User manual soft air® simplex wds · Version 09/2018 9
• During use the mattress and tubes should be cleaned weekly using a damp soft
cloth and mild detergent or non-phenol germicidal solution.
Never use any detergents containing aldehyde!
• After changing to another patient the mattress should be disinfected by an
accredited company approved by the Robert-Koch-Institute. Please ensure
conformity to the European guideline for medical devices , directions for prepa
ration of medicine products and directions for hospital hygiene and prevention
of infections from the Robert-Koch-Institute or other guidelines they rule.
• The PU cover can be washed to a maximum temperature of 95 °C and be re-used
after a thermal process (e.g. Ottalin - Peracet - process).
Also a steam sterilisation process can be applied.
The power unit of the soft air simplex® wds system, should be cleaned with a
damp cloth and soap detergent or a mild non-aldehyde cleaning detergent.
• Disinfection with cleaning on locations is also possible, when instructions for use
and process time are properly kept.
• ADL® GmbH prescribes safety-related revisals of all electric powered parts after
at least 2 years.
The rules for accident prevention BGV A2 will remain unaffected.
The revisals may only be conducted by especially trained personal.
10 Cleaning & Service instructions
Disinfectant recommendations:
BACILLOCID RASANT (BODE), BACILLOL AF (BODE), BACILLOCID SPEZIAL (BODE),
KOHRSOLIN FF (BODE), KOHRSOLIN (BODE), MICROBAC FORTE (BODE), DISMOZON
PUR (BODE), INCIDIN FOAM (ECOLAB), INDUR DES (ECOLAB).
Check with your supplier the disinfectant does not damage plastic surfaces
(PUR, PA, ABS, PVC). Prevent contact on electrical components with damp
or moisture. Cleaning cloths should be damp, not wet. Carefully check no
moisture remains in the system. Execute a functions control procedure before
using the system again.
Check your cleaning process on effectiveness on a regular basis.