microAeth®AE51 Flow Calibration Kit Operating Manual microAeth®AE51 Flow Calibration Kit Operating Manual
Auto Flow Calibration
1. Always make sure that a lter strip is
installed in the microAeth AE51 when
it is operating.
2. When exchanging a lter strip in the
microAeth AE51, make sure that it is
turned off.
3. Install a clean, unused lter strip into
the microAeth AE51. A pre-used lter
with heavy loading may create an offset
in the ow calibration table of the AE51.
4. Make sure that the ow calibration kit is
connected to the AE51.
5. Connect the owmeter and microAeth
AE51 to the computer with the software
installed.
6. Turn on the owmeter and let it stabilize
for at least 10 minutes before use.
7. Turn on the microAeth AE51.
8. Open the AE51 Auto Flow Calibration
software.
9. Select the appropriate COM port from
the dropdown menu in the External
Flowmeter section of the software.
There may be multiple COM port
options depending on the computer
and the peripherals connected to it.
10. Click the Connect to Flowmeter button.
If the port is already in use, a window
will pop up to request that a different
port is selected. If the status bar at the
bottom of the software, does not show
‘microAeth ON, Flow Meter Connected’
status, check your connections and
ensure that communication with the
microAeth AE51 and owmeter has
been initiated as previously described.
11. Select the ow tolerance setting in the
Auto Calibration section.
The sofware will attempt to calibrate
the internal ow system of the
microAeth AE51 with 1% of the
external owmeter measurement. Some
instruments, especially older units that
have not had pump upgrade service
by AethLabs, may need to use the 3%
tolerance setting. Some instruments
may have internal pumps that cannot
reach the higher ow setpoints.
Instruments not sold by AethLabs may
require pump servicing by AethLabs
in order to achieve ow rates of 200
mlpm or higher. Contact AethLabs for
additional information.
12. Click the Calibrate button.
The system will automatically cycle
through all the setpoints of the ow
calibration table of the microAeth
AE51 (from 250 ml/min down to 0 ml/
min). The software will rst check the
maximum achievable ow rate of the
instrument to determine what setpoints
can be calibrated. If the internal pump
cannot achieve the specied ow rate,
the setpoint will not be calibrated.
Contact AethLabs for a pump upgrade,
replacement, or service.
13. Click the View Instrument Values
button. This will display all the values of
the ow calibration table.
14. Please check the values to make
sure that as the ow setpoint
increases from 0 to 250 ml/min, the
pump drive and internal owmeter
values also increase. If this is not the
case, please try again to calibrate the
microAeth AE51 using the auto and
manual ow calibrations. If this issue
persists, please contact AethLabs for
further assistance.