Stocking the Medisafe Plus Refrigerator
The Medisafe Plus vaccine refrigerator is designed to maintain a uniform temperature inside the cabinet. To do so,
sufcient air ow is required to prevent “cold spots” from forming. It is therefore important that an appropriate
level of care is taken in stocking the refrigerator.
• Cool the vaccine refrigerator before stocking with vaccine. We recommend allowing the refrigerator to run for
at least six (6) hours prior to loading with vaccines.
*There will be a small amount of condensation produced inside the refrigerator. This will dissipate over time.
• Distribute stock evenly throughout the refrigerator.
• Ensure vaccines with the shortest expiry are stored at the front of the refrigerator.
• DO NOT place stock directly against the supply air panel at the rear of the refrigerator. This could cause
the stock to freeze.
• DO NOT place large cartons inside the refrigerator (i.e., at the bottom of the cabinet). Remove smaller boxes
from the carton and evenly place them on the shelves.
• Some vaccines are sensitive to ultraviolet (UV) light. Please check the specic vaccine storage guidelines
before removing vaccines from their boxes.
• If trays are used, ensure they are made of plastic, with holes to allow for sufcient air ow, and that they are-
spaced evenly on the shelves.
• DO NOT overstock the refrigerator.
• If very few vaccines are stocked, thermal mass (such as pre-cooled water bottles) should be added to the
refrigerator in areas of open space.
Basic Operation-First use
• As advised on previous page, the power failure alert “PF” should be acknowledged by pressing the red “SET”
button once on the Min-Max temperature display.
• It is recommended the Min/Max Thermometer is reset on rst start up.
Press and hold the red ‘SET’
button for 8 seconds to ensure
the Min-Max values have been
reset.
TheMin-MaxdisplaywillrstshowrT--NNN--rE—°C(rTisthenumberofhourssincethelastreset),
beforedisplayingtherE(reset)conrmation.
The Medisafe Plus refrigerator is pre-set for operation between +2° and +8°.
The AQ Box Cloud Monitoring device will begin to log data as soon as it is powered.