CONTENTS
P.N. 3.05.6.6.15667, Rev A, January 2020
1.0
INTRODUCTION................................................................................... 5
2.0
SPECIFICATIONS................................................................................ 6
3.0
INSTALLATION.................................................................................... 8
3.1
UNPACKING..........................................................................................................8
3.2
LOCATING.............................................................................................................8
3.3
SETTING UP THE SCALES ..................................................................................9
4.0
KEYPAD............................................................................................. 10
5.0
DISPLAY............................................................................................. 12
5.1
SYMBOLS AND INDICATORS............................................................................12
6.0
CALIBRATION COUNTER FOR APPROVED SCALES .................... 13
7.0
BATTERY........................................................................................... 13
8.0
BACKLIGHT....................................................................................... 14
9.0
AUTO POWER OFF ........................................................................... 14
10.0
OPERATION.................................................................................... 14
10.1
ZEROING..........................................................................................................14
10.2
TARING.............................................................................................................15
10.2.1
MANUAL TARE..........................................................................................15
10.3
WEIGHING........................................................................................................16
10.4
PARTS COUNTING..........................................................................................16
10.5
CHECK-WEIGHING..........................................................................................18
10.5.1
SETTING UP WHILE WEIGHING ..............................................................19
10.5.2
SETTING UP WHILE PARTS COUNTING OR % WEIGHING...................19
10.6
LIMITS STORING AND RECALLING...............................................................20
10.7
PERCENT WEIGHING......................................................................................21
10.8.1
ANIMAL WEIGHING PROCEDURE...........................................................24
10.9
ACCUMULATED TOTAL..................................................................................24
10.9.1
MANUAL ACCUMULATION......................................................................25
10.9.2
AUTOMATIC ACCUMULATION................................................................26
11.0
RS-232 SPECIFICATION................................................................. 27
11.1
INPUT COMMANDS FORMAT.........................................................................31
12.0
CALIBRATION................................................................................. 31
13.0
PARAMETER SETTINGS................................................................ 32
13.1
CHECK WEIGHING PARAMETERS................................................................32
13.2
RS-232 PARAMETERS....................................................................................34
13.3
SCALE PARAMETERS....................................................................................36
13.4
PERCENT WEIGHING AND ANIMAL WEIGHING...........................................37
14.0
ERROR MESSAGES ....................................................................... 38
15.0
REPLACEMENT PARTS AND ACCESSORIES.............................. 39
16.0
SERVICE INFORMATION................................................................ 40
17.0
WARRANTY INFORMATION........................................................... 41
18.0
APPENDIX....................................................................................... 42

© Adam Equipment Company 2020
5
1.0 INTRODUCTION
•The GBK/GFK-Mplus scales provides an accurate, fast and versatile
general purpose weighing scales with parts counting, percent weighing and
check-weighing functions.
•The GBK/GFK-Mplus has colour changing LCD illumination to indicate when
a weight is below the low limit, between the limits or above the high limit next
to the display. These can work in conjunction with an audible alarm for
check weighing.
•The GBK/GFK-Mplus scales are supplied with a RS-232 bi-directional
interface and real time clock (RTC).
•The GBK/GFK has a sealed keypad with colour coded buttons and a large
easy to read liquid crystal type display (LCD) supplied with a multicolour
illumination.
•Includes automatic zero tracking, semi-automatic tare and accumulation
facility that allows the weight to be stored and recalled as an accumulated
total.
•OIML Approved models, GBK/GFK-Mplus, only allow “Kg” and “g” as units,
have calibration controlled by jumpers or passcodes and other limitations as
noted in the manual.

© Adam Equipment Company 2020
6
2.0 SPECIFICATIONS
GBK-Mplus Models
Model # GBK 6Mplus
GBK 15Mplus GBK 30Mplus GBK 60Mplus GBK 150Mplus
Maximum Capacity 6kg 5kg 30kg 60kg 150kg
Readability 2g 5g 10g 20g 50g
Repeatability (Std
Dev) 4g 10g 20g 40g 100g
Linearity + 4g 10g 20g 40g 100g
Units of Measure Kilograms, grams
Stabilization Time 2 Seconds typical
Operating
Temperature -10°C to +40°C / +14°F to +104°F
Power Supply 110V - 240 VAC 50/60 Hz
Calibration External
Display 6 digits LCD digital displays with capacity tracker and symbols for units
Pan Size 300mm x 400mm x 50mm / 11.8” x 15.5” x 2”
Overall Dimensions
(w x d x h) 300mm x 520mm x 660mm / 11.8” x 20.5” x 26”
Net Weight 7.4kg
Features Weighing, Check Weighing, Parts counting, Check Counting, Animal Weighing,
Accumulating memory
Interface RS-232 bi-directional interface
English, German, French, Spanish, Italian and Portuguese selectable text

© Adam Equipment Company 2020
7
FK-M Models
Model # GFK 60Mplus GFK 150Mplus GFK 300Mplus
Maximum Capacity 60kg 150kg 300kg
Readability 20g 50g 100g
Repeatability (Std Dev) 40g 100g 200g
Linearity + 40g 100g 200g
Units of Measure Kilograms, grams
Stabilization Time 2 second typical
Operating
Temperature -10°C to +40°C / +14°F to +104°F
Power Supply 110V - 240 VAC 50/60 Hz
Calibration External
Display 6 digits LCD digital displays with capacity tracker and symbols for units
Pan Size 400mm x 500mm
Overall Dimensions
(w x d x h) 400 mm x 620 mm x 860 mm
Net Weight 12.5kg
Features Weighing, Check Weighing, Parts counting, Check Counting, Animal
Weighing, Accumulating memory
Interface RS-232 bi-directional interface
English, German, French, Spanish, Italian and Portuguese selectable
text

© Adam Equipment Company 2020
8
3.0 INSTALLATION
3.1 UNPACKING
The GBK/GFK-Mplus scales have already been adjusted to work with a platform and
have been configured for this application. The platform and indicator have been calibrated
as a pair and must be used together.
3.2 LOCATING
•The scales should not be placed in a location that
will reduce the accuracy.
•Avoid extremes of temperature. Do not place in
direct sunlight or near air conditioning vents.
•Avoid unsuitable tables. The table or floor must be
rigid and not vibrate.
•Avoid unstable power sources. Do not use near
large users of electricity such as welding equipment
or large motors.
•Do not place near vibrating machinery.
•Avoid high humidity that might cause condensation.
Avoid direct contact with water. Do not spray or
immerse the scales in water.
•Avoid air movement such as from fans or opening
doors. Do not place near open windows or air-
conditioning vents.
•Keep the scales clean. Do not stack material on the
scales when they are not in use.

© Adam Equipment Company 2020
9
3.3 SETTING UP THE SCALES
•The pillar is attached to the base using a bracket that must be attached to the base
frame first using the 4 bolts supplied. The pillar is secured to the bracket using 2
sets of screws. The cable from the base to the indicator module is run through the
tube and taken out through the plastic support at the top. Excess cable can be
stored within the tube.
•The GBK/GFK-Mplus series comes with a stainless steel platform packed
separately. Place the platform in the base.
•Level the scale by adjusting the four feet. If the scale rocks re-adjust the feet.
•Attach the indicator module to the pillar by sliding it over the bracket with the
flanges engaged in the groves on the base. Attach the cable from the base to the
connector on the rear of the indicator.
•Attach the power to the indicator. Press the [On/Off] key. The software revision
number will be displayed followed by a self-test showing all digits before the zero is
displayed along with the unit of weight that was selected last.
•If the scale is an approved version, GBK/GFK-Mplus, and the Calibration Counter
has been enabled (see section 6) the current values will be displayed. These
values should match the values marked on the scale at the time of verification, if it
has been verified.

© Adam Equipment Company 2020
10
4.0 KEYPAD
KEYS PRIMARY FUNCTION SECONDARY FUNCTION
[Zero] Sets the zero point for all subsequent
weighing. The display shows zero.
Escape from any setting menus
[Tare] It tares the scale and stores the current
weight in memory as a tare value,
subtracts the tare value from the weight
and shows the results. This is the net
weight.
Accept the set values
[Unit] This is used to select the weighing units
from a preset list of available units.
Allows the weight, unit weight,
and count to be seen when
parts counting or to change from
weight to % in percent weighing
[Low
Limit]
&
[High
Limit]
It sets the limits for check weighing and
allows setting of either the low limit or
the high limit or both.
None
Ce manuel convient aux modèles suivants
9
Table des matières
Autres manuels Adam Échelle

Adam
Adam GC INDICATOR Manuel utilisateur

Adam
Adam AGB35 Manuel utilisateur

Adam
Adam TBB Manuel utilisateur

Adam
Adam Swift SWZ Series Manuel utilisateur

Adam
Adam LBX C Series Manuel utilisateur

Adam
Adam ACBplus Series Manuel utilisateur

Adam
Adam GBK 16a Manuel utilisateur

Adam
Adam AGF Series Manuel utilisateur

Adam
Adam CBC-M Series Manuel utilisateur

Adam
Adam WBW 2L Manuel utilisateur