed for the type of Thermocouple
being tested. Pilot Generators are tested on the 300 or 900 MV scale, as
needed. Insert plug end of cable in millivolts jack of test kit.
Millivolt readings should agree with those given in Table I, with a normal
pilot flame.
II. Procedure for taking fixed load millivolt readings of the thermocouple.
Note:
CABLES AND
THERMOCOUPLE
ADAPTERS FOR
MILLIVOLT
TESTING
Red
Red
THERMOCOUPLE LEAD
A-GREEN
Black
TRANSFORMER
Lc
PtLOT BURNER GAS LINE
Fits series 17 and 50 Thermocouples.
With Adapter 1, Fits 80 Series Thermocouples.
With Adapter 2, Fits B30 and B20 Series Thermbcouples.
With Adapter 4, Fits 97 and 107 Series Thermocouples.
Fits other makes of Thermocouples or Pilot Generators
and Basofron valves.
- Fixed Load Adapter
- Closed Circuit Adapter
2
Disconnect thermocouple from Baso control and attach fixed load adapter A
to thermocouple using proper fitting.
Attach alligator clip cable to terminals of fixed load adapter A. Bent
terminal indicates negative polarity.
Turn selector switch to MV range scale needed for the type of therrnocoupJe
being tested. Insert plug end of clip cable in millivolts jack of test kit.
Millivolt readings should agree with those given in Table I, with a normal
pilot flame.
Procedure Iand Procedure II are to be used for testing Baso Thermocouple
Leads and Procedure IV for drop-out testing of Baso Power Units.
Do not use the closed circuit adapter for testing Baso Automatic
Pilot Control Systems.
POWER SOJRCE I HE RMOS IA