Purpose
To install the Dual Plasma Burner mounting kit on a gas chromatograph (GC)
unit.
Scope
This procedure describes how to install the Dual Plasma Burner mounting kit.
Safety
Make sure all devices are powered off while performing the installation.
Materials
•A Dual Plasma Burner mounting kit
•The Dual Plasma Burner
•A Philips screwdriver
•3 mm Allen wrench
•Sheet metal cutting tool(s)
Procedure
1. Remove the mounting kit from the packaging.
2. Make sure the gas chromatograph (GC) is powered off and gas supplies
are off.
3. On the GC, remove the cover from the detector area to expose the top of
the GC, through which a detector is normally mounted. Find the
appropriate mounting location based on Figure 1.
4. Align the Burner mounting plate with the mounting screw holes on the
GC. Clear the hole into the GC oven of interfering insulation (approx
1.8 inch diameter).
Installing the Dual Plasma
Burner Mounting Kit on a
Shimadzu 2014 GC
Accessory G6600-60621