USER MANUAL MINI CENTRIFUGE
2. Product Description
1-2.Precautions for Use
1. Place the body on a flat, rigid surface.
2. Keep the body dry and clean.
3. When adding samples to the rotor, adjust the weight of the samples
and place them symmetrically.
4. Do not move the equipment while it is spinning.
5. Do not arbitrarily decelerate or stop the rotating rotor.
6. Do not spill liquid or put foreign substances into the body or power adaptor.
If liquid is spilt inside the equipment or around the power connection, disconnect the
power adaptor and clean and dry all parts completely. If it is not working even with
full cleaning, contact A/S center.
7. Stop running if the equipment makes big noise or trembling. Check the rotor coupling
and sample balance. If the problem is not eliminated, contact A/S center.
8. Do not apply a strong impact or move the equipment while it is spinning.
9. Precautions for handling power adaptors
- Do not pull or bend the power adaptor hard.
- Do not heat the cord or use it in humid places.
- Do not keep the power adaptor with power supplying status if it is not in use.
The mini6-RC table -topcentrifuge runs up to 6,500rpm with fast acceleration
and deceleration, 4 seconds and 2 seconds respectively. It startsto spin when the
lid is closedand stops when the lid is open.Simple action of “LidOpenButton ”
makeits operationmuch easier.
The magnet sensor betweenthe lid and the bodyrecognizesthe status of lid and keeps
its safeoperation.