Alert J5 Manuel utilisateur

Instruction Manual


Federal Communications Commission (FCC) Disclaimer
Changes or modications not expressly approved by the party responsible for
compliance could void the user’s authority to operate the equipment.
This device complies with Part 15 of the FCC rules. Operations are subject to the
following 2 conditions:
1. This device may not cause harmful interference
2. This device must accept any interference received, including interference that may
cause undesired operations
Industry Canada
This device complies with Industry Canada licence-exempt RSS standard(s). Operation
is subject to the following two conditions:
1. This device may not cause interference, and
2. This device must accept any interference, including interference that may cause
undesired operation of the device.
Le présent appareil est conforme aux CNR d’Industrie Canada applicables aux
appareils radio exempts de licence. L’exploitation est autorisée aux deux conditions
suivantes:
1. l’appareil ne doit pas produire de brouillage, et
2. l’utilisateur de l’appareil doit accepter tout brouillage radioélectrique subi, même si
le brouillage est susceptible d’en compromettre le fonctionnement.
Alcohol Countermeasure Systems Corp
60 International Boulevard
Toronto, Ontario M9W 6J2 CANADA
ALERT, Alcohol Countermeasure Systems, and the Molly are trademarks of Alcohol
Countermeasure Systems (International) Inc. and are used under license.
© Alcohol Countermeasure Systems Corp 2011
The information disclosed in this document is the valuable property of Alcohol
Countermeasure Systems Corp and all copyright and other proprietary rights
to this document are reserved. No reproduction of this document is permitted
without the prior written consent of Alcohol Countermeasure Systems Corp.

INSTRUCTION MANUAL
TABLE OF CONTENTS
INTRODUCTION .............................................................. v
SAFETY AND PRECAUTIONS ........................................ v
WARRANTY .................................................................... vi
MAINTENANCE .............................................................. vi
CONTACT INFORMATION .............................................. vi
GETTING STARTED ........................................................ 1
Buttons and Menus ..................................................2
BAC/BrAC Format.................................................... 3
Power On/Shut Down .............................................. 3
Activating ................................................................. 4
Battery...................................................................... 4
TESTING .......................................................................... 5
Taking a Breath Test ................................................ 5
Recalling a Breath Test ............................................ 7
PRINTING ........................................................................ 8
Serial Printing........................................................... 8
Inserting/Charging the Printer Battery ................... 10
Loading the Paper Roll .......................................... 11
MAINTENANCE & TROUBLESHOOTING ..................... 12
Troubleshooting Table ............................................ 12
INDEX............................................................................. 13

p. v/viINSTRUCTION MANUAL p. v/vi
INTRODUCTION
Thank you for purchasing the ALERT™ J5 – a breath
alcohol tester (BAT) that measures the alcohol level in
human breath. Based on your jurisdiction, your ALERT J5
comes with preset levels that indicate whether the alcohol
breath test result is a fail (red LED light), warning (amber
LED light) or pass (green LED light).
SAFETY AND PRECAUTIONS
Take the following precautions to ensure smooth and
uninterrupted use of your device:
• Use the product for its intended purpose only
• Use AA Alkaline batteries only
• Do not store in extreme heat or cold and avoid
exposing the device to direct sunlight for extended
periods of time
• Do not blow smoke or any uids into the mouthpiece,
as these can damage the sensor
• Do not open the BAT case. Doing so will void the
warranty and may also damage the components within
• Keep all magnets and transmitting devices (such as
mobile phones) away from the BAT to minimize RF
interference
• If you spill any liquid on the ALERT J5, remove all
excess uid with a cloth and allow the device to dry
thoroughly before attempting to power on. Do not
power on in case of excessive water damage

ALERTTMJ5
p. vi/vi
WARRANTY
All ACS products are warranted to be free of defects in
workmanship and materials for a period of one year from
the date of shipment. ACS agrees to replace or repair any
defective device, provided the defect was not caused by
misuse or mishandling.
The ALERT J5 being returned for warranty repairs
must be properly packaged and shipped prepaid to an
authorized service provider.
MAINTENANCE
ALERT J5 requires periodic maintenance and calibration.
After testing and calibrating at the factory, ALERT J5 is
deactivated so it can be stored without losing any clock
time from the calibration period. You need to activate it to
start using it (see the “Activating” section). Once activated,
ALERT J5 automatically increments the date for the next
calibration by 365 days from the date of activation.
NOTE: ALERT J5 is calibrated for optimal performance at
the point of shipment.
CONTACT INFORMATION
acs-corp.com
Alcohol Countermeasure Systems Corp
60 International Boulevard
Toronto, Ontario M9W 6J2
Canada
T +1 416 619 3500
F +1 416 619 3501
Battery
Compartment

p. 1/13
INSTRUCTION MANUAL
GETTING STARTED
The ALERT J5 is a professional series breath alcohol
tester (BAT) that is accurate and easy to use. It
measures breath alcohol concentration (BrAC), which
is directly proportional to blood alcohol concentration
(BAC).
This section familiarizes you with the basic features, such
as the menus, buttons, and batteries of the device.
Display
Battery
Compartment
Intake Port
Serial Port
Indicator Light
Two-Button
Operation
Mouthpiece

INSTRUCTION MANUALp. 2/13 ALERTTMJ5 p. 3/13
Buttons and Menus
The ALERT J5 has an interface similar to a cellular
phone, which allows you to navigate all menus using the
two buttons.
Either button powers on the device. To navigate using the
two buttons:
• Use the right button to scroll down to your selection
(scrolling up is not possible)
• Use the left button to select (Enter) the option
The device has the following menus:
• Last 100 tests: Recalls and displays the last 100 test
results. You have the option of printing using the right
button. Refer to the “Printing” section for details
• Calibration: Shows the date of last calibration. You
can re-calibrate using the right button. Specialized
equipment is required to accurately calibrate ALERT
J5 and you should not attempt to calibrate it yourself
• Deactivation/Activation: A service menu option for
use by manufacturer or distributer. Units are supplied
to customer in “Activated” state
• Date/Time: Shows the current date and time of the
device. Use the right button to scroll through the digits
and the left button to change the values. You can also
set the time to a 12 or 24 hour format
• Settings:
- Change the units of the BrAC/BAC. Using the
right button, scroll to the desired unit and press
left button to select. The units available are: mg%,
%BAC, g/L, mg/L, and μg/100mL
- Backlight: Set the brightness of the backlight and
the time you would like the light to stay on

INSTRUCTION MANUALp. 2/13 ALERTTMJ5 p. 3/13
- Shutdown timer: Set the time for auto shut off. This
time can be between 30 seconds and 900 seconds
(15 minutes)
• Language: Set the display language using the right
button. Scroll through to view available options
• Shut down: Use this menu to manually shut down
the device without waiting for automatic shut off. This
option does not deactivate the device; it only shuts off
the device
BAC/BrAC Format
The blood alcohol concentration (BAC) or the breath
alcohol concentration (BrAC) format is available in ve
formats; mg%, mg/L, %BAC, g/L and μg/100mL. You can
select the desired unit from the Settings menu.
Power On/Shut Down
Press either of the two buttons to power on the device.
A ashing amber LED light indicates that the device is
warming up for a test, and you will see a WAIT message
on the display (for approximately 3 seconds). A green
ashing LED light with a READY message on display
indicates that the device is ready and you can now do a
breath test.
You can manually shut down the device by selecting
Shut down from the Main Menu. Alternatively, the
device shuts down automatically after a specic period
of time. You can set this time to suit your needs from the
Settings menu option.

INSTRUCTION MANUALp. 4/13 ALERTTMJ5 p. 5/13
Activating
ALERT J5 is shipped deactivated after it is calibrated
and passes inspection at the factory. This allows the
device to be stored without losing any time from the
calibration period. You have to activate the device to start
using it. Once activated, the date for the next calibration
increments by 365 days from the date of activation.
A Not Activated message when you initially power on
the device indicates that it is not activated and no test
can be performed until it is activated.
NOTE: This feature is displayed on the Main Menu, only
if your ALERT J5 has been previously calibrated. It is not
possible for you to deactivate the device.
To activate:
1. From the Not Activated screen, press the left button
to activate it. The date the device was deactivated is
shown.
2. Press the right button to activate the device. The
calibration due date is now 365 days from the current
date.
Battery
The ALERT J5 requires two AA Alkaline batteries to
operate. To insert, remove the battery compartment rear
cover (press the tab to release).
Table des matières
Autres manuels Alert Équipement de test
Manuels Équipement de test populaires d'autres marques

SMART
SMART KANAAD SBT XTREME 3G Series Manuel utilisateur

Agilent Technologies
Agilent Technologies BERT Serial Manuel utilisateur

Agilent Technologies
Agilent Technologies N3280A Manuel utilisateur

Vernier
Vernier Go Direct Voltage Manuel utilisateur

Lifeloc
Lifeloc R.A.D.A.R. Manuel utilisateur

Fluke
Fluke T5-600 Instructions d'utilisation et d'installation