ARA M Series Manuel utilisateur

ARA DEEP FREEZERS
ARA M series
USER MANUAL
M80
M300/MD300
M400/MD400
M500/MD500
M700/MD700
M870/MD870

RESEARCH USE ONLY
Copyrightⓒ2018 Hanil Scientific Inc. All rights reserved.
Contact Us
If you have any questions,
contact Hanil Scientic Inc. or place of purchase.
ARA DEEP FREEZERS
ARA M series

1. Operating Instructions -------- 4
1-1. Before Use ---------------- 4
1-2. Key Features -------------- 4
1-3. Delivery Package ---------- 5
2. Installation -------------------- 6
2-1. Preparing installation ----- 6
2-2. Technical Spectifications - 7
3. Operating --------------------- 8
3-1. Cautions for Use ---------- 8
3-2. Product description ----- 11
3-3. Operation ---------------- 13
4. Matintenance --------------- 19
4-1. Accessories -------------- 19
4-2. Maintenance ------------- 20
4-3. Emergency Door for Use - 20
(for Auto Door)
CONTENTS

USER MANUAL DEEP FREEZERS
1. Operating Instructions
1-1. Before Use
- This operating manual provides details of the device and safety cautions.
- Read this manual completely before using the device.
- This manual must always be easily accessible.
1-2. Key Features
This device adopts 'Simplified Hi-Efficiency Freezing System' which enables fast cooling and
reduce power consumption.
Convenience and safety in use is considered in developing and manufacturing this
product.
Features include:
1. 'Simplified Hi-Effciency Freezing System' minimizes the time required to reach the set
temperature and power consumption.
2. If one of 'Simplified Hi-Effciency Freezing System' fails in operation, the other system still
works and maintains the temperature. (Only Dual Comp. Type).
3. Ergonomic 7″Full Touch Screen TFT
4. 'Foam insulation, inner door insulation, 3layer insulator' systems prevent leaking cold air
and inflowing of warm air from outside.
5. 'Automatic Vacuum Breaker' makes the door easy to re-open.
6. Robust & Flexible design of door handle
7. Data can be check with a high-resolution display and data can be backed up or
updated via the usb port.
8. LCO
2
injection function prevents damage to the sample in case of an unexpected
machine stops.(Optional)
9. Filter-free Mechanism is no need to replace or clean the filter.
10. Multiple accessories are offered, including CO
2
/LN
2
back-up systems, remote monitoring
systems, and more.

USER MANUAL DEEP FREEZERS
ARA DEEP FREEZERS
ARA M series
USERMANUAL
User Manual, 1 ea
1 unit
Accessories are in the chamber.
1-3. Delivery Package
ARA M80
ARA M(D)300
ARA M(D)400
ARA M(D)500
ARA M(D)700
ARA M(D)870
Hand Remover of Ice, 1 ea
Door Lock KeyStainless Steel Shelves
1 ea (ARA M80)
3 ea (except for ARA M80)
for Manual model
for Auto model

USER MANUAL DEEP FREEZERS
2. Installation
1. Check packing conditions very carefully before unpacking. Contact Hanil Scientific Inc.
immediately if damages found.
2. Do not lift the device by hand. Preferred lifting for loading and unloading is by
mechanical lifting equipment.
3. Avoid dropping or subjecting the device to impacts.
4. Connect the device to voltage sources which correspond to the electrical requirements on
the label attached to the device.
5. Avoid installing the device in direct sunlight and in an area of high humidity.
6. Provide a clearance of at least 15cm (6 in) on all sides. (At least 100cm apart from the
ceiling)
7. This device is equipped with a grounding plug.
Plug the power cord into the grounded electrical outlet.
Improper grounding can increase the risk of electrical shock.
8. Grounding electrical work is required when there is no grounding outlet.
9. Remove the brackets on the top of the device before opening the door.
2-1. Preparing installation

USER MANUAL DEEP FREEZERS
2-2. Technical Spectifications
Model ARA M80 ARA M300
ARA MD300
ARA M400
ARA MD400
ARA M500
ARA MD500
ARA M700
ARA MD700
ARA M870
ARA MD870
Capacity 85 Lit 317 Lit 404 Lit 514 Lit 732 Lit 876 Lit
Compressor Single M model : Single / MD model : Dual, Independent
Temp.
Range
-86℃
~ -65℃
M series: -86℃~-65℃
MD series: -95℃ ~ -65℃
(ECO MODE -86℃ ~ -65℃)
M series:
-80℃~ -65℃
MD series:
-90℃ ~ -65℃
(ECO MODE :
-80℃ ~ -65℃)
M series:
-80℃~ -60℃
MD series:
-90℃ ~ -60℃
(ECO MODE :
-80℃ ~ -60℃)
At room temperature of 30℃ and humidity of 70%
Sensor PT 100 Sensor
Condenser Type Block Condenser, NO Pin, Filter Free Mechanism
Refrigerant CFC Free Refrigerant System
Controller & Display 7" Full Touch Screen TFT LCD (Touch Controller)
Shelves Included 3 Stainless Steel Shelves(ARA M80 - 1shelve)
Max. Q'ty of Racks
4 pcs of
12(3x4)
Placed Rack
12pcs of
15(3x5)-
Placed Rack
(RFU3040)
16pcs of
15(3x5)-
Placed Rack
(RFU3040)
16pcs of
20(4x5)-
Placed Rack
(RFU50)
24pcs of 20(4x5)-
Placed Rack
(RFU50)
24pcs of 25(5x5)-
Placed Rack
(RFU50)
Material
Internal Stainless Steel (#304)
External Powder Coated Steel Plate
Inner Door Powder Coated Stainless Steel (#304)
Others CFC-Free Refrigerants, Touch-Screen TFT LCD Controller
Built-in Standard Warning and Alarm Functions
Chamber
Dimension
(wXdXh, mm)
Internal
340 x 485 x 520 470 x 515 x 1,310 600 x 515 x 1,310 600 x 655 x 1,310 880 x 635 x 1,310 880 x 760 x 1,310
External
572 x 830 x 1,147 800 x 892 x 1,990 930 x 892 x 1,990 930 x 1,032 x 1,990 1,210 x 1,012 x 1,990 1,210 x 1,137 x 1,990
Net Weight 100 kg 190 kg/
220 kg
220 kg/
260 kg
245 kg/
300 kg
285 kg/
320 kg
335kg/
385kg
Packing Size
(wXdXh, mm)
960 x 760 x 1,350 1,100 x 1,100 x 2,220 1,160 x 1,160 x 2,220 1,160 x 1,240 x 2,220 1,400 x 1,400 x 2,210 1,400 x 1,630 x 2,210
Power Consumption 300 W 600W / 1,200W 800W/ 1,600W
Power
Requirement
Dual Mode
(MD series) 321 W/h
415 W/h 461 W/h 526 W/h 576 W/h 613 W/h
Eco Mode 364 W/h 386 W/h 469 W/h 508 W/h 547 W/h

USER MANUAL DEEP FREEZERS
3. Operating
Before Use
3-1. Cautions for Use
Carefully read this operating manual before using the device.
The device should be installed on hard and flat place. If the device is installed in an inclined
place, it causes noise and vibration. If necessary, level the device by adjusting the ‘Leveling tool’.
Be sure to operate the device only with the rated power supply, as specified on the label.
Do not connect wrong power source. It may be caused to over-heat, electric shock or fire.
If you move and reinstall the unit in a different place, let the unit stand stable for at least
3 hours.
Install the shelves before connecting the device to the power source.
After installation, wipe them with a soft cloth.
Put the samples to be stored into the chamber after reaching the set temperature.
Cautions for Safe
Use only in rooms(5℃~30℃).
Do not modify, pull, forcibly bend, or perform any other act that may damage the power
cord.
Do not let the product come into contact with water. Do not connect or disconnect the
power cord with wet hands.
Do not unplug the power cord during operation.
It may cause electric shock.
Do not hang any objects on the door. It may cause electric shock.
Do not put anything other than the sample container. When an object that is easy to
be combustible like a stone, a metal or paper goes inside, it may cause a fire, an electric
shock or a break down. Do not put any coolants in the chamber.
Do not touch stored samples, containers and inner wall with wet hands.
It may cause frostbite. Be sure to wear Cryo Gloves.

USER MANUAL DEEP FREEZERS
Do not place any object on the device.
Be sure the holes on the side or rear surface of the device are not blocked by a wall or
an object. It may cause a fire.
Do not install the device in dusty or humid places. It may cause short circuit.
Do not install the device at an altitude above 2,000 meters (6,562 feet).
Do not operate the device at outdoors. Store and operate the device in a place where
you can avoid rain or sunshine.
Do not operate the device near a heater, blowing an air-conditioner.
Do not use volatile, flammable products near the device.
Do not store volatile and/or flammable goods like alcohol, benzene, ether, propane gas
and glue. It could cause explosion or file.
Maintenance, adjustment and repair work should be carried out only by qualified
personnel who have been authorized to undertake such work by Hanil Scientific Inc.
This device must be connected to an earthed power supply.
Clean the power plug with a dry cloth to remove dust and grime.
Dust on the power plug may cause a fire.
Unplug the power plug before cleaning or left unused for a long period of time.
Before moving the device, unplug the power plug from the power outlet. At this time,
be careful not to damage the device.
When unplug and replug the power plug, leave a time interval at least 10 minutes.
If gas leakage occurs near the device, close the gas valve and ventilate it.
Always grasp the power plug when disconnecting the power cord.
If you notice abnormal noise or malfunctions during operation turn off the device and
disconnect the power cord, contact the Hanil Scientific Inc or place of purchase.
The mains voltage fluctuation must be within ± 10% of the nominal voltage.
The maximum relative humidity is 80% at 31℃ / 50% at 40°C.
If device packaging is wet, do not use the device and contact the Hanil Scientific Inc. or
place of purchase.

USER MANUAL DEEP FREEZERS
For product inquiries or repairs, contact Hanil scientific Inc.
When moving the device
If you want to move the product, please contact Hanil scientific Inc. or place of
purchase.
When moving the device, do not tilt or lay down the device.
If the device is tilted or laid down when moving, start operation after 24 hours.
When moving the device, remove all the samples from the device.